Note: The Passport 8 and Passport 12 Quick Reference Guide is not intended as a replacement to the Operating Instructions. Prior to operating equipment, the user must be familiar with the Operating Instructions Manual contents. This document is a guideline only to be used as an aid to comprehensive...
Page 5
Have a split-screen view of Display more QuickKeys minitrend Enter the volume setup Hide the QuickKeys menu Default configurations Enter the main menu Start cardiac output Enter standby mode procedure Perform calculations Change alarm settings Have a split-screen view of Review the patient's data another patient's conditions Have a split-screen view of...
The Patient Setup menu allows the clinician to admit a patient and enter patient demographics. Admitting also restores default (preconfigured) settings and clears previous patient’s information. To admit a patient and enter demographics on the Passport 8 or Passport 12 bedside monitor: a. Select Patient Setup>> using Patient Setup...
Panorama Central Station and WorkStation. Adjust Patient Size The Patient Setup menu allows the clinician to adjust patient size. To adjust patient size on the Passport 8 or Passport 12 bedside monitor: a. Select Patient Setup>> using Patient Setup (quick key) or Main...
Discharge A Patient The Discharge function clears all patient information and restores default (preconfigured) settings. To Discharge a patient on the Passport 8 or Passport 12 bedside monitor: a. Select Patient Setup>> using Patient Setup (quick key) or Main Menu (quick key) b.
Standby as a patient cannot be removed from Standby on the Panorama Central Station. RECORD Real-Time Waveforms To record a strip on a patient on the Passport 8 or Passport 12 bedside monitors’ integrated recorder select Record on the monitor’s front key pad. P/N: H-046-005847-00 Rev 3.0...
PRINT Real-Time Waveforms To print a strip on a patient on the Passport 8 or Passport 12 bedside monitor to a predetermined laser printer select Real-Time Print (quick key). DPM Central Station Interface: Selecting Real-Time Print on the Passport 8 or Passport 12 bedside monitor will print a 10-second, Real-Time Report on the DPM Central Station.
Passport 12 bedside monitor, Panorama Central Station, WorkStation, and ViewStation are independent. Enable Pacemaker Detection And Set Pacemaker Rate The Passport 8 or Passport 12 bedside monitors allow the clinician to enable pacemaker detection. The symbol is displayed when the pacemaker detection is enabled.
NIBP Initiate A Non-Invasive Blood Pressure Measurement (NIBP) To initiate a NIBP measurement on a patient on the Passport 8 or Passport 12 bedside monitor a. Select NIBP on the front panel of the monitor a.
DPM Central Station Interface: Selecting Stop NIBP Measurement on the DPM Central Station will stop a NIBP measurement on the Passport 8 or Passport 12 bedside monitor. To stop all NIBP measurements on a patient on the Passport 8 or Passport 12 bedside monitor: a. Select Stop all measurements (quick key) a.
Quick Reference Guide INVASIVE BLOOD PRESSURES (OPTIONAL) The Passport 8 allows a clinician to monitor up to 2 invasive blood pressures (IBP) with the included Y-type IBP adapter and external module. The Passport 12 bedside monitors allow a clinician to monitor up to 4 invasive blood...
The Alarm Setup tab allows the clinician to configure alarm limits and alarm responses on the Passport 8 or Passport 12 bedside monitors. To adjust a patient’s alarm limit on the Passport 8 or Passport 12 bedside monitor: a. Select parameter’s waveform or numeric tile to open parameters Setup menu b.
Passport ® Quick Reference Guide Adjust Alarm Responses To adjust a patient’s alarm responses on the Passport 8 or Passport 12 bedside monitor: a. Select parameter’s waveform or numeric tile to open parameters Setup menu b. Select Alarm Setup >>...
Quick Reference Guide DPM Central Station Interface: Changing alarm responses on the DPM Central Station will not change alarm responses on the Passport 8 or Passport 12 bedside monitor. Changing alarm responses on the Passport 8 or Passport 12 bedside monitor will not change alarm responses on the DPM Central Station.
(quick key) or to return to the main display RESET ALARMS To silence an alarm for the duration of the alarm on the Passport 8 or Passport 12 bedside monitors select Alarm Reset key on the front panel of bedside monitor.
Selecting Mute on the Panorama Central Station will not silence an alarm on the Passport 8 or Passport 12 bedside monitor. SILENCE ALL ALARMS To silence all alarms on the Passport 8 or Passport 12 bedside monitor for a preconfigured amount of time, select Alarm Pause key on front panel of bedside monitor.
*If enabled DPM Central Station Interface: Transferring patient data from one Passport 8 or Passport 12 bedside monitor to another Passport 8 or Passport 12 bedside monitor will not transfer data within the DPM Central Station. Transferring data within a DPM Central Station or to a different DPM Central Station may affect the Passport 8 or Passport 12 bedside monitor.
PA catheter when the PA distal balloon is inflated and the catheter advances and occludes a distal pulmonary artery. PAWP pressure is a reflection of the pressure in the left ventricle at end-diastole. To initiate a PAWP measurement on the Passport 8 or Passport 12 bedside P/N: H-046-005847-00 Rev 3.0...
Hemodynamic calculations are a set of values that are used to determine the hemodynamic status of the patient. To initiate hemodynamic calculations on the Passport 8 or Passport 12 bedside monitor: a. Select CALC>> from the C.O. menu or PAWP menu or HEMODYNAMIC...
Note: Hemodynamic Calculations performed and stored on the Passport 8 or Passport 12 bedside monitor can be recorded only on the Passport 8 or Passport 12 bedside monitor. DPM Central Station Interface: Hemodynamic Calculations performed on the DPM Central Station can be printed and stored on the DPM Central Station.
Page 24
Normal Screen selection. DPM Central Station Interface: Digital and waveform display selections on the DPM Central Station are independent of the Passport 8 or Passport 12 bedside monitor. Panorama Central Station Interface: Digital and waveform display selections on the Panorama Central Station, WorkStation, and ViewStation are independent of the Passport 8 or Passport 12 bedside monitor.
Passport ® Quick Reference Guide HISTORY Tabular Trends To display Tabular (List) trends on a patient on the Passport 8 or Passport 12 bedside monitor: a. Select Review>> (quick key) or from Main Menu (quick key) b. Select Tabular Trends tab c.
Passport ® Quick Reference Guide To view and print an alarm event on a patient on the Passport 8 or Passport 12 bedside monitor: a. Select Review >> (quick key) or from Main Menu (quick key) b. Select Events tab c.
The alarm volume, key volume, and QRS volume can be adjusted on the Passport 8 or Passport 12 bedside monitors. To adjust the alarm volume on the Passport 8 or Passport 12 bedside monitor: a. Select Volume (quick key) or select Others Tab in Alarm Setup (quick key) b.
Passport ® Quick Reference Guide Beat Tone To adjust the beat tone (QRS tone) volume on the Passport 8 or Passport 12 bedside monitor: a. Select ECG waveform or ECG numeric tile to enter ECG Setup>> b. Select Others>> c. Select QRS Volume d.
Select Yes To remove a patient from Privacy Mode and resume monitoring on the Passport 8 or Passport 12 bedside monitor select any key on the Passport 8 or Passport 12 bedside monitor DPM Central Station Interface: Privacy Mode monitor can be enabled/ disabled on the DPM Central Station.
® Quick Reference Guide DATE AND TIME To adjust the date and time on the Passport 8 or Passport 12 bedside monitor: a. Select Date/Time next to patient size on main display b. Set the date and time c. Select Date Format to adjust the month, day, and year format d.
® Quick Reference Guide PARAMETER COLORS The Passport 8 or Passport 12 bedside monitor provides the ability to adjust the parameter’s numeric and waveform (if applicable) color. To adjust the parameter’s color on the Passport 8 or Passport 12 bedside monitor: a.
® Quick Reference Guide LOCK/UNLOCKING TOUCHSCREEN DISPLAY The Passport 8 or Passport 12 bedside monitor provides the ability to lock and unlock the touchscreen display. To lock the touchscreen display on the Passport 8 or Passport 12 bedside monitor press and hold Main Menu (quick key) for 2-3 seconds until appears.
Passport ® Quick Reference Guide CLEANING/DISINFECTING Cleaning the Main Unit/Module Clean your equipment on a regular basis. Before cleaning, consult your hospital's regulations for cleaning the equipment. The following table lists approved cleaning and disinfecting agents. Product Name Product Type Ingredients Sodium hypochlorite Liquid Sodium hypochlorite bleach 0.5%...
Quick Reference Guide Cleaning and Disinfecting the Accessories For information about cleaning and disinfecting of accessories, please refer to the manufacturer's packaging or the Passport 8/Passport 12 Patient Monitor Operator's Manual. STERILIZATION Sterilization is not recommended for this equipment, related products, accessories, or supplies unless otherwise indicated in the Instructions for Use that accompany the products, accessories or supplies.