result via an audible visual alarm.
4. The XD One will automatically begin to
sample particulate concentrations.

11.7 Accidental activation of power/function switch

The XD One has a single power/function switch that has been designed for all device input
operations. All positive inputs are completed via press, hold and release to prevent accidental
activation of the power switch during operation.
In the event that the XD One power/function switch is accidentally activated during operation,
the device will alert the user to the prolonged button press and hold via a series of sequential
audio alarms.
Once the power/function switch is released, the XD One will automatically perform a self-test
and revert to normal sensing operation.

11.8 Power management

The XD One uses a single battery icon to identify the battery life of the device. The internal
battery has a 16-hour operating capacity when fully charged and it is recommended that the
device is charged between shifts to ensure maximum operational charge during use.
During normal operation, the battery indication icon is used to display the following battery
status.
Green flashing icon (intermittent at three-second intervals)
When the XD One is displaying a green flashing icon, the battery charge level is between
100 and 70%.
Amber flashing icon (intermittent at three-second intervals)
When the XD One is displaying an amber flashing icon, the battery charge level is between
69 and 40%.
Red flashing icon (intermittent at three-second intervals)
When the XD One is displaying a red flashing icon, the battery charge level is between
39 and 10%. The red flashing icon will blink at one-second intervals when the battery
decreases below 10%.
30
XD ONE USER MANUAL
XD One_User manual_P5644.1601_Rev B
'Sample' routine