multi N/C 2100
Computer restart in the analysis device
Cause
−
Internal computer reset
−
Overvoltage
−
Short-term power failure
Internal program restart
Cause
−
Internal program error
Analyzer error: MESSx measurement is canceled
Cause
−
Equipment fault
Minimum sample volume > cup volume
Cause
For sample supply with sampler:
−
The sample volume selected is
too large
−
The number of detections is too
high
Edition 02.17
Remedy
−
If the front LED display "internal com-
puter" lights, initialize the analyzer
−
If this is repeated, monitor precisely at
which time the error occurs (note sta-
tus line)
Remedy
−
Initialize the analyzer
−
If this is repeated, monitor precise-
ly at which time the error occurs
(note status line)
Remedy
−
Initialize the analyzer
−
Check the System state window
for error confirmation
−
Search for the equipment fault and
remove the error
Remedy
Check the configuration in the method:
−
Sample volume/rinsing volume
−
Adjust the number of determina-
tions (repeat measurements) to
the cup volume