Mounting
High viscosity
NOTICE
Highly viscous liquids may cause switching delays.
Make sure that the liquid can run off the tuning fork easily.
Deburr the socket surface.
High viscosity, e.g. viscous oils: < 10 000 mPa⋅s
The tuning fork must be located outside the installation socket!
 6
Installation example for a highly viscous liquid. Unit of measurement mm (in)
4.1.3
Avoid buildup
• Use short installation sockets to ensure that the turning fork can project freely into the
vessel
• Install preferably flush mount on vessels or in pipes
• Leave sufficient distance between the buildup expected on the tank wall and the tuning fork
10
> 40 (1.57)
Liquiphant FTL41
A0037348
Endress+Hauser