General description............................1 Main features................................2 Limits of use................................2 Product safety..............................3 Danger from process............................4 Safety procedures.............................5 Laser safety precautions..........................5 System components............................7 TX8100 Air XS RCS Monitor..........................8 5.2 Particulate flow path............................9 Peripherals and accessories...........................10 Certification.................................11 Compliance................................11 Technical information............................12 Product specification............................12 Product dimensions............................14 Hardware installation............................15 Safety precautions.............................15...
Page 3
Troubleshooting...............................30 15.1 High temperature operating..........................30 15.2 Fault codes.................................30 Glossary and definitions..........................31 Disposal................................32 Get in touch...............................33 18.1 Feedback................................33 Disclaimer................................33 Trademark................................34...
As the Air XS processes and analyses particulate data, an adaptive algorithm is used to calculate and identify the average level of RCS from the overall particle count. Measurement information can be viewed via the instrument display or as historical readings using the accompanying Trolex Breathe application software.
‘Plug and play’ installation 1.2 Limits of use To ensure the optimum performance and safe operation, the Air XS must be operated as per the limits detailed in the technical data section of this user manual. Operation outside these limits may result in damage to the equipment or failure to achieve the performance specification.
In addition, the following symbols are used on the instrument WARNING! – ELECTRIC SHOCK RISK WARNING! – LASER RADIATION The use of controls, adjustments, or procedures other than those specified in this user manual may result in exposure to hazardous optical radiation. AIR XS USER MANUAL...
3. DANGER FROM PROCESS It is possible that the Air XS could be installed in environments that contain process particulates which can be hazardous to health. Unless process conditions are known to be entirely safe, suitable precautions such as the use of breathing apparatus or environmental purging/detoxifying should be employed before entry is made into the installation or maintenance environment.
It is important that you carefully read the entire User Manual before using and installing the Air XS for the first time and keep it in a safe place for future reference. Peripheral components such as the power supply and communications module/peripheral or interface must be installed according to the manufacturer’s instructions and the installation...
Page 9
Sensor housing WARNING – There are no user servicable parts inside the Air XS. Servicing should only be carried out by Trolex or an approved service technician.
5. SYSTEM COMPONENTS The Air XS is typically installed as a stand-alone instrument for general purpose applications. The instrument is supplied with peripherals fitted to allow for the plug and play installation to universal mains/DC power supplies and data outputs.
5.1 TX8100 Air XS Silica Monitor The Air XS RCS monitor uses a custom particle sensor that is located inside a robust stainless-steel housing. This provides isolation and ingress protection between the particle flow path and the main control circuits. Control circuits are housed in a lockable, IP66 rated steel enclosure. Information and settings can be accessed using the keypad and display located on the front of the instrument.
5.2 Particle flow path The Air XS has been designed with the ability to restrict ingress through the particulate flow path during routine maintenance and cleaning periods. A rotational top cap is used to open or close the particle flow path to provide ingress protection during cleaning.
5.3 Peripherals and accessories Power supply The Air XS can be connected to a standard 100V - 240V AC power supply OR a 9 – 18V dc power supply. Before connecting a power supply to the instrument, ensure that the supply source is compatible with the instrument and information outlined on the appropriate rating plate.
6. CERTIFICATION 6.1 Compliance The Air XS complies with the following European Union Directives: Electromagnetic Compatibility (EMC) Directive 2014/30/ - EN 61326-1:2013 Low Voltage Directive (LVD) 2014/35/EU - EN 61010-1:2010+A1:2019 The Air XS complies with the following RoHS Directives: - RoHS Directive 2002/95/EC...
*The instrument can define particulate measurement peak trends up to the quantity specified. Note: Sustained exposure to dust quantities above 25 mg/m³ will be logged; however, it may affect the operating life of the Air XS sensor. As applications of use are outside our control, the information provided is given without legal responsibility.
Page 16
Custom alarm set points Latching/non-latching Indicators 1x Green High Brightness LED – Sensor heartbeat 1x Blue High Brightness LED – Communications Self-test routine Sensor hardware, Circuitry and Communications on ‘power on’ Manual self-test during use Certification CE compliant AIR XS USER MANUAL...
8. HARDWARE INSTALLATION 8.1 Safety precautions Refer to section 4 of this user manual before undertaking the installation of the Air XS device. The installation location of the Air XS device is the prerogative of the installer and care should be taken to ensure an appropriate position has been selected.
Note: Ensure the Air XS is mounted vertically during installation. 4x M10 The figure and tables below detail the connections available internally in the Air XS. The connections can be accessed by opening the front housing of the instrument using the supplied key.
Supply voltage Supply voltage V out RS485/USB switch 0V return 0V return RS485 B Earth Earth RS485 A V out RS485 0V V out Ethernet Table 1: Power, RS485, External power, USB and Ethernet terminal connections. AIR XS USER MANUAL...
8.4 I/O terminals Power, RS485, 4-20mA and relay connection terminal data is highlighted below. Actuation type Operating tool Solid/stranded conductor 0.08 - 2.5mm² / 28 - 12 AWG Conductor with ferrule 0.25 - 1.5mm² Strip length 5 - 6mm / 0.2 - 0.24 inch 8.5 Power connections For instruments connected to an AC power supply, it is the responsibility of the installer to ensure that the instrument is installed with the following protection:...
9. COMMISSIONING Once installed and powered for first time use, the Air XS will begin to monitor and sample the passing environmental particulates. The first-generation Air XS does not allow for custom user parameters or set-up adjustment. 9.1 First power on Prior to commissioning and first use, the instrument should be inspected for any visible damage and integrity of the enclosure.
9.4 Application software The Trolex Breathe application software is available for download and installation from the Trolex website and is specifically designed for the import and review of data sets collected by the Air XS instrument. Note: The Trolex Breathe application software is not required for general Air XS operation.
10. CONTROLS AND INDICATORS 10.1 Navigation The Air XS user interface is controlled and navigated using the on-device keypad and display. The keypad consists of four directional keys and a central enter key to allow the scrolling, selection and input of data into the instrument.
11. GENERAL OPERATION 11.1 User interface display On initial power up of the Air XS, particulate sampling will automatically start, and the graphical display will show the following screens. Once the instrument has been auto-configured, the Air XS will begin sampling the environmental dust mixtures.
11.2 Particulate display information The following information is presented on the Air XS display during particle sampling and operation. RCS content rolling USB connected icon averaging peridod RCS content (mg/m3) Total particles/litre AIR XS USER MANUAL...
Page 27
During normal operation the Air XS will capture the associated function data and generate a single log file for the relevant time span – this typically results in one file a day. Note: If the Air XS is restarted, a new file will be generated and saved for the associated period. 11.2.1 Data download The Air XS allows users to download collected data sets via the ‘data download’...
12. INSTRUMENT CONFIGURATION 12.1 Default settings The Air XS has been programmed with factory default settings prior to delivery which are detailed in the table below: RCS averaging period 15 minutes, 1 hour, 4 hours, 8 hours and 12 hours (rolling avg)
12.3 Custom alarms The first-generation Air XS has a single-user configurable alarm that can be used to warn users of a specific particulate threshold breach based on a chosen limit or value. To enable the alarm functionality, follow the sequence outlined below: 1.
13. UPDATING UNIT FIRMWARE On notification of firmware update and release, the Air XS can be updated via the on-device firmware update functionality. The instrument requires the update to take place via the connection of an external USB device loaded with the latest firmware.
The particulate exit aperture is protected by a stainless-steel grille to minimise the ingress of flora and fauna into the Air XS instrument. It is recommended that the grille is checked and cleaned during maintenance periods to ensure that it has not become clogged with ingress that may obscure the particulate sensing airflow.
14.4 Filter replacement The Air XS uses two HEPA inlet and outlet filters for the purpose of providing a clean sample air flow within the device. It is recommended that these filters are replaced every month to maintain the performance of the device.
Air XS is installed in environments with high ambient temperatures. 15.2 Fault codes The following codes relate to on-screen warnings that the Air XS will display when a fault is encountered during normal operations. Code...
Particulate matter (PM) General term for a mixture of solids and liquid droplets suspended in the air from typical processes including combustion, industrial activities or natural sources is particulate matter x micrometer or less in diameter AIR XS USER MANUAL...
17.1 Waste of Electrical and Electronic Equipment (WEEE) Directive (2012/19/EU) The Air XS operates a protective thermal cut-out sequence when the temperature inside the instrument exceeds the maximum operating temperature specification, detailed in section 7. This protective measure is in place to maintain the lifespan and operating functionality of the optical sensor assembly when the Air XS is installed in environments with high ambient temperatures.
Please contact our technical services team: Telephone: +44 (0)161 483 1435 Email: service@trolex.com 18.2 Feedback If you have any suggestions for improvements or amendments, or find errors in this publication, you can contact marketing@trolex.com directly. AIR XS USER MANUAL...
No part of this document may be reproduced in any form or by any means, electronic or mechanical, including photocopying, without express written permission of Trolex. Trolex is a registered trademark of Trolex Ltd. The use of all trademarks in this document is acknowledged.
Page 38
Our aim is to become the world’s leading name in health and safety technology, through pioneering products that provide real-world benefits to our customers, whenever workers operate in hazardous environments. For more information about Trolex, please contact us at: Enquiries Trolex Ltd sales@trolex.com...