Source; Frequency; Level Offset State,Level Offset; Filter - Rohde & Schwarz R&S SGS100A User Manual

Sgma rf source
Hide thumbs Also See for R&S SGS100A:
Table of Contents

Advertisement

®
R&S
SGS100A
Remote command:
:SENSe<ch>[:POWer]:ZERO
Source ← Sensor Settings
Selects the source for measurement.
"RF"
"User"
Remote command:
:SENSe<ch>[:POWer]:SOURce
Frequency ← Sensor Settings
Defines the frequency value if "Source > User" is used.
Remote command:
:SENSe<ch>[:POWer]:FREQuency
Level Offset State,Level Offset ← Sensor Settings
Activates and defines a level offset which is considered in the power measurement
result. The level offset value is always expressed in dB, irrespective of the display of
the measurement result.
This function allows you to consider, for example, an attenuator in the signal path.
Remote command:
:SENSe<ch>[:POWer]:OFFSet
:SENSe<ch>[:POWer]:OFFSet:STATe
Filter ← Sensor Settings
Selects the way the length of the used filter is defined.
See also
stable results"
"Auto"
User Manual 1173.9105.02 ─ 13
Assigns the signal to the RF path of the R&S SGS.
The R&S SGS transfers the RF frequency and level settings to the
R&S power sensor automatically. Thus you achieve power readings
of high accuracy, irrespective from the connected sensor type.
Sets a user defined frequency.
Example:
If you have a frequency converting device between the generator and
the DUT. If the frequency converter doubles the frequency, you can
set twice the frequency in the R&S SGS. The R&S power sensor con-
siders this RF frequency setting.
Set the parameter
"About the measuring principle, averaging filter, filter length, and achieving
on page 118.
Selects the filter length automatically and adjusts it to the measured
value. The value is indicated with the parameter
When high output power is applied, the filter length and therefore the
measurement time can be short.
When low output power is applied, the filter length and therefore the
measurement time is increased which reduces the considered noise
content in your measurement.
on page 259

Frequency

to the measurement's frequency.
on page 257
on page 256
on page 257
on page 257
Signal generator settings
Level and power-on settings

Filter

Length.
122