You must install the correct peristaltic pump tubing for the wavelength
NOTE
calibration:
4100 MP-AES
4200 and 4210 MP-AES
2
3
4
5
6
7

Creating/Opening a Worksheet

Creating a new worksheet

To create a new worksheet, click New from the Start page or the File
menu.
A list of recently used files will be presented when creating a new
worksheet from a template; otherwise you may Browse for more files.
The 'New From Template' dialog box will be displayed in this
instance.
Agilent Microwave Plasma Atomic Emission Spectrometers User's Guide
Click the Instrument button.
Click Calibration.
Place the solution inlet tubing into the wavelength calibration
solution and allow the sample to reach the plasma.
Click Calibrate Instrument. The torch will be aligned and then a
wavelength calibration and a calibration check will be performed
automatically.
After a short while, an indication of the success or failure of the
calibration check will appear, as well as an indication of the
wavelength offsets.
If the calibration fails, check the sample introduction system. If
the system seems fine, prepare a new wavelength calibration
solution and try again, or consult the Troubleshooting section of
the MP Expert Help.
Pump
White/white
Orange/green
Operation
Drain
Blue/blue
Blue/blue
39