5 ADDITIONAL INFORMATION

5.1

Maintenance

Your spectrometer comes fully NIST traceable calibrated.
The spectrometer requires a recalibration yearly, regardless of the operating hours. If a spectrometer
calibration is necessary, please contact Image Engineering.
Please send the complete device to Image Engineering.
Pack the CAL3 with a notation for calibration in the hard case from its delivery.
Please contact support@image-engineering.de for conditions and procedures.
After calibrating the spectrometer, perform a spectral calibration (iQ-LED calibration).
We also suggest you again generate all the illuminants you use.
5.2

Care instructions

Do not touch, scratch, or pollute the diffuser.
If there is any dust on the diffuser, clean it with an air blower.
Do not remove the fiber from the spectrometer. Otherwise, the calibration is invalid, and
the spectrometer must be recalibrated!
Only store and transport the CAL3 in the delivered hard case.
5.3

Disposal instructions

After the service life of CAL3, it must be disposed of properly. Electrical and electromechanical
components are included in CAL3. Observe your national regulations and ensure that third parties
cannot use CAL3 after disposing of it.
Contact Image Engineering if assistance for disposal is required.