CONTROL PANEL

The control panel provides a user interface consisting of a membrane panel keyboard, Liquid Crystal Display (LCD), external power present indicator,
Power Cartridge capacity indicator, alarm status indicators, and an audio transducer. The user interface informs the user of the system status and
allows the user to set the desired flow rate and flow mode.
User Control Panel (3 LED model)
ON/OFF Button (Green) Indicator: This button powers
the device ON or OFF. The Green Indicator is
illuminated when the device is ON and functioning
properly.
Increase or Decrease Flow Setting Buttons: Use these buttons
to set the flow to your prescribed setting.
Delivery Mode Button and Indicator: The button toggles between
Continuous Flow and Pulse Dose Mode. The Pulse Dose Mode activates
autoSAT Technology and allows a significant increase in the operating
time while powered by the battery. When the Pulse Dose Mode is
activated, the green Pulse Dose Mode Indicator illuminates and a pulse of
oxygen is delivered with each inspiratory effort.
ALERT (Yellow) Indicator—Low and Medium Priority Alerts: When
illuminated, this indicates a low priority awareness condition or Caution.
Continue to use your system and refer to the Troubleshooting Table for
the proper response, or contact your Home Care Provider. A flashing
yellow indicates a medium priority alert. A prompt response is necessary.
(Typically low oxygen or low battery.)
ALARM (Red) Indicator—High Priority Alarms: Indicates a high priority
alarm condition. An immediate response is necessary. Refer to the
Troubleshooting Table and contact your home care provider. (Typically loss
of power, low oxygen, or malfunction.)
Flow Setting Indicator: This is the main focus on your control panel. Your
home care provider will correctly set your prescribed flow for the
Continuous Flow Mode (LPM) and/or your Pulse Dose Mode (mL) settings.
Each time you power the device ON, the previous mode and/ or setting has
been saved and will be used at start-up.
Power Cartridge (battery) Status Gauge: This indicator displays the
charge remaining in the battery. Each of the five horizontal gray bars
represents approximately 20% of the total battery charge. When the
battery is being charged, the charge indicator bars will blink in a waterfall-
type fashion. If the battery is not installed, or if it is improperly installed,
the Power Cartridge (battery) Status Gauge will not be illuminated or will
flash on and off.
External Power Indicator: When the Eclipse 5 is properly plugged in and is
using the AC or DC Power Supply, this indicator will appear on the
User Control Panel.
Buzzer: An audible alarm (or buzzer) alerts you to the operating condition
of the device, either a warning or failure, and confirms a valid key press.
User Control Panel (2 LED model)
ON/OFF Button (Green) Indicator: This button powers the
device ON or OFF. The Green Indicator is illuminated when the
device is ON and functioning properly.
Increase or Decrease Flow Setting Buttons: Use these buttons
to set the flow to your prescribed setting.
Delivery Mode Button and Indicator: The button toggles between
Continuous Flow and Pulse Dose Mode. The Pulse Dose Mode
activates autoSAT Technology and allows a significant increase in the
operating time while powered by the battery. When the Pulse Dose
Mode is activated, the green Pulse Dose Mode Indicator illuminates
and a pulse of oxygen is delivered with each inspiratory effort.
Low Priority Technical Alarm: When illuminated, this indicates a low
priority awareness condition. Continue to use your system and refer to
the trouble shooting table for the proper response, or contact your
Home Care Provider.
Flow Setting Indicator: This is the main focus on your control panel. Your
home care provider will correctly set your prescribed flow for the
Continuous Flow Mode (LPM) and/or your Pulse Dose Mode (mL)
settings. Each time you power the device ON, the previous mode and/or
setting has been saved and will be used at start-up.
Power Cartridge (battery) Status Gauge: This indicator displays the
charge remaining in the battery. Each of the five horizontal gray bars
represents approximately 20% of the total battery charge. When the
battery is being charged, the charge indicator bars will blink in a
waterfall-type fashion. If the battery is not installed, or if it is improperly
installed, the Power Cartridge (battery) Status Gauge will not be
illuminated or will flash on and off.
External Power Indicator: When the Eclipse 5 is properly plugged in
and is using the AC or DC Power Supply, this indicator will appear on
the User Control Panel.
Buzzer: An audible alarm (or buzzer) alerts you to the operating condition of the
device, either a warning or failure, and confirms a valid key press.
Oxygen Concentrator Technical Service Manual | PN 20631679 Rev G
14

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents