Pulse Dose Mode Specifications; Power Accessory Specifications; Ac Power Supply; Dc Power Supply - CAIRE Eclipse 5 with autoSAT Technical & Service Manual

Table of Contents

Advertisement

Back-Up Alarm Power
Filters
Device Classification

Pulse Dose Mode Specifications

Pulse Settings
Trigger Sensitivity
Adjustable Bolus Rise Time
Trigger Criteria
Minimum time between breaths
Response to Missing Breaths
The expected service life is a minimum of 5 years.
Pulse Dose
Setting
1.0
2.0
3.0
4.0
5.0
6.0
7
8
9

Power Accessory Specifications

AC Power Supply

Input Voltage
100-240VAC, 50-60 Hz
Input Power
245-260 VA
Output Voltage
28 VDC
Output Power
200W

DC Power Supply

Input Voltage
11.5-16VDC
Output Voltage
26 VDC
Output Power
150W Max
9V Internal Battery
Air Inlet, HEPA, Compressor Intake
IEC Class I, Type B Applied Part, IP0 (3 LED model)
IEC Class II, Type BF Applied Part, IP22 (2 LED model)
1.0 to 6.0: 16mL, 24mL, 32mL, 40mL, 48mL, 56mL, 64mL, 72mL, 80mL, 88mL, 96mL;
7-9: 128mL, 160mL, 192mL
Adjustable between settings of 1 (most sensitive) to 3 (least sensitive)
Adjustable settings of Fast(factory setting), Medium, or Slow
Cannula pressure has dropped below the trigger point (typically between 0.135—0.37 cm of H
of negative pressure)
At least 1¼ seconds has passed since the last pulse began
1.25 seconds (max. 3 consecutive breaths)
While in Pulse Dose Mode, the Eclipse 5 is always monitoring for breath detection. After 15 seconds of no
breath detected, the Eclipse 5 "delivers" Continuous Flow at the last Continuous Flow setting.
After another 15 seconds, the Eclipse 5 stops delivering Continuous Flow and waits 15 seconds for a
breath. The Eclipse 5 will stay in this modality until a breath is detected. (See Power Cartridge
Conservation Feature, page 26.)
Bolus Size
(± 15%) mL
16
32
48
64
80
96
128
160
192
NOTE: Bolus volume decreases as breath rate exceeds published range.
Output voltage
Capacity
Nominal Power
Cartridge Life
Power
Cartridge
Recharge Time
AC Power Supply and Power
Cartridge (Battery)
Max Breath
Rate
40
40
40
40
37
31
23
18
15

Power Cartridge (Battery)

14.8 VDC
Quantity (2) 97.5 W-hrs batteries
(Each containing 7.92 grams equivalent Lithium content)
80% Capacity after 500 Charge/Discharge cycles
1.8 to 5.0 hours, dependent on flow setting, to achieve 80%
capacity from a fully discharged Power Cartridge
Oxygen Concentrator Technical Service Manual | PN 20631679 Rev G
O
2
DC Power Supply
Max Breath
Rate
40
40
40
31
25
20
15
12
10
8