Step 5
Secure Clips
1. SCRUBBED PERSON: After aseptically removing the sterile power shell from the
packaging, carefully open the shell by holding the back handle of the shell so the
front handle is facing up and away from the back handle.
2. SCRUBBED PERSON: Align and fully seat a clean, sterilized reusable insertion
guide onto the back handle of the open power shell to provide an aseptic transfer
guide when inserting the power handle into the shell.
PRECAUTION
Ensure that the reusable insertion guide is properly seated onto the power shell
and all clips are covered before inserting a power handle device.
3. CIRCULATING PERSON: While maintaining aseptic transfer techniques, insert the
power handle into the insertion guide and power shell.
4. CIRCULATING PERSON: After the power handle is fully seated in the power shell,
carefully remove the reusable insertion guide using the extended remover handle.
5. SCRUBBED PERSON: Taking care not to touch the power handle, close the front
portion of the power shell until there is tactile confirmation the base of the shell is
closed and the front clips are secured. Check all three clips to confirm the shell is fully
closed and securely locked.
PRECAUTION
Ensure that both front clips are secured before use.
6. Confirm successful connection of the sterile control shell with the power handle and
the functional status of the stapler on the display screen (see Table 10).
Display Screen Status
17
Instructions for Use