5 ADDITIONAL INFORMATION

5.1 Maintenance

The spectrometer requires a recalibration once a year, regardless of the operating hours. If a
spectrometer calibration is necessary, please contact Image Engineering. We will provide you
with removal instructions for the spectrometer. The spectrometer then has to be sent to our
office with a notation for calibration. Use appropriate packaging to prevent the spectrometer
from being damaged during shipping.
A temporary replacement spectrometer can be provided to bridge the time gap while your
spectrometer is recalibrated. Please contact us for conditions and procedure.

5.2 Care instructions

Do not touch, scratch, or pollute the diffusor.
If there is any dust on the diffusor, clean it with an air blower.
Do not remove the fiber from the spectrometer. Otherwise, the calibration is invalid, and
the spectrometer has to be recalibrated!
Only store and transport the CAL1 in the delivered hard case.

5.3 Disposal instructions

After the service life of CAL1, it must be disposed of properly. Electrical and electromechanical
components are included in CAL1. Observe your national regulations. Make sure that CAL1
cannot be used by third parties after disposing of it.
Contact Image Engineering if assistance for disposal is required.