Warnings and Guidelines ......................7 - 4 Test Equipment and Special Tools Required ..................7 - 4 Inputs............................7 - 4 Diagnostics and Calibration Procedure ..................7 - 5 NIBP Diagnostics ........................7 - 5 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 5
Sterilization and Cleaning of Reusable Cuffs ................8 - 4 Battery Replacement and Maintenance ..................8 - 5 Recorder Paper Replacement....................8 - 5 Care and Storage of Thermal Chart Paper ................8 - 5 Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420...
Page 6
Table of Contents This page intentionally left blank. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420...
This Service Manual is intended as a guide for technically qualified personnel during repair and calibration procedures for the Passport 5-Lead, 5L, LT and XG monitors. The information has been divided into the eight main sections as listed in the table of contents. A detailed table of contents of each section is also provided on the first page of each section.
Page 8
WARNING: Many patients suffer from poor peripheral perfusion due to hypothermia, hypovolemia, severe vasoconstriction, reduced cardiac output, etc. These symptoms may cause a loss in vital sign readings. 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 9
Vacuum (negative pressure) should not exceed 1 mmHg at the Gas Module Pump Exhaust fitting (50). Excessive scavenge vacuum may result in damage to the Gas Module’s internal pump. 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 10
Testing should include verification that the maximum delay for cardioversion synchronization does not exceed 60 ms (monitor and defibrillator). Cable should be labeled with the devices they have been tested for use with. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 viii...
Page 11
ETO sterilization may also be used. CAUTION: When ironing or pressing the cuffs, be aware that the Velcro® fasteners can melt at temperatures above 325°F, 162°C. 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 12
Introduction Warnings, Precautions and Notes This page intentionally left blank. 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
The pictures of the unit and the displays should be used as samples and may not be exactly as your unit appears. Table 1-1 on page 1-2 lists the functions available on the Passport 5-Lead/5L and the Passport LT. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 1...
Page 14
Sections 1.2 and 1.3 are included as a review of instrument functions and operation, although the reader is encouraged to refer to the Operating Instructions, P/N 0070-00-0324, for more complete details. 1 - 2 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
As the cursor is moved, the highlighted menu window is displayed. 2. SELECT (Set-Up) A repeat action key used to select the function or value indicated by the highlighted cursor. 1 - 4 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Choices are: Off, Continuous, 1, 2.5, 5, 10, 15, 20, 30, 60, and 120 minutes. Five seconds after the selection is made and the key is released, the new interval takes effect. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 5...
Page 18
Pressing and holding the key for 3 seconds clears all trended data. 17. TIME UP p and q TIME DOWN (Trends) Two repeat action keys which are used to scroll through the trend data. 1 - 6 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 19
A green LED used to indicate that the POWER Switch (33) is in the ON position. ECG AREA MENU PARAMETER AREA AREA MULTI-FUNCTION AREA MESSAGE AREA FIGURE 1-2 Display Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 7...
Page 20
C. PARAMETER Display Area D. MULTI-FUNCTION Display Area MESSAGE Display Area NOTE: Only one key function will be recognized at any time. The Passport will ignore multiple key selections. 1 - 8 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
All Patie nt Connections Ele ct rica lly Is olated IBP 1 PI 1 IBP 2 PI 2 BATT. CO 2 FIGURE 1-4 Left Side Panel with Nellcor® SpO Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 9...
Page 22
A six-pin male connector (Datascope P/N 0131-00-0094) used for Datascope specified pressure transducers listed in Chapter 5, Accessories. 30. CO (Optional) A 20 pin connector used to attach the CAPNOSTAT C sensor to the monitor. 1 - 10 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Monitor, Remote Color Display, Nurse Call, DPD Defibrillator or other peripheral devices. 34. POWER A recessed rocker switch which interrupts power to the main unit but does not prevent charging of the batteries. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 11...
For instructions on mounting the remote display to a wall, refer to Operating Instructions. WARNING: The use of remote displays will enlarge all waveforms and may alter aspect ratios depending on the remote device and its set-up. 1 - 12 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
2. Attach sensor to Sp connector and apply to the patient. 3. Set either waveform 2 or 3 in the set-up menu to display the Sp waveform, if desired. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 13...
PROGRESS” message will display and once the self tests have been completed a “SELF TEST COMPLETE” message will display. After this the main screen is displayed. If any failures occur, see “Status Messages” on page 1-57 for further instructions. 1 - 14 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
ECG Display Area - displays the ECG trace, ECG information and pacer enhancement status. PARAMETER Display Area - displays the current values of patient parameters. 1 - 16 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 29
The “IBP”, “Recorder” and “CO ” menu items only appear only on models equipped with these options. NOTE: The display sample shown above may not be an exact representation of your unit. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 17...
Page 31
The ECG Display Area contains the ECG waveform, the ECG size, scale, lead information, Pacer Enhancement ON/OFF message, and when appropriate a message indicating there is no ECG waveform available. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 19...
Page 32
3-Lead cables. The “Artifact” message is also displayed in the Heart Rate window. 1.3.3.3 Multi-function Display Area FIGURE 1-10 Multi-Function Display Area 1 - 20 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 33
The trend lists display consists of a heading that includes the page #, the time, the parameters with their units, and the listed data in columns under the appropriate heading. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 21...
Page 34
After 1 hour the trend data will be cleared. NOTE: The display sample shown above may not be an exact representation of your unit. 1 - 22 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
1. Using the UP p and DOWN q arrow keys (1), move the cursor to select the desired main menu item. NOTE: As the cursor is moved up and down the list of menu items, view windows for each menu item are displayed. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 23...
Page 36
1 mV FIGURE 1-13 4. Press the END (3) key to return to the normal monitoring mode when all of the sub-menu items have been set as desired. FIGURE 1-14 1 - 24 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Any cuffs that exhibit this effect should be replaced. NOTE: Ensure that the pressure tubes are not compressed or restricted. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 25...
Page 38
NIBP. The HR source menu selection must be in the Auto mode (i.e., not selected for ECG, IBP or SpO ) and no heart rate alarm limits are set. (see “Alarms” on page 1-49). 1 - 26 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 39
After the first measurement in the timer mode, the inflation pressure is the previous systolic +50 mmHg in the Adult Mode and +30 mmHg in the pediatric and neonate mode. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 27...
Page 40
Only one measurement is taken for each measurement cycle - even if the unscheduled measurement coincides with the scheduled measurement. 1 - 28 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
• Attach electrodes to lead wire first before placing onto patient. NOTE: Using more than one type of electrode on the same patient should be avoided because of variations in electrical resistance. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 29...
Page 42
However, there may be excessive noise on these waveforms which can corrupt the heart rate calculation. See Section 5 in the Operating Instructions for a list of the proper ECG cables. 1 - 30 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 43
• Place green electrode on lower right abdomen under the ribs. • Place the brown lead in any one of the V lead (V1 - V6) positions, see Figure 1-17 on page 33. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 31...
Page 44
• V3-Midway between V2 and V4 on a line joining these 2 locations. • V4-Fifth interspace in mid-clavicular line. • V5-Fifth interspace in anterior axillary line. • V6-Fifth interspace in mid-axillary line. 1 - 32 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 45
The thoracic impedance is measured between the RA and LA electrodes therefore, the electrodes must be placed across the chest from each other to optimize measurement of chest movement. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 33...
Connector (25) on the left side panel. 3. Apply sensor and connect to cable. NOTE: Do not place the sensor on an extremity with an invasive catheter or blood pressure cuff in place. 1 - 34 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 47
• pulse levels • rejection of ambient light • Can be re-sterilized (ETO sterilization - 3 times) • patient isolation • ease of application and removal • long term patient comfort Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 35...
Page 48
Patient Connector (25). Confirm that the cable connector is securely in place. NOTE: To obtain maximum cable use, do not twist the cable connector when attaching to or disconnecting from the Passport Monitor. 1 - 36 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 49
If any of the above mentioned indications occur, immediately remove the sensor and find an alternate site. NOTE: Check the sensor site more frequently on infant and active patients. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 37...
Page 50
Non-adhesive bandages are recommended for premature infants to minimize prenatal skin damage. See Accessories, Chapter 5, for more detailed information *** Additional choices: 0060-00-0026-02 (10’ sensor cable) 0020-00-0071-01 (3’ sensor cable plus 7’ extension cable) 1 - 38 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
SpO readings will result. Verification of oxygenation should be made, especially in preterm infants and patients with chronic lung disease, before instituting any therapy or intervention. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 39...
Page 52
Light sources that can affect performance include surgical lights, especially those with a xenon light source, bilirubin lamps, fluorescent lights, infrared heating lamps, and direct sunlight. 1 - 40 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 53
The instrument’s software reads this calibration resistor to determine the appropriate calibration coefficients for the measurements obtained by that sensor. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 41...
Choke blocks are electrical filters that may be used in electro-cautery environments where ECG interference can be substantial. These filters remove the electro-cautery noise, but also block the signal used by the Passport Monitor to measure respiration. 1 - 42 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
The Passport offers the Capnostat mainstream method of measuring respiratory rate and end tidal CO . Capnostat requires a hardware option within the Passport Monitor. It utilizes a large lumen adapter which is placed directly into the endotracheal tube. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 43...
Page 56
(See “Use of Menus” on page 23.) NOTE: If the monitor detects changing CO levels (breaths) during an adapter calibration, an “Adapter Cal Failure” message displays. Remove the source of CO and repeat the calibration. 1 - 44 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 57
To prevent “rain-out” and moisture from draining into the airway adapter, DO NOT place the airway adapter in a gravity dependent position. CAPNOSTAT CO Sensor Ventilator Elbow Cable Clip Adult Airway Adapter Patient FIGURE 1-20 Adult Airway Adapter Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 45...
Page 58
6. The sensor cable should face away from the patient. To secure the sensor cable safely out of the way, attach the sensor cable holding clips to the airway tubing, then connect the sensor cable to the clips. 1 - 46 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 59
6. The sensor cable should face away from the patient. To secure the sensor cable safely out of the way, attach the sensor cable holding clips to the airway tubing, then connect the sensor cable to the clips. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 47...
Page 60
However, if a second sensor is connected in place of the original, the second sensor must be calibrated and if the original sensor is used again, it too will have to be recalibrated. 1 - 48 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
The Datascope Passport Monitor provides high and low alarm limits for heart rate (HR), systolic pressure (BP1/NIBP Sys), diastolic pressure (BP1/NIBP Dia), mean pressure (BP2 Mean), respiration rate, ETCO , and SpO . An alarm for apnea delay is also provided. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 49...
Page 62
Respiration rate measurement range is limited to 2 - 150 rpm when Mainstream CO is selected as the rate Symbol source. Values above 150 rpm will be displayed as 150 rpm.. 1 - 50 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 63
• The violated parameter is displayed in reverse graphics in the parameter window. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 51...
Page 64
Approximately 15 minutes of operating time remains when the low battery symbol displays. Battery recharge time is 16 hours. NOTE: Recorder is not operational when the battery charge is low. 1 - 52 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
It is a two trace thermal strip chart recorder with an integral paper spool. The recorder uses plain white thermal paper 5 cm wide (see Section 4.7 for replacement instructions). All grid patterns and data are printed by the recorder. 1 - 54 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 67
• If the RECORD key (19) is pressed while List Trend is displayed (and the ECG waveform is not frozen), the recorder prints the list trend report. • See Section 4.7, in the Operating Instructions, of paper installation. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 55...
Page 68
The list trend data is printed with text running along the length of the strip. If more than one page of data is available then all additional pages are printed along the length of the strip. 1 - 56 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
The monitor uses the Message Display Area to provide messages to the user relating to monitor status. The following lists these messages and a description of the message. The messages are grouped by function. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 57...
Page 70
• Always have an alternate method of BP verification available. • On vasoconstricted patients, failure to evacuate air from the cuff can distort BP measurement. • Do not place cuff on extremity that has an IV. 1 - 58 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 71
Hardware settings are being adjusted in order to Change to site where SEARCH discriminate a pulse waveform. pulse is stronger if patient is vasoconstricted. Change or readjust sensor if loose. No detectable pulse is measured. PULSE Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 59...
Page 72
: CHECK SENSOR The CO sensor is out of calibration or A zero calibration damaged. should be performed if the message persists. This is accessed through the CO menu. 1 - 60 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 73
: FAILURE The CO subsystem has failed. Cycle the power on the monitor to determine if the problem persists. If the message returns then the unit needs to be serviced. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 61...
-Ensure proper connection. secured.-Electrodes dry. (Electrode to lead, lead to -Cable or lead wires damaged. cable, cable to monitor.) -Re-prep skin and apply fresh moist electrodes. -Check with continuity tester. 1 - 62 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 75
-Change scale via menu. Too Large Resp. Waveform -Patient breathing shallow or turned on -Change lead position to better Too Small side. detect respirations. -Scale set inappropriately. -Change scale. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 63...
For each menu item highlighted, the Choices bar indicates the available selections. The menu Instructions bar indicates the available actions to take. See “Use of Menus” on page 23 for detailed instructions on menu operation. 1 - 64 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 81
The range for pediatrics is 60-180 mmHg, and the range for neonates is 40-120 mmHg. CHANGE NIBP mmHg Start Pressure: Range: 100-260 mmHg = Adjust valve EXIT = quit SELECT = Enter/move FIGURE 1-34 NIBP Menu Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 69...
Page 82
CHANGE HR AUTO Source: Choices: Auto, ECG, IBP1, SpO2 = Adjust valve END = quit SELECT = Enter/move FIGURE 1-37 HR/ECG Menu 1 - 70 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 83
O Compensation - Select ON to compensate for interference caused by nitrous oxide present in the gas being analyzed. Compensation - Select ON to compensate for interference caused by oxygen present in the gas being analyzed. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 71...
Page 84
Day: Hr SpO2 size Alarm Audio delay = Adjust valve END = quit Aud alm standby SELECT = Enter/move Serial output type NIBP: Idle SPO2 FIGURE 1-40 Date Menu 1 - 72 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 85
= A djust valve E XIT = quit CO 2 S ELE CT = E nter/m ove Co lor Setti ngs NIB P : Idle S PO 2 FIGURE 1-42 Trend Menu Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 73...
Page 86
Alarm Audio delay Choices: OFF, 4,6,8 sec Aud alm standby = Adjust valve END = quit Serial output type SELECT = Enter/move NIBP: Idle SPO2 FIGURE 1-45 Audio Alarm delay Menu 1 - 74 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 87
Alarm Audio delay Choices: Visa,Accutorr,Msg,Diap Aud alm standby Serial output type = Adjust valve END = quit SELECT = Enter/move NIBP: Idle SPO2 FIGURE 1-47 Serial Output Type Menu Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 75...
Page 88
Alarm Audio delay CO2 Units: Aud alm standby Choices: Torr, % Serial output type = Adjust valve END = quit SELECT = Enter/move NIBP: Idle SPO2 FIGURE 1-48 CO Menu 1 - 76 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
3. EXIT (Set-Up) A single action key which causes the display to return to the main screen display (as specified in the DISPLAY section). This key is always available. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 79...
Page 92
A single-action key which mutes the alarms of the corresponding channel on the VISA Central Station. The amount of time the alarms are silenced, is determined by the VISA. 1 - 80 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 93
A delayed action key which zeros the current pressure in the BP2 channel. If the transducer zero process is unsuccessful, “UNABLE TO ZERO” is displayed in the IBP2 data window. This window is only available when a pressure transducer is present. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 81...
Page 94
PUMP (Optional) A single action key which turns on the sidestream CO pump and initiates an adapter and pump calibration. Holding this key for 3 seconds turns the pump off. 1 - 82 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 95
Press SELECT to change the alarm values. Press the LIMITS key again to access additional alarm parameters (when the Gas Module is installed). To return the display to the main screen, press the EXIT key. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 83...
Page 96
When “Aud alm standby” is set to “ON and active on power up,” the unit will power up in the Aud alm standby mode. All audio alarms will be muted. 1 - 84 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 97
When a remote (VISA) recording is printed, the information that is printed is determined by the set-up in the VISA. 23. BATTERY CHARGING LED A green LED used to indicate that the batteries are charging. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 85...
Page 98
They are: A. MENU Display Area B. ECG Display Area C. PARAMETER Display Area D. MULTI-FUNCTION Display Area MESSAGE Display Area See Section 1.6.3 on page 1-102 for details. 1 - 86 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Press to Open CUFF ECG/EKG E.S.I.S. All Patient Connections Electrically Isolated IBP 1 PI 1 IBP 2 PI 2 PUMP INPUT CO 2 FIGURE 1-52 Left Side Panel with Nellcor® SpO Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 87...
Page 100
5, Accessories of the Passport Operating Instructions Manual. 32. IBP2 (Optional) A six-pin male connector used for Datascope specified pressure transducers listed in Chapter 5, Accessories of the Passport Operating Instructions Manual. 1 - 88 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 101
33. PUMP INPUT (Optional) This is used to connect the sampling tubing for Sidestream CO 34. CO (Optional) A 20 pin connector used to attach the CAPNOSTAT® CO sensor to the monitor. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 89...
A recessed rocker switch which interrupts power to the main unit but does not prevent charging of the batteries. 39. PUMP EXHAUST (Optional) This panel mount coupling is used for attaching a gas scavenging system. 1 - 90 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Passport CO filter (P/N 0103-00-0452) is located inside this housing. This filter should be replaced at least once a year. This filter is not part of the optional Gas Module. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 91...
0070-00-0397, 0070-00-0440, and 0070-00-0503 of the Passport Operating Instructions Manual. WARNING: The use of remote displays will enlarge all waveforms and may alter aspect ratios depending on the remote device and its set-up. 1 - 92 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Refer to section 4.11 of the Passport Operating Instructions for more details. 44. Dust Filter Cover The Dust Filter Cover is removed to access the filter. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 93...
Page 106
A switch used to power the unit on and off. It is located on the front of the Gas Module SE and on the back of the Gas Module II. 1 - 94 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
This port is used only to measure room air. This port is not to be connected to anything. Do not block this port. 52. External Interface Port A communication interface port used to connect the Gas Module to the Passport XG. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 95...
Page 108
55. Gas Module II Mounting Bracket This bracket (P/N 0406-00-0729-01) mounts the Gas Module and Passport XG together. The Passport XG Power Pack can be mounted to the rear of this bracket as well. 1 - 96 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
4. To enable automatic measurements, press INTERVAL key to set desired period. The NIBP measurement will start when the INTERVAL time has expired. 5. Press DEFLATE to suspend measurement. This also stops interval measurements. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 97...
Page 110
3. Empty and clean water trap if more than half full or whenever changing Patients. 4. Set Waveform 2 or 3 to CO , Agent or O or Auto Display CO to Waveform 2 or 5. Set the Resp. Source to CO /ECG. 1 - 98 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Full Disclosure* Arrhythmia* This item only appears if the serial output type is set to Visa or Visa w/ Admit. These items only appear if the related function is installed. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 99...
Page 112
150 mmHg This item only appears if the serial output type is set to Visa or Visa w/ Admit. These items only appear if the related function is installed. 1 - 100 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 113
Beep Alarm This item only appears if the serial output type is set to Visa or Visa w/ Admit. These items only appear if the related function is installed. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 101...
Fahrenheit is the default setting for English units only. All other language units default to Centigrade. 1.6.3 Display ECG AREA MENU PARAMETER AREA AREA MULTI-FUNCTION AREA MESSAGE AREA FIGURE 1-59 Window Layout 1 - 102 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 115
For a graphic representation of each of these menus see “Use of Menus” on page 1-109. NOTE: The “IBP”, “Recorder”, “CO ” and “Gas Module” (Gases) menu items only appear on models equipped with these options. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 103...
Page 116
These items only appear if the related function is installed. This item only appears if the Gas Module option is installed and serial output is set to Gas Module. 1 - 104 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 117
• ECG WAVEFORM - The display shows the ECG waveform at a user selected speed. This provides 4 seconds of data on the display (at 25mm/sec). The vertical scale of the waveform is determined by the ECG size selected. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 105...
Page 118
The scale information for invasive pressures consists of a 0 at the bottom of each window to represent zero pressure level and the scale range value at the top of the window. The waveform provides four seconds of data when the selected speed is 25mm/sec. 1 - 106 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 119
To clear the trend data, press and hold the DISPLAY key (19) for 3 seconds. NOTE: Trend data will be stored in memory for 1 hour after the unit has been powered down. After 1 hour the trend data will be cleared. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 107...
Page 120
This area of the display contains messages relating to NIBP, SpO , CO , Gas Module and recorder operation. See “Status Messages” on page 1-158 for a complete list of these messages. 1 - 108 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
If no change is required to the highlighted item, press SELECT to move the cursor down to the next item on the list. Keep pressing SELECT until the required item to change is highlighted. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 109...
25 - 35 Adult - Tan 0998-00-0003-24 18 - 26 Child - Red 0998-00-0003-23 10 - 19 Infant - Green 0998-00-0003-22 6 - 11 New Born - Blue 0998-00-0003-21 1 - 110 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 123
NOTE: The NIBP cuff should not be placed on a limb that is being utilized for any other medical procedure. For example, an I.V. catheter or an SpO sensor. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 111...
Page 124
Make a valid NIBP measurement. During or after an NIBP measurement, one of several advisory messages may be displayed. Refer to “Status Messages” on page 1-158, for their explanations. 1 - 112 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 125
As a safety precaution, a five minute limit is placed on continuous measurements. After 5 minutes, the NIBP interval will automatically switch to measurements taken every 5 minutes. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 113...
Page 126
“RETRY” or “UNABLE TO MEASURE” message will display in the NIBP message window. 1.6.5.5 Start and Deflate Functions The START and DEFLATE functions have the following effects on the timed measurement sequence. 1 - 114 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
• Attach electrodes to lead wire first before placing onto patient. NOTE: Using more than one type of electrode on the same patient should be avoided because of variations in electrical resistance. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 115...
Page 128
See Section 5 in the Datascope P/N’s 0070-00- 397, 0070-00-0440, 0070-00-0503 of the Passport Operating Instructions for a list of the proper ECG cables. 1.6.6.1 Adult Electrode Placement For 5 Lead ECG Monitoring 1 - 116 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 129
• Place black electrode on right sternal under clavicle. border, 4th intercostal space. • Place red electrode on lower left • Place red electrode on midaxillary line, abdomen under the sixth rib. 5th intercostal space. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 117...
1.6.7 Invasive Pressure Acquisition 1. Plug the pressure transducer into the PRESSURE TRANSDUCER connector (31) or (32) on the side panel. 1 - 118 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Connector (28) on the left side panel. 3. Apply sensor and connect to cable. NOTE: Do not place the sensor on an extremity with an invasive catheter or blood pressure cuff in place. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 119...
Page 132
The detector receives that amount of light not absorbed by the blood or tissue components. The Passport then uses the relative absorption of the two light wavelengths to compute and display SpO and Rate measurements. 1 - 120 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 133
C. Sensor Connection to the Passport Monitor 1. Align the cable connector on the sensor assembly with the SpO Patient Connector (28) on the Passport XG Monitor. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 121...
Page 134
If any of the above mentioned indications occur, immediately remove the sensor and find an alternate site. NOTE: Check the sensor site more frequently on infant and active patients. 1 - 122 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 135
Non-adhesive bandages are recommended for premature infants to minimize prenatal skin damage. See Accessories, Chapter 5, for more detailed information. *** Additional choices: 0060-00-0026-02 (10’ sensor cable) 0020-00-0071-01 (3’ sensor cable plus 7’ extension cable) Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 123...
Carefully read the sensor directions for use, the Passport operating instructions, and all precautionary information before use. WARNING: Excessive ambient light may cause inaccurate measurements. Cover the sensor site with opaque material. 1 - 124 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 137
“ON”, the Passport will give a one time series of 5 triple beeps. Set to “CONT”, the Passport will give a series of 5 triple beeps every 30 seconds. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 125...
Page 138
Interference” message to be displayed. Encourage the patient to remain still. If this is not possible try moving the sensor to another site. If patient movement presents a problem: 1 - 126 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
See “Use of Menus” on page 1-109. 1.6.10 Sequence for Establishing SpO with Nellcor® Pulse Oximetry 1. Select the appropriate sensor for the patient see table on page 1-51. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 127...
Page 140
Set the Auto Display for IBP1 and CO to either waveform 3 or Off. If either of these functions are set to waveform 2, they will have a higher priority to display, if they are detected. 1 - 128 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 141
NOTE: Nellcor®, OXIBAND® and DURASENSOR® are registered trademarks of Nellcor® Incorporated. OXISENSOR™ and DURAFORM™ are trademarks of Nellcor Incorporated. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 129...
Page 142
• Move the sensor to a less active site. • Use a type of sensor that tolerates some patient motion, such as the OXISENSOR™ D-25, D-20, N-25, or I20 oxygen transducer. 1 - 130 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 143
(default setting), 2, and 3. 1. The default operating mode, uses a 5 to 7 second averaging time and is useful in situations in which the patient is relatively inactive. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 131...
If the patient’s airway is obstructed and the patient attempts to breath, then the chest wall can move and create a respiratory signal even though no gas flow is occurring to the patient. 1 - 132 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 145
Refer to Section 1.6.12 on page 1-134 through Section 1.6.14 on page 1-146 for more details on the above options for CO monitoring. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 133...
Up” message is displayed. NOTE: The CO waveform displays while the sensor is warming up, however numeric data will not be available until the “Sensor Warming Up” message disappears. 1 - 134 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 147
(breaths) during an adapter calibration, a “Breaths Detected ... Retry” message displays, and then an “Adapter Cal Failure” message will display. Remove the source of CO and repeat the calibration. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 135...
Page 149
Position the airway adapter with it’s windows in a vertical and NOT a horizontal position. This helps keep patient secretions from “pooling” on the windows. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 137...
Page 150
“Sensor Warming Up” message must not be displayed. To perform a Capnostat® sensor calibration: 1. Verify the Passport XG is turned on and the Capnostat® is plugged in and warmed-up. 1 - 138 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
6 in Section Section 1.6.13.1 on page 1-140. NOTE: For sidestream operation, always set the apnea and CO alarms. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 139...
Page 153
Passport XG CO connector (34). The “CO Sensor Warming Up” message is displayed. The CO numeric dataand waveform will not be available until the “CO : Sensor Warming Up” message disappears. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 141...
Page 154
1-164 for additional CO messages. NOTE: This pump calibration also calibrates the adapter. It is not necessary to perform a separate adapter calibration at this time. 1 - 142 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 155
CO capability, there is automatic compensation for barometric pressure. NOTE: See “Sidestream CO Messages (on Passports with Sidestream CO only)” on page 1-164 for more details on messages. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 143...
Page 156
(breaths) during an adapter calibration, a “Breaths Detected ... Retry” message displays, and then an “Adapter Cal Failure” message will display. Remove the source of CO and repeat the calibration. 1 - 144 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 157
However, if a second sensor is connected in place of the original, the second sensor must be calibrated and if the original sensor is used again, it too will have to be recalibrated. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 145...
Vacuum (negative pressure) should not exceed 1 mmHg at the Gas Module Pump Exhaust fitting (50). Excessive scavenge vacuum may result in damage to the Gas Module’s internal pump. 1 - 146 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 159
The pressure indicator on the Datascope gas regulator must operate in the green zone during the entire calibration session. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 147...
Page 160
NOTE: To avoid premature emptying of the gas canister, always remove the regulator at the end of the procedure, prior to storage. 1 - 148 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
AL AR MS Depress the Alarms VOL UME Limits Key to access the Alarms Menu LIMIT S MUT E FIGURE 1-83 Alarm Menu Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 149...
Page 162
This Alarm Limit is found on page one of the Alarm Limit Menus, since it applies whether the Gas Module is configured or not. However, when the Gas Module is configured this Alarm parameter will have different ranges and it will be saved separately. 1 - 150 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 163
• The violated parameter is displayed in reverse graphics in the parameter window. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 151...
Page 164
“Aud Alm Sby” message is displayed above the menu selections. When in this mode, press MUTE to re-activate alarms. 1 - 152 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
The User Configuration Mode is accessed via a special power-up sequence. To enter the User Configuration Mode: 1. Turn the POWER switch (34) ON. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 153...
Page 166
The trend interval is independent of the NIBP interval. 1 - 154 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
(22) is pressed. If a continuous printing is required, press and hold PRINT (22) for 3 seconds (until a beep is heard). Press PRINT again to stop a real time printing. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 155...
Page 168
The upper and lower borders have the date, time and physiologic parameters currently available as well as the ECG size and lead configuration the recorder is printing. All parameters include their units. 1 - 156 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 169
The list trend data is printed with text running along the length of the strip. If more than one page of data is available then all additional pages are printed along the length of the strip. FIGURE 1-87 Sample Printout, Trend List Format Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 157...
The monitor uses the Message Display Area to provide messages to the user relating to monitor status. The following lists these messages and a description of the message. The messages are grouped by function. 1 - 158 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 171
• Always have an alternate method of BP verification available. • On vasoconstricted patients, failure to evacuate air from the cuff can distort BP Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 159...
Page 172
Change to site where to discriminate a pulse waveform. pulse is stronger if patient is vasoconstricted. Change or readjust sensor if loose. NO PULSE No detectable pulse is measured. 1 - 160 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 173
SpO measurement. • Intra-vascular dyes, depending on concentration, may affect SpO measurements. • SpO measurements may be difficult on patients undergoing IABP treatments. This monitor measures functional hemoglobin. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 161...
Page 174
: ZERO CAL IN The CO sensor is placed on the zero Wait for the zero PROGRESS cell and a CO calibration is in calibration to complete. progress. 1 - 162 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 175
: FAILURE The CO subsystem has failed. Cycle the power on the monitor to determine if the problem persists. If the message returns then the unit needs to be serviced. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 163...
Page 176
Check exhaust line for blockage and clear if possible. If message persists call for service. 1 - 164 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 177
Isoflurane as the primary agent and administered with the the manually selected agent is neither Agent selected, or select Isoflurane nor the detected secondary agent. Agent Auto ID. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 165...
Page 178
GM: ET AGENT Appears when the End Tidal Anesthetic Agent Check patient. HIGH measurement is greater than or equal to the value set for the ET Agent High Alarm. 1 - 166 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 179
ERROR persists, call for service. CALIBRATION Appears when the Gas Module cannot Repeat calibration ERROR ZEROING perform a Zero during calibration. procedure. If problem ERROR persists, call for service. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 167...
-Press ECG SIZE key. Excessive Electro- -Inadequate skin prep prior to application of -Repeat skin prep and surgical electrode. electrode placement Interference procedures. -Add additional gel to electrodes. 1 - 168 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 181
-Readjust as required. With No key. -Check insertion. Waveform -Lead wires and patient cable not fully inserted into -Check with lead proper receptacle. continuity tester. -Cable or lead wires damaged. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 169...
Page 182
-Poor skin prep. -Clean and abrade skin before applying electrodes. “CVA” Message -Can be caused by shallow breathing or an apnea -Check the patient- event. adjust scales or leads if necessary. 1 - 170 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 183
-+12/-23 Volt Board Failure-CPU Board Failure Board. Failure Interface Board Failure (optional) -Check External Supply Output voltage. -Replace F1 fuse on +12/-23 Volt Board. -Replace CPU Board. -Replace CO Interface Board Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 171...
VISA operation with the Passport XG. The keys that are used on the Passport XG in conjunction with the VISA are: ADMIT (4), DISCHARGE (5), SILENCE (6) and PRINT (22). 1 - 172 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 185
30 seconds until the admit is confirmed on the VISA. To confirm the admit on the VISA either click on the yellow tile with the mouse or tap on the tile if the VISA has a touch screen. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 173...
Page 186
Press the SILENCE key (6) to mute alarms that are initiated from the Passport and sent to the VISA Central Station. The amount of time the alarms are silenced, is determined through the VISA. 1 - 174 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Arrhythmia: If ON is desired, it must be selected prior to selecting Admit “YES”. Arrhythmia “ON” requests Arrhythmia detection to be enabled at the VISA. Pacer Enhancement: Set to ON to display the pacer signal as a full scale square wave. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 175...
Page 188
Passport XG is powered on. Select “Restore Factory” to return to the factory default settings. NOTE: The choices for Waveform 2 and Waveform 3 only include the Agent and O selections when the Gas Module is installed and configured. 1 - 176 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 189
Gas Module selections when these items are installed and configured. NOTE: Record Destination is only displayed when Serial Output Type is set to “VISA” or “VISA WITH ADMIT”. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 177...
Page 190
The range shown is for the adult patient size. The range for pediatrics is 60-180 mmHg, and the range for neonates is 40-120 mmHg. CHANGE NIBP mmHg Start Pressure: Range: 100-260 mmHg = Adjust valve EXIT = quit SELECT = Enter/move FIGURE 1-96 NIBP Menu 1 - 178 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 191
, if CO is present. If CO is not present, the Resp Source is set to ECG. If option is not installed, Resp Source choices are OFF or ECG. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 179...
Page 193
Start Zero Calibration: Select Yes to initiate the calibration of the sensor. Start adapter cal: Select Yes to initiate the calibration of the airway adapter with the sensor. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 181...
Page 194
Resp Menu (see “Resp Menu” on page 1-179). Start Zeroing: Select Yes to manually initiate re-zeroing of the Agent, CO , and N 0 Gas Module parameters. 1 - 182 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 195
Choices: 3.125, 6.25, 12.5, 25 mm/sec Date of last mixture calibration: 09/15/97 Date of last mixture calibration: 09/15/97 = Adjust SELECT=enter/move EXIT = Adjust SELECT=enter/move EXIT FIGURE 1-102 Gas Module Menu Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 183...
Page 196
When the Gas Module is installed and configured, two more pages with additional alarm limits become available, refer to “Alarm Limits” on page 1-150 for complete details on limits. 1 - 184 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 198
C o lo r S e ttin g s G a s M o n ito r N IB P : Id le S P O 2 FIGURE 1-107 Trend Menu 1 - 186 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 199
Choices: OFF, 4,6,8 sec Serial Output Type = Adjust valve EXIT = quit SELECT = Enter/move Color Settings Gas Monitor NIBP: Idle SPO2 FIGURE 1-110 Alarm Audio Delay Menu Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 187...
Page 200
Passport XG Monitor. The baud rate for the Accutorr protocol can be set for 1200, 2400, 4800 or 9600 bps. NOTE: The VISA Central Station cannot be used in conjunction with the Gas Module option. 1 - 188 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 201
S E L E C T = E n t e r / m o v e G a s M o n i t o r N I B P : Id le S P O 2 FIGURE 1-113 CO Menu Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 1 - 189...
Page 202
Operation (XG) Operation (5-Lead, 5L, LT and XG) This page intentionally left blank. 1 - 190 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 205
Passport 5 Lead/5L - All Options Available VISA OPTION 0334-00-1197 Passport LT - SpO2, NIBP Optional Only Datascope SpO2 Is Available As An Option On The Passport LT FIGURE 2-1 Passport 5L/LT Block Diagram Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 2 - 3...
Page 206
E.L. W/O IBP 331-00-0064-02 0160-00-0009 BLANK LABEL FOR NO RECORDER OPTION: 334-00-0973 640 X 200 CENTRAL SILENCE LABEL VISA OPTION 0334-00-1197 FIGURE 2-2 Passport 5L with CO Block Diagram 2 - 4 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 207
Theory of Operation Block Diagrams FIGURE 2-3 Passport XG Block Diagram Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 2 - 5...
Page 208
3=17VDC French 4=GND Spanish PLANAR VGA E.L. DISPLAY 0160-00-0026 Italian Keypad P/N - 0330-00-0017-XX LCD Color Display 0160-00-0028 640 x 480 pixels FIGURE 2-4 Password XG, (Masimo) Block Diagram 2 - 6 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
RA, LA, RL, LL, and C, to 200 volts differential to “Shield” potential. “Shield” is ultimately connected to isolated ground on the Front End Module. 2 - 8 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 211
The connection to the non isolated monitor ground is via a direct wire strap from the resistor and spark gap combination non isolated side, to the monitor’s internal center chassis structure. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 2 - 9...
The non-isolated microcontroller is a high-performance 32 bit integrated processor with direct memory access. 2 - 10 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 213
(supplied from the front end board) so that it cannot exceed the +5 volt supply rail of U4. Were this to occur, U4 would malfunction, and lead fault would not be reliably detected. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420...
Page 214
3-lead system, the 5-lead board is identified by grounding pin 21 of the microcontroller, which is high on the 3-lead systems. 2 - 12 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 215
I, II or III, but will see a voltage as little as -2 volts if one of LA, RA or LL is detached in lead C. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420...
Page 216
FIGURE 2-6 Overall Front End Block Diagram ECG Pre-Amp ESIS ESIS Lead Stage 2 Stage 3 Wilson Select Network U2,U3 Lead Buffers SHIELD Drive FIGURE 2-7 ECG Pre-Amp Block Diagram 2 - 14 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 217
Interconnect board. The capacitors behave as a short to the 114 kHz sinewave carrier, but as a very high (greater than 5 megohms at 10 Hz) shunting impedance to the ECG electrodes so as not to attenuate the ECG signal. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 2 - 15...
Page 218
The impedances of capacitors on the Interconnect Board connected to the RRA and RLA wires, along with the clamping of CR32, CR38, and shield in T2, protect U11 from defibrillator discharges. 2 - 16 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 219
The operation of the second channel, P2, is identical to the above description of P1 (with different input and output labels), using U20 and associated components. The P1 and P2 signals are measured by the analog to digital converter, U21. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 2 - 17...
Page 220
Sleeve Temperature Pre Amp Current Summing Voltage Amplifier Amplifier Temp Q4,U9A Programmable Current Source Comparator Ring T flag FIGURE 2-10 Temperature PreAmp Block Diagram 2 - 18 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 221
R2 replaces the probe and R3 serves as the standard. Resulting ratio is R2/R3 = 3539/9428 = 0.37537. Nominally this ratio corresponds to 37.48ÉC. Assuming worst case errors, this number will be 37.5, ±0.1°C. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 2 - 19...
Page 222
DC voltage by components CR30,31,33,36, L1 and L2, and C169 and C171. The isolated DC voltage will vary between approximately 7.1 volts and 9.3 volts, depending on VDC and the secondary load current requirement. 2 - 20 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 223
SPI, which is used to communicate with the A/D converter. Two of the lines, IRXD and ITXD, carry the serial data received from and transmitted to the isolation block. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420...
Page 224
Diode D41 allows ORing of the reset signal originating at the microcontroller with that generated by the ECG subsystem internally. Resistor R249 provides isolation between the comparator and the 2.5 volt supply. 2 - 22 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 225
Each data item is internally identified as to its contents. Packets received by the microcontroller always consist of two bytes, a single command byte followed by a confirmation. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 2 - 23...
Page 226
The A/D converter is of the successive approximation charge redistribution type. It features an internal capacitor ladder network which is charged from the signal source when the multiplexer samples a signal. 2 - 24 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 227
+5 volt rail and the analog ground plane half. The same reference is used excite the pressure transducer bridge. The temperature function makes use of the ratio of two A/D converter readings and is thus independent of the exact reference value. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 2 - 25...
(lungs). Specifically the device, within limits, determines the fraction of hemoglobin molecules carrying oxygen from the lungs to the body cells. Termed % SPO2, this fraction is normally about 97 percent. 2 - 26 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 229
CURRENT PROGRAMABLE SUBTRACTION PROGRAMABLE SAMPLE DEMULTIPLEXER REMULTIPLEXER FILTER U8A GAIN I GAIN II HOLD VOLTAGE 18 HZ PREAMP U10A U11A U11B U11B FILTER U8B FIGURE 2-13 SpO Block Diagram Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 2 - 27...
Page 230
Next signal is amplified by the dc gain stage. Having the ability to apply a different gain to the two components, this block functions as a coarse equalization of the multiplexed signal. 2 - 28 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 231
R/W* buffers to assume a high-impedance state. When a valid ISA bus access cycle is detected it is synchronized using ISA IOCHRDY signal. Because ISA access cycle is synchronized to 63C09 clock, 63C09 always has SRAM access priority over ISA bus cycle. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 2 - 29...
Page 232
INPUT OUTPUT DESTINATION BIT 4 DAC Bit 8 BIT 3 DAC Bit 9 BIT 2 DAC Bit 10 BIT 1 DAC Bit 11 ACG0 AC GAIN ACG1 AC GAIN 2 - 30 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 233
R24. This will turn Q4 “ON” and Q1 “OFF”, cutting off the voltage supply to the sensor. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 2 - 31...
Page 234
Subtraction function is used to get rid of most of the signal’s DC pedestal. The circuit then amplifies the resulting signal according to the gain set with signals ACG0 and ACG1. 2 - 32 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 235
SA16 through SA19 according to timing relationship described later in the document. 63C09 signals are asynchronously set (disabled) when a valid write cycle is detected and reset (enabled) during the read cycle. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 2 - 33...
Page 236
This is accomplished by manipulating gains of C/V, AC, and DC amplification blocks and by controlling Sequencer, DAC, Remultiplexer, and Sample and Hold sections, while monitoring COMPOUT* and INTERF* outputs. Control of the blocks mentioned above is achieved through Control Register. 2 - 34 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
The SpO2 Modules communicate serially at 9600 baud. A microcontroller is used as a “smart” UART. A microcontroller can be programmed to understand either the Masimo protocol or the Nellcor protocol, and can thus buffer the data in an intelligent manner. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 2 - 35...
Page 238
When a write is made to the status port address, the inverted state of D0 is latched in flip-flop F/F 3 inside the EPLD, U4 (Figure 2-14 on page 2-35). 2 - 36 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 239
D0 of the status port. Thus, the ISA host can poll to determine when data is available. A 1K resistor, R2, limits the current when the CO2 module drives the line high. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 2 - 37...
Page 240
RAM operation. During this self test period, lasting several ms, bit 4 of port C, SRST*, is configured as an output and driven low. This asserts reset to the SpO2 Module. The PM/NIM then enters a data echo mode. 2 - 38 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 241
In order to disable this supply voltage, the SCTS* line must be taken to a logic level Low. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 2 - 39...
FRONT PANEL -15V PARALLEL KEYPAD INTERFACE +12V AR-42 RECORDER INTERFACE (40) U18, U19 INTERFACE U12, U13, U14 16-BIT +12V RCRST DATA BUS -23V FIGURE 2-17 CPU Board Block Diagram 2 - 40 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 243
Passport. This jumper must be removed if the Datasette is to be used in a PROM programmer. The CPU module operating at 16.5888Mhz will necessitate the use of 1 wait state to access SRAM and 0 wait states to access FLASH ROM. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 2 - 41...
Page 244
The display controller outputs VSY, HSY, and LD.4 are also connected to the external interface connector , but are not used at this time. 2 - 42 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 245
3 timing. This device requires a 240 nsec. write pulse. This is the worst case timing on chip select 3. With 3 wait states inserted, the pulse with respect to UDS* is 241 nsec., which is barely met. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 2 - 43...
Page 246
In this mode, the volume signal is enabled only at the start of the beep and the volume will be allowed to decay by not updating the sample and hold. 2 - 44 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 247
In writing to the recorder, the processor should check the SYNC* and WRRDY* lines from the recorder to make sure that the recorder is ready to accept data. In reading from the recorder, the lower byte contains readback information from the recorder. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 2 - 45...
Page 248
The receiver is always enabled; the transmitter is enabled by SEREN (port B pin 7). This line should normally be low to disable the transmitter. This interface is connected to SCC2. 2 - 46 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 249
The DAC voltage input can be used to check the operation of the DAC. The ADC diagnostic voltage can be used to check the operation of the ADC. The external analog input signal is currently not used. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 2 - 47...
Page 250
A defib sync output DSYNC (port A bit 11) is used to provide the external defib sync signal. This signal should be normally high. The E trigger obtained from the analog front end should be output to this signal as soon as possible. 2 - 48 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
BAT CHG B 20 SEC DELAY BAT CHG A BAT CHG B Battery A Battery B +15V VOLTAGE SHUT DOWN -15V Battery A Battery B UNIPOINT FIGURE 2-18 Line/Battery Block Diagram Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 2 - 49...
Page 252
When the system is off, diodes CR11 and CR12 allow a charging path from the battery charger to each battery. Resistors R10 and R13 provide a path to ground, to bleed accumulated charge from CR11 and CR12. 2 - 50 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 253
Capacitor C1 and R8 forces Q1 to turn “ON” and Q2 to turn “OFF” during power-up. This is to insure that the hysteresis resistor R4 will not be connected during power up, to establish the proper state for normal operation. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 2 - 51...
Page 254
IC U2. R9 bleeds charge off of capacitors C5 and C6 to shut off the FET when the oscillator is disabled. R14 limits current to the FET gate when the FET is initially turned “ON”. 2 - 52 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 255
This module will accept input voltages ranging from 9V DC to 18V DC and produces outputs of 5V (@1.5A), 15V (@250ma) and -15V (@250ma). 1 GND 15 W POWER MODULE FIGURE 2-22 Power Module Diagram Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 2 - 53...
C17 filters leading edge spikes from the reconstructed current waveform due primarily to discharge of Q1’s drain capacitance. CR3 allows enough voltage backswng to reset T2’s core and yet clamps the voltage so as not to destroy Q1. 2 - 54 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Eight output and four input lines interconnect the 8 by 4 keyboard matrix. Detailed description of the signals and power leads is given in Chapter 3 on page 3. The DC/DC converter generates the power for the EL panel. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 2 - 55...
Page 258
CPU board and the 8 by 4 keyboard matrix. Speaker Interconnections The speaker signals for alarm and beep tone are interconnected between the CPU board and the loudspeaker. 2 - 56 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
A filter is used to assist in handling the pulsatile load component. Due to even more stringent power supply requirements, the Sidestream CO2 control module requires an additional cleaner -12 volt supply (-12V capnostat on the block diagram). Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 2 - 57...
Page 260
Capnostat II (J92) and one is for the Capnostat III (J94). They differ in that the Capnostat II uses a 15 V supply and also has a separate 5V heater supply, while the Capnostat III uses a dual 12V supplies. 2 - 58 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
The BUSWIDTH input is tied low since the eight bit data bus mode is used for external memory access. The EA/ pin is tied low since external EPROM and static RAM are used for program and data memory. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 2 - 59...
Page 262
Reference voltages +5VR and -5VR are generating by amplifying a 2.5V reference generator U12 by a factor of 2 and -2 respectively. The sampling rate of the 80C196 A/D is about 250 Hz. 2 - 60 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 263
J30. Power to the motor is supplied by means of a decoupling network L2/C12 from the host’s +12 volt bus. A common mode coil, L1, provides further attenuation of the motor brush noise. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 2 - 61...
Page 264
R25. This reduces the effective resistance in the grounded resistor in the voltage divider, causing a lower overpressure trip point in the neonatal mode. FIGURE 2-27 Linear Bleed Block Diagram 2 - 62 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
VRAM, while the output video data is clocked directly from the VRAM to the EL panel. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 2 - 63...
Page 266
DT/ and the write control signal WB/ are changed accordingly). The last 5 states are used to write two 4-bit chunks of input video data into the VRAM. 2 - 64 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 267
LED. The function is powered by an external +17 V which improves the internal power management and allows the system only to work when the external power supply is plugged in. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 2 - 65...
Page 269
This was necessary because these signals are generated from high speed CMOS logic devices whose outputs are under damped and ring. This is shown in Figure 2-30 below: FIGURE 2-30 Termination Board Terminated Signals Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 2 - 67...
Page 270
Block Diagrams Theory of Operation This page intentionally left blank. 2 - 68 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
20 second intervals. As per ANSI/AAMI EC13-1983 par. 3.2.2.2. Recovery Time: Time for trace to return to screen from defibrillator mono pulse 3 sec. maximum Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 3 - 1...
Mean error less than 7 mmHg, standard deviation less than 8 mmHg (adult/pediatric) Mean error less than 4 mmHg, standard deviation less than 7 mmHg (neonatal) Heart Rate Meter Range: 30--250 bpm 3 - 2 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 273
Neonate. The software monitor is disabled in the CAL mode. If the hardware over pressure switch is tripped in normal operation then the power must be cycled to reset the switch. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 3 - 3...
Page 274
12 seconds to 95% of final step change of pulse value (rate from 60-120 bpm) • Temperature Scale: Selectable ºC or ºF Range: 15 ºC to 45 ºC 59 ºF to 113 ºF 3 - 4 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 275
NOTE: 6.25 and 3.125 are for Respiration only. Chart Size: 50 mm ± 2 mm • Battery Performance (Passport 5-Lead, 5L & LT) Run Time: 2 hours from full charge of new battery (5-Lead, 5L, & LT) at 25 °C with SpO2, 1 NIBP every 15 minutes and no recording.
Page 276
±2 Torr max additional, 564 to 760 mmHg Reference Cell Reading: 36 to 40 TORR Sensor on Zero Cell: 0 ± 1 TORR Zero Calibration: Less than 30 seconds Adapter Calibration: Less than 30 seconds 3 - 6 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 277
36 to 40 TORR Sensor on Zero Cell: 0 ±1 TORR Zero Calibration: Less than 30 seconds Adapter Calibration: Less than 30 seconds Pump Calibration: Less than 30 seconds Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 3 - 7...
The Passport 5L and XG are designed to comply with the following industry standards: CSA C22-2 No. 125-M 1984 IEC 601-2-30: 1995 UL 544, Third Edition IEC 601-2-34: 1994 IEC 601-1: 1988/EN60601-1:1990 ISO 9919: 1992 IEC 601-2-27: 1994 ISO 9918: 1993 3 - 8 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 279
The Passport 5L, 5L-CE and XG, XG-CE are registered with the following agencies: UL-USA CSA-Canada The Passport 5L-CE and XG-CE comply with the requirements of the medical device directive 93/42/EEC. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 3 - 9...
Page 280
Agency Compliance Specifications This page intentionally left blank. 3 - 10 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
The instrument covers must not be removed by other than technically qualified personnel who have received supplementary instructions regarding maintenance of medical electronic equipment or have has equivalent experience in this area. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 4 - 1...
Datascope offers an exchange program for certain assemblies in the instrument. In many cases, replacement of the complete assembly will result in the most expedient repairs. See “Exchange Program” on page 6-2 for details concerning the exchange program. 4 - 2 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
The E.L. assembly consists of the E.L. display (121) and The E.L. Power Supply Board (121). 1. Place the front housing, with the E.L. display down on a protective surface. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 4 - 3...
Page 284
3. Lift the Panel Board up and out of the front housing. G. Removal of the Passport XG LCD Color Display 1. Place the front housing with the LCD color display down on a protective surface. 4 - 4 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 285
3. Install a new glare screen by removing the protective coating from the adhesive tape already mounted on the glare screen and press in place inside the front housing. Removal of the Board Sub-Assembly (A) Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 4 - 5...
Page 286
3. Remove the four screws at the corner of the SpO2 Board and lift the board from the sub-assembly. NOTE: For removal of the Interconnection Board and Front End Board, note the position of any mylar shields for proper placement during reassembly. 4 - 6 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 287
2. Remove the four screws (26) at the corners of the Front End Board. NOTE: When installing the Front End and Interconnection Boards, make sure all mylar shields are reinstalled. j. Removal of the Front End Daughter Board Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 4 - 7...
Page 288
3. Disconnect P51 from the line battery board (76). 4. Lift the battery can from the rear housing. C. Removal of the Recorder Assembly (23) 1. Open the recorder paper door and remove paper roll, if installed. 4 - 8 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 289
7. Do NOT touch the reflective surfaces inside the backlight holder, or the fluorescent lamp, or any connector contacts with your bare hands. Oil and other contaminants on your hands can damage these surfaces and contacts. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 4 - 9...
Page 290
7. Reassembly of the display into the Passport is the reverse of the disassembly. Figure 1. Bottom Connector to High Voltage Power Supply High Voltage Board New Backlite Assembly Backlite Detail "A" Assembly Frame Assembly Detail "A" Side View 4 - 10 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
(See Figure 4-5 ) 1. Secure the Mounting Plate (P/N 0386-00-0156) to the power supply bracket (P/N 0406-00-0544) using the four #8-32 x .38 flat head screws (P/N 0212-14-0806). Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 4 - 11...
Page 292
2 inches (5 cm) in diameter. 4. Place the power supply into the slots in the back of the holding bracket. 4 - 12 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 293
Repair Information Mounting Instructions FIGURE 4-1 FIGURE 4-2 FIGURE 4-3 Rolling Stand Bedrail Hook Mount Bed Mount 0040-00-0121-02 0040-00-0121-04 Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 4 - 13...
Page 294
Mounting Instructions Repair Information FIGURE 4-4 Passport Mounting FIGURE 4-5 Wall Mount 4 - 14 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
M o unt A r m Pl a te A r m M o u n ti n g Pl a te FIGURE 4-6 Remote Monitor with Wall Mount Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 4 - 15...
Resistor shall be located inside DIN plug and connected between Pin 6 and Pin 8. Resistor shall be mechanically secure and suitably insulated. Bed Clip: Each cable assembly shall include a bed clip, Datascope P/N: 0343-00-0042. 4 - 16 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 297
DESCRIPTION AND LENGTH 0012-00-0620-01 5 Wire, 10 Foot, USA (AHA) 0012-00-0620-02 5 Wire, 20 Foot, USA (AHA) 0012-00-0620-03 5 Wire, 10 Foot, International (IEC) 0012-00-0620-04 5 Wire, 20 Foot, International (IEC) Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 4 - 17...
Page 298
DESCRIPTION AND LENGTH 0012-00-0722-01 5 Wire, 10 Foot, USA (AHA) 0012-00-0722-02 5 Wire, 20 Foot, USA (AHA) 0012-00-0722-03 5 Wire, 10 Foot, International (IEC) 0012-00-0722-04 5 Wire, 20 Foot, International (IEC) 4 - 18 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 299
DESCRIPTION AND LENGTH 0012-00-0723-01 5 Wire, 10 Foot, USA (AHA) 0012-00-0723-02 5 Wire, 20 Foot, USA (AHA) 0012-00-0723-03 5 Wire, 10 Foot, International (IEC) 0012-00-0723-04 5 Wire, 20 Foot, International (IEC) Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 4 - 19...
Page 300
DESCRIPTION AND LENGTH 0012-00-0724-01 5 Wire, 10 Foot, USA (AHA) 0012-00-0724-02 5 Wire, 20 Foot, USA (AHA) 0012-00-0724-03 5 Wire, 10 Foot, International (IEC) 0012-00-0724-04 5 Wire, 20 Foot, International (IEC) 4 - 20 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 301
ECG Cable Assembly HP 3 Lead - 0012-00-0846 Pin 1 ECG Cable Assembly HP 5 Lead - 0012-00-1002-01 3 Lead HP ECG Lead Wired - 0012-00-0847 5 Lead HP ECG Lead Wired - 0012-00-1003-01 Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 4 - 21...
Page 303
Data Link, Shielded, 16 Pin (XG-CE) - 0012-00-1028 (A)* Terminal 1 Designation 10', +/- 0.5' This version cable is exactly the same as the other except it also has a ferrite on it for CE approval. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 4 - 23...
Page 304
0012-00-1031-01 7’ ± .5’ 0012-00-1031-02 50’ ± .5’ This version cable is exactly the same as the other except it also has a ferrite on it for CE approval. 4 - 24 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 305
NOTE: Cable shield must be electrically isolated from P1 connector shield and ferrule. This version cable is exactly the same as the other except it also has a ferrite on it for CE approval. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 4 - 25...
Page 306
1. CABLE SHIELD SHALL NOT CONTACT CONNECTOR SHIELD OR FERRULE. This version cable is exactly the same as the other except it also has a ferrite on it for CE approval. 4 - 26 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 307
8' 1" N.C. N.C. N.C. N.C. 6-16 N.C. N.C. This version cable is exactly the same as the other except it also has a ferrite on it for CE approval. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 4 - 27...
Page 309
Output, Vertical sync Ground Output, Horizontal sync N.C. (No connection) N.C. (No connection) Ground, Red video return Ground Ground, Green video return Ground Ground, Blue video return N.C. (No connection) Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 4 - 29...
Page 310
DESCRIPTION AND LENGTH 0012-00-0990-01 3 Lead, 10 Foot, USA (AHA) 0012-00-0990-02 3 Lead, 20 Foot, USA (AHA) 0012-00-0990-03 3 Lead, 10 Foot, International (IEC) 0012-00-0990-04 3 Lead, 20 Foot, International (IEC) 4 - 30 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 311
DESCRIPTION AND LENGTH 0012-00-0991-01 3 Lead, 10 Foot, USA (AHA) 0012-00-0991-02 3 Lead, 20 Foot, USA (AHA) 0012-00-0991-03 3 Lead, 10 Foot, International (IEC) 0012-00-0991-04 3 Lead, 20 Foot, International (IEC) Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 4 - 31...
Page 312
DESCRIPTION AND LENGTH 0012-00-0992-01 3 Lead, 10 Foot, USA (AHA) 0012-00-0992-02 3 Lead, 20 Foot, USA (AHA) 0012-00-0992-03 3 Lead, 10 Foot, International (IEC) 0012-00-0992-04 3 Lead, 20 Foot, International (IEC) 4 - 32 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 313
DESCRIPTION AND LENGTH 0012-00-0993-01 3 Lead, 10 Foot, USA (AHA) 0012-00-0993-02 3 Lead, 20 Foot, USA (AHA) 0012-00-0993-03 3 Lead, 10 Foot, International (IEC) 0012-00-0993-04 3 Lead, 20 Foot, International (IEC) Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 4 - 33...
Page 314
DESCRIPTION AND LENGTH 0012-00-0994-01 Male to Male, 10 foot, Terminated 0012-00-0994-02 Male to Male, 25 foot, Terminated 0012-00-0994-03 Male to Male, 50 foot, Terminated 0012-00-0994-04 Male to Male, 75 foot, Terminated 4 - 34 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 315
PART NUMBER DESCRIPTION AND LENGTH 0012-00-1010-02 Male to Male, 25 foot, Unterminated 0012-00-1010-03 Male to Male, 50 foot, Unterminated 0012-00-1010-04 Male to Male, 75 foot, Unterminated Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 4 - 35...
Product Variations and Options Product variations, due to differences for various line voltages, may require different components. These variations are reflected, where necessary, on the parts lists. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 6 - 1...
Screw, #4-40 x .25 lg. Pan Head, Serial No. XXXX NOTES: Datascope Corp. maintains a policy of continuous development for product improvement and reserves the right to change materials, specifications, and prices without notice. 6 - 2 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 405
The use of other than Datascope components in these applications may result in degradation of reliability or instrument performance characteristics. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 6 - 3...
Keyboard Light Emitting Diode Metal Film MONO Monostable MYLR Mylar NTWK Network Operational Push Button Peripheral Interface Adaptor Potentiometer PRESS Pressure Power Random Access Memory Receiver Stage Stacked Supply 6 - 4 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 407
Replacement Parts Abbreviations ABBREVIATION TERM Switch SYST Systolic TANT Tantalum TRANS Transistor TRANSIS Transistor Variable Versatile Interface Adapter XDCR Transducer XFMR Transformer XSTL Crystal XSTR Transistor Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 6 - 5...
Page 408
Abbreviations Replacement Parts This page intentionally left blank. 6 - 6 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 409
Replacement Parts Available Replacement Parts and Sub-Assemblies FIGURE 6-1 Main Assembly Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 6 - 7...
Page 410
Available Replacement Parts and Sub-Assemblies Replacement Parts FIGURE 6-2 Planar Display Assembly 6 - 8 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 411
Replacement Parts Available Replacement Parts and Sub-Assemblies FIGURE 6-3 XE El Display Assembly Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 6 - 9...
Page 412
Available Replacement Parts and Sub-Assemblies Replacement Parts FIGURE 6-4 Detail A Electronic Frame - Top View 6 - 10 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 413
Replacement Parts Available Replacement Parts and Sub-Assemblies FIGURE 6-5 Detail A Electronic Frame - Bottom View With Datascope or Nellcor SpO Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 6 - 11...
Page 414
Available Replacement Parts and Sub-Assemblies Replacement Parts FIGURE 6-6 Detail A Electronic Frame - Bottom View With Datascope or Masimo SpO 6 - 12 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 415
Replacement Parts Available Replacement Parts and Sub-Assemblies FIGURE 6-7 Detail B/C NIBP Subassembly Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 6 - 13...
Page 416
Available Replacement Parts and Sub-Assemblies Replacement Parts FIGURE 6-8 Detail C Power Supply Subassembly 6 - 14 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 417
Replacement Parts Available Replacement Parts and Sub-Assemblies FIGURE 6-9 XG Color LCD Assembly Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 6 - 15...
Page 418
Available Replacement Parts and Sub-Assemblies Replacement Parts FIGURE 6-10 XG Rear Housing with Sidestream Parts 6 - 16 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 419
See tables on page 6-26 for the part numbers of various versions of the keypad overlays. This cable is part of the Daughter Bd. for all other versions other than XG-CE. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 6 - 17...
Page 420
See tables on page 6-26 for the part numbers of various versions of the keypad overlays. This cable is part of the Daughter Bd. for all other versions other than XG-CE. 6 - 18 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 421
See tables on page 6-26 for the part numbers of various versions of the keypad overlays. This cable is part of the Daughter Bd. for all other versions other than XG-CE. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 6 - 19...
Page 422
See tables on page 6-26 for the part numbers of various versions of the keypad overlays. This cable is part of the Daughter Bd. for all other versions other than XG-CE. 6 - 20 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 423
See tables on page 6-26 for the part numbers of various versions of the keypad overlays. This cable is part of the Daughter Bd. for all other versions other than XG-CE. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 6 - 21...
Page 424
See tables on page 6-26 for the part numbers of various versions of the keypad overlays. This cable is part of the Daughter Bd. for all other versions other than XG-CE. 6 - 22 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 425
See tables on page 6-26 for the part numbers of various versions of the keypad overlays. This cable is part of the Daughter Bd. for all other versions other than XG-CE. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 6 - 23...
Page 426
See tables on page 6-26 for the part numbers of various versions of the keypad overlays. This cable is part of the Daughter Bd. for all other versions other than XG-CE. 6 - 24 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 427
See tables on page 6-26 for the part numbers of various versions of the keypad overlays. This cable is part of the Daughter Bd. for all other versions other than XG-CE. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 6 - 25...
Page 428
VISA & Gas Module II/ Masimo SpO2 w/ XG2 units only This datasette provides an upgrade for 5 Lead, 5L, LT units to allow Admit from Bedside feature. 6 - 26 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 429
0334-00-1044-06 0334-00-1044-07 0334-00-1044-08 0334-00-1044-11 0334-00-1044-12 Left Side Panel Label with Nellcor Sp0 0334-00-1161-03 0334-00-1161-04 0334-00-1161-05 0334-00-1161-06 0334-00-1161-07 0334-00-1161-08 Left Side Panel Label with Masimo Sp0 0334-00-1457-01 0334-00-1457-02 0334-00-1457-03 0334-00-1457-04 Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 6 - 27...
Page 430
RES, 10K 1%, 1/8W R118 0307-06-2103 RES NET, 10K, 10 PIN SIP RP1, RP2 0307-00-0029 RES NET, 10K, 14 PIN DIP RP3, RP4 This component is used on the 0670-00-0657 board only. 6 - 28 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 431
DIODE, IN5817 CR25 0151-00-0195 XSTR, IRLD014 PWR MOSFET 0151-00-0052 XSTR, 2N4401 0155-00-0205 IC, 74LS244 OCTAL BUFFER 0155-00-0440 IC,DS1244Y, 32K x 8 This component is used on the 0670-00-0657 board only. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 6 - 29...
Page 432
SOCKET, IC PLCC 68 PIN 0136-01-1028 SOCKET, IC, 28 PIN 0136-01-1020 SOCKET, IC, 20 PIN XU7, XU9 0432-01-0002 MOUNTING PAD 0214-00-0074 INSULATOR BEAD This component is used on the 0670-00-0657 board only. 6 - 30 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 433
PART NUMBER DESCRIPTION REFERENCE 0125-01-0001 CABLE TIE WRAP 0285-10-0474 CAP, 0.47uF, 2%, 50V *C126 0307-00-1213 RES, 121K, 1%, 1/8W *R124 This component is used on the 0670-00-0657 board only. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 6 - 31...
Page 436
ITEM NO. PART NUMBER DESCRIPTION REFERENCE 0214-00-0074 Insulator Bead used on C1, C3 and C4, two each 0210-20-0004 Terminal Crimp Ring, 22-16 Use on 0670-00- AWG, #4 0621 only 6 - 34 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Key Pressed is Displayed on Screen Spo2 Diagnostics Accuracy Test 85% +/-1 DAC Test J24-5 Shared Ram Test Memory Test 1 Pass/Fail Memory Test 2 Pass/Fail Memory Test 3 Pass/Fail Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 7 - 3...
User input is provided through the following keys: • UP ARROW • DOWN ARROW • SELECT • END • RECORD (only when printable test results are displayed) • DEFLATE (for NIBP diagnostics only) 7 - 4 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
When this menu option is selected the Main Menu is covered by the NIBP Diagnostics Menu. The NIBP Diagnostics consists of three separate sub menus as shown in Figure 7-2. These tests are performed to check the pneumatics components of the unit. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 7 - 5...
NIBP module will open to release the pressure in the test chamber. NOTE: The NIBP hose with a safety leak fitting cannot be used, otherwise the test results are invalid. 7 - 6 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 509
The current pressure is displayed in the pressure window during the entire test. The format for displaying the test results is shown in Figure 7-5. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 7 - 7...
Page 510
The purpose of this test is to determine whether the overpressure sensor is operating properly. This sensor prevents an overpressure condition in the event that the normal sensing circuit fails. a. Connect the dummy cuff/test chamber to the side panel luer fitting. 7 - 8 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 511
Pressure Calibration The purpose of this test is to adjust the sensitivity of the transducer circuit for optimal accuracy and for checking the linearity at three separate pressure points. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 7 - 9...
Page 512
(if installed) and forwarded to a Datascope Service Representative or Datascope Technical Support department. 7 - 10 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
SpO2 Diagnostics FIGURE 7-7 Communication Test Menu 7.5.4.1 Front End Communication Test When this test item is selected, the following tests and test results are executed and displayed as follows: Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 7 - 11...
CPU Board are displayed. This test consists of putting the external communication channel into the loop mode and transmitting data, and verifying that the correct data is read back. 7 - 12 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
The results of the power up internal loop test for the NIBP communication channel on the CPU Board are displayed. The test consists of putting the NIBP communication channel into loop mode and transmitting data, and verifying that the correct data is read back. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 7 - 13...
Press the END key to return to the Main Diagnostics Menu. 7.5.7 Display Test When this test is selected the Main Service Menu is covered by the Display Tests Menu, which offers a choice of two different diagnostic tests. 7 - 14 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Press the END key (5L, LT), EXIT key (XG) to return to the Display Tests Menu. NOTE: Pattern #4 will appear when version “Y” software or higher is installed. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 7 - 15...
This will produce a square wave output on the four bits (LCD0-3) driving the video controller which can help identify a faulty CPU Board. 7 - 16 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Whenever a key is pressed the name of the key will be displayed in the key name window. e. Exercise each key to verify correct operation. f. Press the END key for three seconds to exit test and return to the Main Diagnostic Menu. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 7 - 17...
Accuracy Test Specification % SpO2 = 85% Press the RECORD key to obtain a printout of the test results. c. Press the END key to return to the SpO2 Test. 7 - 18 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Memory Test 2 (Databus Test): The main processor will write a bit pattern of walking one’s (8-bit) to each memory location and read it back to verify that the data bus is not cross talking. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 7 - 19...
At the end of the download, verify unit sets up display screens for signal inputs. NOTE: If no display is observed when unit is powered up, adjust front panel “contrast ” control. 7 - 20 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
1. Observe that the trace display sweeps across the Waveform 1 screen in two seconds. There should be two complete ECG cycles. The same display and timing should be seen in the Waveform 2 screen. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 7 - 21...
Page 524
Decrease the ECG HR to 45 bpm. The ALARM should sound. Then set the ECG HR to 60 bpm and verify the alarm condition is canceled. 7 - 22 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
4. Verify that the temperature displays is 37.0º +/-0.3. º º NOTE: If temperature exceeds 45 C or is below 0 C, the temperature display goes to “dashed lines”. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 7 - 23...
If the monitor detects changing CO2 levels (breaths) during an adapter calibration, a “Breaths Detected ... Retry” message displays and then an “Adapter Cal Failure” message displays. Remove the source of CO and repeat the calibration. 7 - 24 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 527
“Sensor Warming Up” message must not be displayed. To perform a Capnostat® sensor calibration: 1. Verify the Passport XG is turned on and the Capnostat® is plugged in and warmed-up. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 7 - 25...
(breaths) during a pump calibration, a “Breaths Detected ... Retry” message displays, and then a “Pump Cal Failure” message will display. Remove the source of CO and repeat the calibration. 7 - 26 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
If the monitor detects changing CO2 levels (breaths) during a adapter calibration, a “Breaths Detected ... Retry” message displays, and then an “Adapter Cal Failure” message will display. Remove the source of CO2 and repeat the calibration. 7 - 28 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Passport XG CO connector (34). The “CO2: Sensor Warming Up” message is displayed. The sensor calibration cannot be performed until the “CO2: Sensor Warming Up” message disappears. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 7 - 29...
The Passport was evaluated with the external supply by UL and CSA with no grounding impedance verification. If you check U-blade resistance, the safety analyzer will read “open”. 7 - 30 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 533
4. On the Safety Analyzer depress the “APPLY 115VAC” button and note the reading. 5. Repeat the test for normal and open ground and reverse polarity combinations. Verify the current reading for any test is less then 20uA. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 7 - 31...
DOWN arrow key to turn the Blank Screen Menu on or off. Press the Select key to choose the selection. when the Blank Screen Menu is turned on, the display will be blank until a key is pressed or power is cycled. 7 - 32 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
Page 535
Barometric Pressure (XG Sidestream units only): 1. This menu is non-selectable. Barometric Pressure is displayed in mmHg’s and is updated as per real time pressure. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 7 - 33...
Page 536
Special Options Menu Calibration This page intentionally left blank. 7 - 34 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
SpO2, refer to sections 1.3.9 (Nellcor) or 1.3.10 (Masimo) and/or the individual instruction sheets that are packaged with each sensor. This feature applicable only if available / installed in your unit. 8 - 2 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
It is recommended that gloves be worn while performing this procedure. 1. Shut off the monitor and disconnect the external power supply. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 8 - 3...
Allow the cuff to air dry, then insert the latex inflation bag. CAUTION: When ironing or pressing the cuffs, be aware that the Velcro® fasteners can melt at temperatures above 325°F, 162°C. 8 - 4 0070-00-0420 Passport 5-Lead, 5L, LT, XG Service Manual...
• Ultraviolet Light We recommend storing the recordings in a filing cabinet within a few days of printing. Long term exposure to natural or artificial U.V. sources may be detrimental. Passport 5-Lead, 5L, LT, XG Service Manual 0070-00-0420 8 - 5...