multi N/C 2100
Computer restart in the analysis device
Cause
Internal computer reset
Overvoltage
Short-term power failure
Internal program restart
Cause
Internal program error
Analyzer error: MESSx measurement is canceled
Cause
Equipment fault
Minimum sample volume > cup volume
Cause
For sample supply with sampler:
The sample volume selected is
too large
The number of detections is too
high
Edition 02.17
Remedy
If the front LED display "internal com-
puter" lights, initialize the analyzer
If this is repeated, monitor precisely at
which time the error occurs (note sta-
tus line)
Remedy
Initialize the analyzer
If this is repeated, monitor precise-
ly at which time the error occurs
(note status line)
Remedy
Initialize the analyzer
Check the System state window
for error confirmation
Search for the equipment fault and
remove the error
Remedy
Check the configuration in the method:
Sample volume/rinsing volume
Adjust the number of determina-
tions (repeat measurements) to
the cup volume
Troubleshooting