In 'Normal' mode, the user-selected alarm PM size readings are taken every second and used
to update the applicable STEL and TWA calculations which are then checked against the
configured alarm thresholds.

12.2 'In-cab' mode

The XD One 'In-cab' mode allows for the same operating functionally as 'Normal' mode but
can also be continuously used in a vehicle when connected or powered via USB.
This mode is indicated via the solid illumination of the battery symbol during operation as
highlighted below.
The XD One will turn itself off after five minutes after USB power has been removed to save
power.

12.3 'Data log' mode

The XD One 'Data log' mode follows the operational functionality of 'Normal' mode, but
readings are not checked against alarm threshold levels and on-device warnings/alarms will
not trigger.
This mode is intended for use during passive monitoring of environments where data is
captured for download and post-analysis using the
BreatheLITE
software.
This mode is indicated via two turquoise illuminated icons as highlighted below.

12.4 'Live readout' mode

The 'Live readout' mode allows the XD One to stream live particulate data to the
BreatheLITE
software via USB connection.
Particulate data is also logged to the internal memory for download and analysis as required.
This mode is indicated by two violet illuminated icons as highlighted below.