PM100 Personal Single Gas Monitor
If calibration fails, the icon
representative or AimSafety Technical Support.
6.4 Span Calibration
When performing a Span Calibration only use certified calibration gas at the required
concentration level. Do not use expired calibration gas.
To perform a Span Calibration:
•
Access the Calibration menu.
•
"CAL" and the
•
Press the [Function key] once and "CAL" and the
•
Attach the calibration (cal) cap on top of the sensor inlet.
•
Connect the hose from the gas regulator of the calibration gas bottle to the cal cap
•
Press and hold the [Function key] for 5 seconds to start the Span Calibration.
•
When the calibration starts, a 90-second countdown displays.
Note: The countdown is only 60 seconds for O
•
Turn on the calibration gas.
•
Once completed, the
the display.
•
The device will return to Measuring mode.
•
Turn off the calibration gas and remove the calibration cap.
•
Clear the Peak values for the sensor. (See
All alarms are muted for approximately one minute after a successful span calibration.
REV – 3.0
Figure 7-2
will appear on the display. If this continues, please contact the sales
WARNING
icon are displayed.
icon and the current gas measurement readings will appear on
Clear
WARNING
[34-2900-0206-2]
icon will appear.
2
Peaks)
18 |