How to proceed if a repair is necessary ................25 Dismount..........................26 Dismounting steps......................26 Disposal ......................... 26 Supplement ..........................27 Technical data ........................ 27 Dimensions ........................35 Industrial property rights ....................38 Trademark ........................38 VEGASWING 61 • Transistor (NPN/PNP)
Page 3
Contents Safety instructions for Ex areas: Take note of the Ex specific safety instructions for Ex applications. These instructions are attached as documents to each instrument with Ex approval and are part of the operating instructions. Editing status: 2023-01-19 VEGASWING 61 • Transistor (NPN/PNP)
Symbols used Document ID This symbol on the front page of this instruction refers to the Docu- ment ID. By entering the Document ID on www.vega.com you will reach the document download. Information, note, tip: This symbol indicates helpful additional infor- mation and tips for successful work.
During work on and with the device, the required personal protective equipment must always be worn. Appropriate use The VEGASWING 61 is a sensor for point level detection. You can find detailed information about the area of application in chapter " Product description". Operational reliability is ensured only if the instrument is properly used according to the specifications in the operating instructions manual as well as possible supplementary instructions.
That is why we have introduced an environment management system with the goal of continuously improving company environmental pro- tection. The environment management system is certified according to DIN EN ISO 14001. Please help us fulfil this obligation by observing the environmental instructions in this manual: • Chapter " Packaging, transport and storage" • Chapter " Disposal" VEGASWING 61 • Transistor (NPN/PNP)
– If necessary, further certificates Information: Optional instrument features are also described in this operating instructions manual. The respective scope of delivery results from the order specification. Constituent parts The VEGASWING 61 consists of the components: • Housing lid • Housing with electronics •...
Principle of operation Application area VEGASWING 61 is a point level sensor with tuning fork for point level detection. It is designed for industrial use in all areas of process technology and can be used in liquids.
The instructions for the listed accessories can be found in the down- load area on our homepage. Flanges Screwed flanges are available in different versions according to the following standards: DIN 2501, EN 1092-1, BS 10, ASME B 16.5, JIS B 2210-1984, GOST 12821-80. VEGASWING 61 • Transistor (NPN/PNP)
Page 10
For connecting the sensors with a separator to voltage supply or sig- nal processing, the sensors are also available with plug connectors. The following plug connectors are available: • M12 x 1 • ISO 4400 • Harting HAN 7D • Harting HAN 8D • Amphenol-Tuchel VEGASWING 61 • Transistor (NPN/PNP)
Switching point In general, VEGASWING 61 can be installed in any position. The instrument only has to be mounted in such a way that the tuning fork is at the height of the desired switching point.
Page 12
Switching point approx. 27 mm (1.06 in) Fig. 3: Horizontal mounting Switching point Marking with screwed version, facing up Fig. 4: Horizontal installation (recommended mounting position, particularly for adhesive products) Switching point Marking with screwed version, facing up In the case of flange versions, the fork is aligned as follows. VEGASWING 61 • Transistor (NPN/PNP)
Page 13
To maintain the housing protection, make sure that the housing lid is closed during operation and locked, if necessary. Caution: Transport Do not hold VEGASWING 61 on the tuning fork. Particularly with flange or tube versions, the tuning fork can be damaged just by the weight of the instrument. Transport coated instruments very carefully and avoid touching the tuning fork.
Mounting instructions Welded socket VEGASWING 61 has a defined thread starting point. This means that every VEGASWING 61 is in the same fork position after being screwed in. Remove therefore the supplied seal from the thread of VEGASWING 61. This seal is not required when using a welded socket with O-ring in front.
Page 15
See " Technical data". Caution: If it is determined (e.g. via a fault signal from VEGASWING 61) that medium has already penetrated into the vibrating element, the instru- ment must be exchanged immediately.
Connect the voltage supply according to the following diagrams. Take note of the general installation regulations. As a rule, connect VEGASWING 61 to vessel ground (PA), or in case of plastic vessels, to the next ground potential. On the side of the instrument housing there is a ground terminal between the cable entries.
Plastic (not with Ex d) Aluminium Stainless steel (not with Ex d) Stainless steel, electropolished (not with Ex d) Filter element for pressure compensation or blind plug with version IP66/ IP68, 1 bar (not with Ex d) VEGASWING 61 • Transistor (NPN/PNP)
Page 18
Ground terminal Connection terminals Wiring plan We recommend connecting VEGASWING 61 in such a way that the switching circuit is open when there is a level signal, line break or failure (safe state). The instrument is used to control relays, contactors, magnet valves, warning lights, horns as well as PLC inputs.
Wire assignment, con- nection cable Fig. 11: Wire assignment connection cable. The numbers of the wires corre- spond to the terminals of the instrument. brown (+) voltage supply White Yellow blue (-) voltage supply Shielding VEGASWING 61 • Transistor (NPN/PNP)
DIL switch for adjustment of the density range (3) Note: Always immerse the tuning fork of VEGASWING 61 in a liquid to test its function. Do not test the function of VEGASWING 61 with your hand. This can damage the sensor.
Switching status Control lamp Mode A closed Overflow protec- tion Green Mode A open Overflow protec- tion Mode B closed Dry run protection Green Mode B open Dry run protection Failure of the sup- open ply voltage (mode A/B) VEGASWING 61 • Transistor (NPN/PNP)
Page 22
6 Setup Level Switching status Control lamp Fault open flashes red VEGASWING 61 • Transistor (NPN/PNP)
24 hour service hotline Should these measures not be successful, please call in urgent cases the VEGA service hotline under the phone no. +49 1805 858550. The hotline is also available outside normal working hours, seven days a week around the clock.
In Ex applications only an electronics module with respective Ex ap- proval may be used. You can find all the information you need to carry out an electronics exchange in the handbook of the new electronics module. In general, all electronics modules of series SW60 can be inter- changed. If you want to use an electronics module with a different signal output, you carry out the complete setup. You find the neces- sary, suitable operating instruction on our homepage. VEGASWING 61 • Transistor (NPN/PNP)
Clean the instrument and pack it damage-proof • Attach the completed form and, if need be, also a safety data sheet outside on the packaging • Ask the agency serving you to get the address for the return ship- ment. You can find the agency on our homepage. VEGASWING 61 • Transistor (NPN/PNP)
If personal data is stored on the old device to be disposed of, delete it before disposal. If you have no way to dispose of the old instrument properly, please contact us concerning return and disposal. VEGASWING 61 • Transistor (NPN/PNP)
Ʋ The Second Line of Defense (SLOD) is a second level of the process separation in the form of a gas-tight feedthrough in the lower part of the housing, preventing product from penetrating into the housing. Ʋ Supporting material 316L VEGASWING 61 • Transistor (NPN/PNP)
Page 28
Floating transistor output, permanently shortcircuit-proof Load current < 400 mA Voltage loss < 1 V Switching voltage < 55 V DC Blocking current < 10 µA Modes (switchable) Ʋ A Max. detection or overflow/overfill protection Ʋ B Min. detection or dry run protection VEGASWING 61 • Transistor (NPN/PNP)
Page 29
200 °C (482 °F) (122 °F) (212 °F) (302 °F) (392 °F) (32 °F) Fig. 13: Influence of the process temperature on the switching point Shifting of the switching point in mm (in) Process temperature in °C (°F) Switching point at reference conditions (notch) Tuning fork VEGASWING 61 • Transistor (NPN/PNP)
Page 30
Shifting of the switching point in mm (in) Process pressure in bar (psig) Switching point at reference conditions (notch) Tuning fork Non-repeatability 0.1 mm (0.004 in) Hysteresis approx. 2 mm (0.08 in) with vertical installation Switching delay Ʋ Standard approx. 500 ms (on/off) VEGASWING 61 • Transistor (NPN/PNP)
Page 31
Ʋ VEGASWING 61 of 316L/Alloy C22 -50 … +250 °C (-58 … +482 °F) (2.4602) Ʋ VEGASWING 61 with ECTFE coating -50 … +150 °C (-58 … +302 °F) Ʋ VEGASWING 61 with PFA coating -50 … +250 °C (-58 … +482 °F) Ʋ...
Page 32
(302 °F) (392 °F) (482 °F) Fig. 18: Process temperature - Process pressure with switch position ≥ 0.5 g/cm³ (sensitivity switch) Process pressure in bar (psig) Process temperature in °C (°F) SIP process temperature (SIP = Sterilization in place) PFA and ECTFE coatings are not suitable for SIP cleaning Vapour stratification up to 2 h +150 °C (+302 F) VEGASWING 61 • Transistor (NPN/PNP)
Page 33
25 mm (0.984 in) with 25 °C (77 °F) Ʋ Diameter approx. 8 mm (0.315 in) Ʋ Colour - standard PE Black Ʋ Colour - standard PUR Blue Ʋ Colour - Ex-version Blue Adjustment elements Mode switch Ʋ A Max. detection or overflow/overfill protection VEGASWING 61 • Transistor (NPN/PNP)
Page 34
For that reason the associated approval documents of these instruments have to be carefully noted. They are part of the delivery or can be downloaded by entering the serial number of your instrument into the search field under www.vega.com as well as in the general download area. A suitable cable is required for maintaining the protection rating.
Les lignes de produits VEGA sont globalement protégées par des droits de propriété intellec- tuelle. Pour plus d'informations, on pourra se référer au site www.vega.com. VEGA lineas de productos están protegidas por los derechos en el campo de la propiedad indus- trial. Para mayor información revise la pagina web www.vega.com.