Page 3
Preface For this manual, the issued Date is October 2005 (Version: 1.5). Safety Symbols Warning A Warning indicates that failure to follow proper instructions can cause death or injury to the patient, the operator, or serious damage to the equipment. Caution A Caution indicates that failure to follow proper instructions may cause serious injury to the patient, the operator, or may cause damage to the...
Page 4
Illustrations All illustrations in this manual are provided as examples only. They may not necessarily reflect your setup or data displayed on your Pulse Oximeter.
Chapter 1 Safety Information This chapter contains important safety information related to general use of the Pulse Oximeter. Other important safety information appears throughout the manual in sections that relate specifically to the precautionary information. Note Important! Before use, carefully read this manual, all safety information and specifications.
Page 8
Safety Information Warning The Pulse Oximeter is to be operated only by medical doctors or trained healthcare personnel to measure the SpO of clinical patients (including adult, pediatric and neonatal patients) in hospitals and healthcare agencies or in the process of transporting patients.
Page 9
Safety Information recommended methods to sterilize and disinfect the Pulse Oximeter or its accessories when the packing material is damaged, or it has not been used for a long time. Caution The Pulse Oximeter should be appropriately placed. Keep it from falling, strong vibration or other mechanical damage.
Chapter 2 General 2.1 Introduction The Pulse Oximeter is a portable measuring device powered by common or rechargeable batteries. It is compact, light, flexible to use and easy to learn. Parameters measured by the Pulse Oximeter include: arterial oxygen saturation (SpO ), pulse rate (PR) and pulse strength.
General 2.3 Appearance ① ② ③ ④ ⑤ ⑥ ⑦ ⑧ Figure 2-1 Front Panel Figure 2-2 Back Panel Table 2-1 Appearance description Description Remarks ① Dual-purpose It connects SpO2 sensor or PC communication cable. socket ② It displays information listed in Table 2-2. ③...
General 2.4 Displayed Information Figure 2-3 shows information displayed on the LCD screen. ⑤ ① ② ⑥ ③ ⑦ ⑧ ④ ⑨ Figure 2-3 Displayed Information Table 2-2 Description of displayed information Description Remarks ① ID number It displays current ID number ranging from 000 to 100 ②...
General 2.5 Button Operation Three soft buttons are available on the front panel. Power Backlight Confirm ID Figure 2-4 Buttons 2.5.1 Power Button Power-on: Press to turn on the device. Power-off: Press and hold for two seconds to turn off the device. Note The Pulse Oximeter is powered by batteries only.
General 2.5.3 Confirm ID Button The Confirm ID button is used to add data. It enables the user to add data, without using a new ID number, to a re-test for the same patient after the sensor is accidentally disconnected. Refer to section 5.1 Data Management for detailed use of this button.
Chapter 3 Installation 3.1 Unpacking and Inspection Please carefully remove the Pulse Oximeter and its accessories from the package and check items by comparing them to the packing list. Check the device for any mechanical damages. Check all cables and accessories for damage. In case of any problem, please contact our Customer Service.
Page 18
Installation shown in Figure 3-2. 5. Push back the battery cover. Note Before disposing of the battery, check with your local solid waste officials for details in your area for recycling options or proper disposal. Caution Please use AA alkaline batteries or rechargeable batteries. Do not use carbon or poor quality batteries.
Installation 3.3 Power-on Press the Power button to turn on the Pulse Oximeter. The startup interfaces and version information will be displayed, and then the initialization interface as shown in Figure 3-3. After that, the Pulse Oximeter will automatically switch to the standby mode as shown in Figure 3-4.
Installation 3.5 Connect Computer The Pulse Oximeter can be connected to a Personal Computer through a communication cable to transmit patient’s trend to the computer for printing. Simply connecting one end of the PC communication cable to the dual-purpose socket and the other end to the PC’s serial port. This symbol will be display in the LCD screen indicating that the Pulse Oximeter has connected with the PC successfully.
Chapter 4 Measurement 4.1 Measuring Principle The Pulse Oximeter is capable of measuring SpO , PR and pulse strength. The Pulse Oximeter measures SpO by a method called pulse oximetry. It is a continuous, non-invasive method based on the different absorption spectra of reduced hemoglobin and oxyhemoglobin.
Measurement 4.2 Precautions Note Do not perform SpO monitoring and NIBP measurements on the same arm simultaneously. Obstruction of blood flow during NIBP measurements may adversely affect the reading of the SpO value. It is recommended that each measurement last longer than 15 seconds.
Page 23
Measurement Warning Prolonged and continuous monitoring may increase the risk of burns at the site of the sensor. If you have to use the Pulse Oximeter for Prolonged and continuous monitoring, it is especially important to check the sensor placement, and ensure proper attachment on neonates and patients of poor perfusion or skin sensitive to light.
Measurement 4.3 Measuring Steps The measurement is usually done using the adult finger SpO sensor. Palm or foot sensors may be adopted for infants. Before measurement, check the SpO sensor and its cable for damages. Do not use the sensors if any damage is found. 4.3.1 Adult Measurement Please follow the steps below to use the adult finger SpO sensor:...
Measurement 4.3.2 Neonatal Measurement Please follow the steps below to use the neonatal SpO sensor: Insert the SpO sensor‘s connector into the dual-purpose socket. Turn on the Pulse Oximeter to enter the standby mode. Appropriately place the neonatal SpO sensor (refer to 4.3.3 Placing Neonatal SpO Sensor).
Page 26
Measurement Figure 4-3 Placing Neonatal SpO Sensor 2 Wind the SpO sensor around a hand or foot of a neonate patient. Hold the sensor, pull the belt and fit one of its sides with “V” edge into the “V” groove on the corresponding side of the sheath. Appropriately elongate the belt to about 20mm, and fit the “V”...
Measurement Figure 4-4 Placing Neonatal SpO Sensor 3 4.4 Measuring Restrictions If the accuracy of any measurement seems unreasonable, first check the patient’s vital signs by an alternate method, and then check the device for proper function. Inaccurate measurements may be caused by the following reasons. Incorrect sensor application or use;...
Page 28
Measurement Placement of a sensor on the same extremity with a blood pressure cuff, arterial catheter, or intravascular line. Loss of pulse signal can occur in the following situation: The sensor is too tight; There is excessive illumination from light sources such as a surgical lamp, a bilirubin lamp, or sunlight;...
Chapter 5 Other Functions 5.1 Data Management 5.1.1 Data Storage The Pulse Oximeter has its internal memory to store data. The memory is divided into the ID Data Zone and Trend Data Zone. The ID Data Zone is capable of storing 100 patients’ ID data at most. When the number exceeds 100, new data will automatically cover the old one from the earliest stored data.
Other Functions The first trend data will be stored 15 seconds after the pulse is found. Thereafter one trend data will be stored every 2 minutes. One trend data includes: Average SpO value within the 2 minutes. Average PR value within the 2 minutes. Note The first trend data are the instant values of SpO and PR.
Other Functions 5.1.3 Data Protection The Pulse Oximeter has data protection function. When the power is turned off accidentally during the process of storing a data, the Pulse Oximeter will evaluate completeness of the last stored data when it is restarted.
Other Functions 5.2 Messages Prompting The Pulse Oximeter can display various prompt messages. In table 5-1, prompt messages as well as their causes and solutions are listed. Table 5-1 Table of Indications Message Cause Solution Batteries energy lower “Low Battery” Replace batteries in time than 4.0 Voltage.
Page 33
Other Functions The Pulse Oximeter can also display technical error messages. In table 5-2, error messages as well as their causes and solutions are listed. If the LCD screen can’t display anything, it may be damaged or error occurs during system self-test. Please shut down the device (if can’t, remove the batteries) and contact our Customer Service.
Other Functions 5.3 Power Management Battery Detection The Pulse Oximeter can detect the battery energy and display message when battery voltage is less “Low Battery” than 4.0 V; shut down the device automatically when battery voltage is less than 3.85V. Energy Saving The Pulse Oximeter can save the energy of the batteries by switching to the standby mode automatically when the finger...
Chapter 6 Maintenance 6.1 System Check Before using the Pulse Oximeter, perform the following steps. Check for any mechanic damages. Check for all cables and accessories for damage. Check all functions of the Pulse Oximeter to make sure the Pulse Oximeter is in proper working condition. If any damage, malfunction or potential safety hazard is found, stop using the Pulse Oximeter on patients.
Maintenance 6.2 General Cleaning The Pulse Oximeter must be kept dust-free. Regular cleaning (depending on the policy of your institution) of its exterior surface and LCD screen is strongly recommended. First absorb non-corrosive detergent with clean and soft cloth or cotton ball. Manually twist the cloth or cotton ball to a proper degree and then use the dried cloth or cotton ball for cleaning.
Page 37
Maintenance the amount of organisms (blood or mucus) on the surface to be cleaned. Caution Do not use ammonia-based or acetone-based cleaners such as acetone. Most detergents must be diluted before use. Please follow the manufacturer’s directions carefully. Do not use abrasive material, such as steel wool. Do not allow any liquid inside the device.
Maintenance 6.3 Sterilization/Disinfection Sterilization or disinfection may, to a certain degree, damage the Pulse Oximeter and SpO sensor. It is not recommended unless stipulated as necessary in the Hospital Maintenance Schedule. Cleaning is recommended before sterilization or disinfection. Recommended disinfection materials: ethylate, and acetaldehyde. The SpO cable can be cleaned with 3% hydrogen peroxide solution or 70% isopropanol solution.
Chapter 7 Pulse Oximeter Management System The Pulse Oximeter Management System software (PMS software) is developed to realize more functions of the Pulse Oximeter. The PMS software runs in Windows 98/2000/XP operation system. In conjunction with the internal software of the Pulse Oximeter, the following functions can be realized.
Pulse Oximeter Management System and then click “Next” in the next dialog box. Input the correct serial number, and click “Next” to resume. Select the serial port to connect the Pulse Oximeter with your PC, and click “Next” to resume. Choose the destination folder where the PMS software is to be installed.
Pulse Oximeter Management System 7.2 Main Interface Double clicking the shortcut icon for PMS software on the desktop of your computer, the Main Interface of PMS software will be displayed as shown below. ① ② ③ ③ Figure 7-1 Main Interface ①...
Page 42
Pulse Oximeter Management System <File> Click the <File> menu to see its pull-down menu as shown in Figure 7-2. There are five submenus: <File Management>: Click to open the “File Management” dialogue box. <Print>: Click to print the current patient data. <Print Preview>: Click to preview the data to be printed.
Page 43
Pulse Oximeter Management System When you use the Data Output or Software Upgrade function, the default serial port might have been occupied. At this time, you can click the <Serial Port Selection> menu to select other serial port. <Operation> Click the <Operation> menu to see its pull-down menu as shown below. There are two submenus: Figure 7-4 Operation Menu <Data Output>: Click to conduct the “Data Output”...
Pulse Oximeter Management System Note When you open the “Help” document, if a dialog box pops up and informs you to install language pack, please select “Never install any language packs”, and then click “Cancel” button. 7.2.2 Tool Bar In the tool bar, you can see the following shortcut icons. Table 7-1 Shortcut Icons Icon Icon Name...
Pulse Oximeter Management System It displays the value of the measured SpO and PR as well as the corresponding Index No and Save Time of each measurement. The content in the Information Area can’t be directly inputted or changed. Please select <Patient Information> under the <Setup> menu, or click the shortcut icon.
Page 46
Pulse Oximeter Management System When data outputting is finished, the “Save As” dialog box will pop up as shown below. Figure 7-7 Save the Outputted File You can choose the file directory where the data is to be stored, and change the file name. The default file directory is the “Files”...
Pulse Oximeter Management System When this situation occurs, please check whether the serial port is correctly connected, and try to select another serial port by clicking the <Serial Port Selection> menu. 7.3.2 Software Upgrade With the PMS software, you can upgrade the internal software of Pulse Oximeter.
Page 48
Pulse Oximeter Management System Figure 7-11 Open Upgrade File Select the Upgrade File and click the “Open” button. The version of the Upgrade File will be displayed in the figure 7-10. Click the “Upgrade” button, the system will check the validity and verify version of the Upgrade File.
Page 49
Pulse Oximeter Management System Figure 7-13 Software Upgrade If the version of Upgrade File is lower or the same as that loaded on the Pulse Oximeter, corresponding different message will be displayed in Figure 7-12. Click “Yes” to continue to upgrade, and “Cancel” to cancel upgrading.
Pulse Oximeter Management System 7.3.3 File Management File management function helps you to open or delete the data outputted conveniently. Click <File Management> from the <File> menu, the “File Management” dialogue box will pop up as shown in Figure 7-16. Figure 7-16 File Management You can choose the folder where the outputted data is saved in the File List.
Pulse Oximeter Management System 7.3.4 Modify Patient Information Click <Patient Information> from the <Setup> menu, the “Modify Patient Information” dialog box will pop up as shown in Figure 7-17. Figure 7-17 Modify Patient Information You can input the following information: Name: 30 characters at maximum.
Pulse Oximeter Management System 7.3.5 Print Data Click <Print Setting> from the <File> menu to display the “Print Setting” dialogue box. You can set the print range of the patient ID. Figure 7-18 Print Range Click “OK” and set the properties according to the printer your personal computer installed in the pop-up dialog box.
Pulse Oximeter Management System You can also click <Print> from the <File> menu, and click <OK> button in the pop-up dialogue box to start printing. 7.3.6 Prompt Message Table 7-2 Prompt Message Error Message Cause Solution Communication Restart the Pulse Oximeter, The connection between Error, and check the connection...
Page 54
Pulse Oximeter Management System FOR YOUR NOTES...
Appendix A Specifications Classification Item Description Electroshock Type IIb device according to the 93/42EEC directive. proof type: Anti-electroshock Powered by internal batteries. type: Anti-electroshock Type BF degree: Harmful liquid IPX0. Ordinary sealed equipment without liquid proof function proof degree: Disinfection Refer to Chapter 6 Maintenance /Sterilization: Working mode:...
Page 57
Specifications Dimension: 65mm×140mm×32mm Weight: About 130g(not including battery and SpO2 sensor) Functions Item Description Memory Full ID Full Indications: Low battery Standby Technical error Power saving Automatic standby/shutdown features: Printer: The PC’s printer Printing: Paper: Content: ID data and trend data Downloading: Download new software revision through PC serial port.
Page 58
Specifications Electrical specifications Item Description Working voltage: 4.0 to 6.4 VDC Power supply: Batteries Battery Common 1.5V AA alkaline or rechargeable batteries specifications: Shutdown < 200uA leakage current: Battery run time: 15-hour continuous operation with alkaline batteries. Power 720mW Consumption &PR specifications Item Item...
Need help?
Do you have a question about the PM-50 and is the answer not in the manual?
Questions and answers