Cleaning Labels; Particulate Entry/Exit Apertures; Compliance Audit Check - Trolex XD One User Manual

Hide thumbs Also See for XD One:
Table of Contents

Advertisement

XD One_User manual_P5644.1601_Rev B

15.3 Cleaning labels

It is recommended to periodically clean the instrument with a damp cloth, to ensure the
instrument user interface and keypad is clean and legible.

15.4 Particulate entry/exit apertures

The particulate exit aperture is protected by a stainless-steel grille to minimise the ingress of
flora and fauna into the XD One device. It is recommended that the grille it is checked and
cleaned during maintenance periods to ensure that it has not become clogged with ingress
that may obscure the particulate sensing airflow.

15.5 Compliance audit check

The XD One has been designed with an inbuilt compliance check routine/test to allow for the
infrequent checking of device functionality against a selection of sized reference particulates.
The Compliance Check uses certified sample material that can be passed through the sensor
to ensure that all sensing and sizing parameters are functioning as intended. Reference
material with a size spread ranging from 0.35 to 40µm allows each sensing region to be
populated with reference data during the process.
To run the Compliance Check sequence, connect the XD One to the
BreatheLITE
software
application, position in the supplied compliance base and cover with the particulate dispersion
hood before selecting the automated test routine within the device maintenance menu.
The use of the compliance base and particulate dispersion hood allows for units to be isolated
within a known volume during testing as highlighted below.
1. Insert the XD One into
2. Cover with particulate
3. Run
BreatheLITE
the compliance dock.
dispersion hood.
compliance check.
42
XD ONE USER MANUAL