Turning On

CAUTION
Turn on the power supply (PAT-TMX series)
10
The CONFIG settings can be configured so that the output is automatically turned on when
the POWER switch is turned on. When this function is enabled, the PAT-TM/TMX series
powers up with the output turned on even if the output was off when the PAT-TM/TMX series
was turned off the last time. However, there is a possibility that a load may break, if you
connect a different load and turn the power and output on simultaneously without changing
the OVP and OCP settings to appropriate values.
With the PAT-TMX series, you can turn the power on or off by flipping the breaker switch while
each power supply
s POWER switch is in the on position.
'
1
Check that the power cable is correctly connected.
2
Check that the POWER switch of each power supply unit is turn on.
3
Raise the lever of breaker switch to turn the power on.
If an odd sound, odd odor, fire, or smoke occurs around or in the PAT-TMX series,
lower the breaker lever to turn it off.
4
Check the firmware version on the front panel display of each power
supply.
The voltmeter and ammeter show the PAT-T series firmware version for approximately
1 second.
Indicates the version
OUTPUT
ALARM
Firmware version display at power-on (example for Ver.3.00)
The measuring value will be shown after the firmware version is displayed.
The master unit information can also be obtained using the *IDN? remote control command.
The model name, serial number, and firmware version will be returned.
With the PAT-TMX series, an inrush current of approximately 500 Amax. (approximately 200
Amax. for 400 V input model) flows when the breaker switch is flipped to the on position (in
the case of the 40 kW systems). Check that sufficient current capacity is available in the AC
power line and the switchboard.