Multimeter - Fluke 5320A Getting Started Manual

Multifunction electrical tester calibrator
Hide thumbs Also See for 5320A:
Table of Contents

Advertisement

Uncertainty............................................. 0.02 %
Settling Time ..................................................... 300 ms to 3 s, depending on output value
AC Voltage
Range
3 – 29.99 V
30 – 99.99 V
100 – 299.9 V
300 – 600 V
DC Voltage
Range
3 – 29.99 V
30 – 149.9 V
150 – 600 V
AC Output Signal Distortion ............................ 0.2 % ±10 mV (harmonic distortion and non-harmonic noise from 20
Sensing Ammeter Current Range ................... 500 mA
Resolution ......................................................... 1 mA
Uncertainty ........................................................ ±5 mA

Multimeter

Voltage
Range ................................................................. 0 to 1100 V ac rms or dc
Resolution ......................................................... 4½ digits
Internal Ranges ................................................. 10, 100, and 1100 V (Autoranging only)
Frequency Range.............................................. DC, 20 Hz to 2 kHz
Input Resistance ............................................... 10 MΩ ±1 %
Time Constant ................................................... 1.5 s
Readings/Second.............................................. 2
Measurement Category .................................... 1000 V CAT I, 300 V CAT II
Range
10 V
100 V
1100 V
Current
Range ................................................................. 0 to 20 A continuous, 30 A for up to 30 minutes, ac rms or dc
Resolution ......................................................... 4½ digits
Internal Ranges ................................................. 300 mA, 3 and 30 A (Autoranging only)
Frequency Range.............................................. DC, 20 to 400 Hz
Time Constant ................................................... 1.5 s
Readings/Second.............................................. 2
Uncertainty and Maximum Burden Current
Resolution
0.001 V
0.01 V
0.1 V
0.1 V
Uncertainty and Maximum Burden Current
Resolution
0.001 V
0.01 V
0.1 V
Hz to 500 kHz), for output power lower than 10 VA on each range.
AC/DC Voltage Uncertainty
Resolution
0.001 V
0.01 V
0.1 V
Multifunction Electrical Tester Calibrator
Uncertainty
±(% of Reading + mV)
0.1 % + 9
0.1 % + 30
0.1 % + 90
0.1 % + 180
Uncertainty
±(% of Reading + mV)
0.1 % + 9
0.1 % + 45
0.1 % + 180
±(% of Reading + mV)
Electrical Specifications
Maximum Burden Current
500 mA
300 mA
150 mA
50 mA
Maximum Burden Current
2 mA
3 mA
5 mA
Uncertainty
0.15 % + 5
0.20 % + 50
0.20 % + 550
9

Advertisement