Bump Test Reminder - Macurco AimSafety PM100 User Instructions

Personal single gas monitor
Hide thumbs Also See for AimSafety PM100:
Table of Contents

Advertisement

PM100 Personal Single Gas Monitor
When enabled, the monitor will prompt the user to perform the test by displaying "StS" on the display
when the test is due.
To activate the test, press the [Function key].
The unit will then perform the following:
Buzzer test
LED test
Vibration
LCD test
Display Alarm set point 1
Display Alarm set point 2
End
The user is required to ensure that all the tests pass successfully and that the alarm values are set to the
proper levels.
Do not use the monitor if any of the tests fail or if the alarm values are incorrect

4.7 Bump Test Reminder

The Bump Test due reminder option can be enabled using the PM Link and software or the Bump Test-
Calibration Station. The Bump Test reminder can be configured from n/a (Off) to 365 days. The default
configuration is set to "Off", no Bump Test reminder will be displayed.
When enabled, the monitor will prompt the user to perform the test by displaying "btS" on the display
when the test is due.
Ensure that the sensor is reading zero (or 20.9% for Oxygen)
Attach the calibration (cal) cap on top of the sensor inlet
REV – 3.0
Figure 5-6
WARNING
[34-2900-0206-2]
12 |
P a g e

Advertisement

Table of Contents
loading

Table of Contents