If calibration fails, the icon
representative or AimSafety Technical Support.

6.4 Span Calibration

When performing a Span Calibration only use certified calibration gas at the required
concentration level. Do not use expired calibration gas.
To perform a Span Calibration:
Access the Calibration menu.
"CAL" and the
Press the [Function key] once and "CAL" and the
Attach the calibration (cal) cap on top of the sensor inlet.
Connect the hose from the gas regulator of the calibration gas bottle to the cal cap
Press and hold the [Function key] for 5 seconds to start the Span Calibration.
When the calibration starts, a 90-second countdown displays.
Note: The countdown is only 60 seconds for O
Turn on the calibration gas.
Once completed, the
the display.
The device will return to Measuring mode.
Turn off the calibration gas and remove the calibration cap.
Clear the Peak values for the sensor. (See
All alarms are muted for approximately one minute after a successful span calibration.
REV – 3.0
Figure 7-2
will appear on the display. If this continues, please contact the sales
WARNING
icon are displayed.
icon and the current gas measurement readings will appear on
Clear
WARNING
[34-2900-0206-2]
icon will appear.
2
Peaks)