Measuring the Replenishment Volume
Figure 9
Replenishment
Calibration
Switch
H148_0112GCA
H148_0112GA
[17] Press the "UPDATE CAL VOL" key.
[18] Use the up and down arrow keys to increase or decrease the volume displayed on the screen until the volume
displayed matches the average volume of replenishment as determined in Step 15.
60 mL = MEASURED DEVELOPER VOLUME
[19] Press the "DONE/RETURN" key repeatedly to return to the menu shown below.
[20] Repeat Steps
6
through
[21] Press the "DONE/RETURN" key repeatedly until you return to the display shown below. Then press the "YES"
key to return to the Walk-Up Menu.
EXIT SETUP?
[22] Install the Top Cover of the Processor.
5B6729 – May 1997
Fixer
Replenishment
Hose
Graduated
Cylinder
Quick
Disconnect
UPDATE
CAL VOL
DEV
FIX
CAL
CAL
18
to calibrate the fixer (or developer) replenishment.
YES
Caution
Wear protective eyewear when executing the following
procedure. Replenishment solutions are pumped
quickly and may splash.
[7] Remove the Top Cover of the Processor.
[8] Disconnect the Quick Disconnect for the fixer
replenishment (or developer replenishment).
[9] Insert the end of the Fixer Replenishment (or
Developer Replenishment) Hose into a Graduated
Cylinder.
[10] Hold the Graduated Cylinder outside of the
Processor so that spills will not contaminate the
processing solutions.
[11] Press the Replenishment Calibration Switch
located next to the replenishment hoses.
[12] Measure and record the volume of replenishment
delivered by the system.
[13] Dispose of the solution in the Graduated Cylinder.
[14] Do Steps 8 through 11 at least 2 more times.
[15] Determine the average volume of replenishment
delivered.
[16] Install the Replenishment Hose into the respective
quick disconnect Valve.
DONE/
RETURN
DONE/
RETURN
DONE/
RETURN
CANCEL
REQUEST
Basic Setup Options
27