Endress+Hauser analytikjena multi N/C 2100 Operating Manual
Endress+Hauser analytikjena multi N/C 2100 Operating Manual

Endress+Hauser analytikjena multi N/C 2100 Operating Manual

Toc/tnb analyzer
Table of Contents

Advertisement

Quick Links

Operating Manual
multi N/C 2100
TOC/TN
Analyzer
b

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the analytikjena multi N/C 2100 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Endress+Hauser analytikjena multi N/C 2100

  • Page 1 Operating Manual multi N/C 2100 TOC/TN Analyzer...
  • Page 2 Analytik Jena AG Manufacturer Konrad-Zuse-Str.1 07745 Jena  Germany Phone + 49 3641 / 77 70 + 49 3641 / 77 92 79 Email info@analytik-jena.de Analytik Jena AG Service Konrad-Zuse-Str. 1 07745 Jena  Germany Phone + 49 3641 / 77-7407 (Hotline) Email service@analytik-jena.de http://www.analytik-jena.com...
  • Page 3: Table Of Contents

    Index of Figures Table of Contents Basic information........1 4.1.5 Detectors ............. 23 User manual notes ........1 4.1.5.1 NDIR detector ..........23 Intended use ..........1 4.1.5.2 Electrochemical NO detector (ChD, optional) ..........23 1.2.1 General ............1 4.1.6 Indicator and control elements, 1.2.2 Analyzer area of application ......
  • Page 4 Index of Figures 4.6.1.3 Eluate blank value ........37 8.5.3 Filling the combustion tube ......67 4.6.2 Boat blank value .......... 37 8.5.4 Inserting the combustion tube ...... 69 First commissioning ........ 38 Regenerating the TIC reactor ...... 71 Site requirements ........38 Cleaning the TC condensate vessel ....
  • Page 5 Index of Figures Index of Figures Fig. 1 Front view (doors open) ......14 Fig.20 Media connections on the backplate of the solids module ........... 42 Fig. 2 Lateral view left (side wall removed) ..15 Fig.21 Window Calibration - Data of new Fig.
  • Page 6: Index Of Figures

    Index of Figures Edition 02.17 multi N/C 2100...
  • Page 7: Basic Information

    Basic information Basic information User manual notes The analyzer multi N/C 2100 is intended for operation by qualified specialist personnel ob- serving this user manual. The user manual contains information about the design and operation of the analyzer and provides personnel familiar with TC/TN analysis the necessary know-how for the safe han- dling of the equipment and its components.
  • Page 8: Analyzer Area Of Application

    Basic information The operational safety of the analyzer multi N/C 2100 is only ensured during proper use according to the information in this user manual. The intended use also includes the adher- ence to the installation conditions prescribed by Analytik Jena AG which are available from the customer service address stated above.
  • Page 9: Warranty And Liability

    Basic information Warranty and liability The warranty duration and liability comply with the legal requirements and the provisions in the general terms and conditions of Analytik Jena AG. Deviations from the intended use described in this user manual will result in limitations of warranty and liability in the event of damage.
  • Page 10: Technical Data

    Technical data Technical data General characteristics Designation/type Analyzer multi N/C 2100 Dimensions Basic device (W x H x D) (513 x 464 x 540) mm Mass approx. 28 kg Procedural data Digestion principle Thermocatalytic oxidation Digestion temperature up to 950°C, depending on the catalyst Catalyst −...
  • Page 11 Technical data Environmental conditions Temperature during storage 5 – 55°C Temperature during operation 10 – 35°C Humidity during operation max. 90% at +30°C Humidity during storage 10 – 30% (use drying agent) Air pressure 0.7 – 1.06 bar Minimum equipment for the control and analysis unit Operating system: Windows 7 Professional or higher Processor:...
  • Page 12: Safety Instructions

    Safety instructions Safety instructions For your own safety and to ensure error-free and safe operation of the analyzer multi N/C 2100, please read this chapter carefully before using the appliance. Observe all safety notes listed in this user manual and all messages displayed by the control and analysis software on the monitor.
  • Page 13: Symbols And Signal Words Used

    Safety instructions Chinese directives The analyzer contains regulated substances (according to directive SJ / T 11363-2011). Analytik Jena AG guarantees that these substances will not leak out during the next 25 years if the analyzer is used in accordance with its intended use and thus do not pose a threat to the environment or health within this period.
  • Page 14: Technical Condition

    Safety instructions Fire, naked flame and smok- Warning against dangerous Warning against hazardous ing prohibited! electrical voltage substances Warning against hot surface Warning against corrosive substances The device contains restricted substances (according to di- rective SJ/T 11363-2011). Analytik Jena AG guarantees, that those hazardous substances may not leak out during the next 25 years when the device is used in accordance with its in- tended purpose.
  • Page 15: Safety Instructions, Transport And Installation

    Safety instructions Therefore, every person entrusted with the operation of the analyzer must have read and understood this user manual and the user manuals of any additional equipment before carry- ing out the respective tasks. This also applies if the respective person has already worked with or been trained on this kind of analyzer.
  • Page 16: Safety Instructions - Operation

    Safety instructions Safety instructions - operation 3.7.1 General The operator of the analyzer must make sure before each commissioning that the condition of the analyzer including the safety equipment is sound. This applies in particular after each modification or extension of the analyzer or its repair. The following has to be observed: ...
  • Page 17: Safety Instructions For Compressed Gas Containers And Systems

    Safety instructions  The electrical components must be checked regularly by a qualified electrician. Any defects, such as loose connections, faulty or damaged cables, must be repaired without delay.  The analyzer must be switched off immediately at the power switch (on the equip- ment backplate) and the power plug disconnected from the mains if there is any in- terference with the electric components.
  • Page 18: Safety Instructions - Service And Repair

    Safety instructions  Auxiliary and operation materials may never be placed in containers or vessels for food. The approved containers for the relevant material are to be used and these have to be labeled accordingly. The notes on the labels have to be observed! ...
  • Page 19: Behavior In An Emergency

    Safety instructions Behavior in an emergency The analyzer must be switched off at the power switch (on the equipment backplate) and the power plug must be disconnected from the mains in the case of dangerous situations or ac- cidents. Because a rapid response can save lives during an emergency, the following has to be en- sured: ...
  • Page 20: Technical Description

    Technical description Technical description System design The analyzer multi N/C 2100 is a compact bench-top device with permanently installed main components. The complete measuring design also includes accessory parts and reagents which must be connected to the analyzer or made ready prior to a measurement. The control of the analyzer and the analysis of the measurements takes place via the control and analysis software multiWin.
  • Page 21: Components For Sample Preparation

    Technica al description Condensa ation coil Combustio on system Gas box Fig. 2 Lateral view left (sid de wall remov ved) Fig. 3 TC lock k at the top of f the device 4.1.1 Comp ponents for samp ple prep aration 4.1.1.1 1 TIC lo...
  • Page 22: Hose System

    Technical description For the analysis of particulate samples, the use of cannulas with a larger inside diameter (particle cannula) is recommended. Special microliter syringes (special geometry, with gas connector for analysis in NPOC mode) are used for the samplers. They are not calibrated and therefore not suitable for anal- ysis in manual operation.
  • Page 23: Fig.4 Hose Diagram

    Technical description Fig.4 Hose diagram multi N/C 2100 Edition 02.17...
  • Page 24: Components For Flow Adjustment

    Technical d escription 4.1.2.1 Compo nents fo or flow ad djustmen The carrier r gas flow is a adjusted via the needle v valve at the f ront side of t the device. T The inlet flow is mea asured using an MFM (m ass flow met ter) and disp...
  • Page 25: Condensate Pump

    Technical description Fig.7 Condensate pump 4.1.2.3 Connection method Within the device most gas connections have been implemented via a FAST connector (FAST – Fast, Save, Tight). These connectors provide a tight transition between the hoses and connections with different diameters. The soft sleeves prevent the risk of glass fractures compared to rigid screw connections.
  • Page 26: Combustion System

    Technical d escription 4.1.3 Combu stion sys stem The combu ustion system m is on the re ear left side w wall of the an nalyzer. The combu ustion furnac e is a resisto or-heated ve rtical furnace e for digestio on temperatu res up to 950°C.
  • Page 27: Tic Condensate Container

    Technica al description Fig.11 Conden nsation coil 4.1.4.2 2 TIC c condensa ate conta ainer In the T TIC condensa ate container r the TIC rea ctor and the gas/liquid se eparator are combined. It is loca ated on the le eft front side .
  • Page 28: Water Traps

    Technical description 4.1.4.3 Water traps The analyzer multi N/C 2100 contains water traps to protect the detector and the gasbox. Two water traps are installed between the TIC condensation vessel and the halogen trap. They prevent condensed water from entering the measuring gas line. The larger water trap (TC pre-filter) retains aerosol during operation, the smaller water trap (disposable retention filter) retains rising water.
  • Page 29: Detectors

    Technical description 4.1.5 Detectors 4.1.5.1 NDIR detector The NDIR detector (NonDispersive Infrared absorption Detector) is behind the right-hand side wall of the analyzer. Gases with molecules from different atoms have specific absorption bands in the infrared wavelength range. When a light beam is sent through an arrangement of cells which con- tains gases in the IR active range, these gas components absorb a proportional part of the total radiation according to their concentration part in the gas mixture on their characteristic wavelengths.
  • Page 30: Indicator And Control Elements, Connections

    Technical description 4.1.6 Indicator and control elements, connections LED displays The green LED at the left door of the analyzer lights after the device has been switched on. Fig.15 LED to indicate readiness for operation The LED strip behind the right door indicates different operating states of the analyzer. Internal firmware controller Device voltage Furnace heating...
  • Page 31: Accessories

    Technical description Connection of the neutral conductor at the sam- "internal" gas connection pler "NPOC" NPOC purging gas connection Main switch to switch the analyzer on and off Carrier gas "O /Air" connection Holder for mains fuse "Waste" Mains connection RS 232 interface for the "sampler"...
  • Page 32: Extension Options For The Analyzer

    Technical description 4.1.8 Extension options for the analyzer 4.1.8.1 Sampler (optional) The sampler APG 60 is attached to the basic device using two Allen screws. It is suitable for both homogeneous and inhomogeneous particulate samples. Each sample can be stirred individually immediately before analysis. The stirring speed can be selected. In NPOC mode, the samples can be automatically acidified and purged.
  • Page 33: Principle Of Operation

    Technical description Principle of operation The analyzer multi N/C 2100 is a compact and powerful device for the determination of the total carbon content and/or total nitrogen content in aqueous samples. Fig.18 Principle of operation The digestion is performed in the multi N/C 2100 by thermocatalytic high-temperature oxida- tion in the presence of special catalysts.
  • Page 34: Measuring Method

    Technical description Inorganic carbon is determined by injecting a sample aliquot into the acidic TIC reactor and driving out the formed CO via the NDIR detector. The CO or NO concentration is detected several times every second. An integer over time is calculated from this signal sequence.
  • Page 35: Tic Analysis

    Technical description Two sequential measurements are used on one sample to determine TIC and TC. The cal- culated difference is given as TOC. The differential method detects volatile as well as non- volatile organic carbon compounds. TOC analysis should be used when the sample contains easily purgeable organic substanc- es as benzene, cyclohexane, chloroform, etc.
  • Page 36: Tn B Analysis

    Technical description 4.3.6 analysis Parallel to all analyses via high temperature combustion, the determination of the total bound nitrogen is possible in the TN analysis. The thermocatalytic oxidation results in nitrogen oxides which can be determined using an optional electrochemical detector (ChD). Catalysts Various suitable solids can be used as catalysts or oxygen carriers in the multi N/C 2100 supporting the combustion of the components of the sample material in a temperature range...
  • Page 37: Calibration

    Technical description Calibration 4.5.1 Calibration strategies 4.5.1.1 Multiple point calibration with constant sample volume For most applications, a multiple point calibration with constant sample volume and varia- ble concentrations is carried out. The concentration series for the ranges to be calibrated are prepared and the settings measured in the selected method.
  • Page 38: 4.5.3.1 Tc/Npoc

    Technical description ∗ ∗ ∗ c … target concentration of the standard V … sample volumes … net integer … calibration coefficient The net integer is the raw integer corrected by the preparation water. The regression type (linear or quadratic) can be defined by the user. It is possible to select individual measuring points or measured values for the calculation of the current calibration (manual outlier selection).
  • Page 39: Toc (Diff)

    Technical description 4.5.3.3 TOC (Diff) Generally, separate calibration functions are calculated for the TC and TIC channels in ac- cordance with the equations (6) or (7) for concentrations > 0.5 mg/l. The equations (8) and (9) apply. The calculation of the analysis results is based on the calculated calibration functions for TC and TIC.
  • Page 40: Method Characteristics

    Technical description 4.5.4 Method characteristics Remaining standard deviation The remaining standard deviation (remaining variance) expresses the dispersion of the inte- gers around the regression function (regression precision). Standard deviation of the method The standard deviation of the method describes in a unique and general way the quality of the calibration.
  • Page 41: Other Calculations

    Technical description 4.5.5 Other calculations For all measurements where multiple injections are carried out the mean value (MV), stand- ard deviation (SD) and variation coefficient (VC) are calculated and displayed. For each sample a tenfold determination can be carried out as a maximum. Outlier selection The control and analysis software multiWin offers the option for an automatic outlier selec- tion.
  • Page 42: Water For Diluting Solutions Diluting Blind Value

    Technical description 4.6.1.2 Water for diluting solutions diluting blind value If the sample needs to be diluted, the blank value of the dilution water might be of interest. This value can be determined separately and entered in multiWin. This value can change over time and must be redetermined before the start of each measuring series.
  • Page 43: 4.6.1.3 Eluate Blank Value

    Technical description The integer values determined for a sample can be easily entered. If the primary sample has been diluted and the dilution ratio entered in multiWin, the concentration of the primary sam- ple is specified in the analysis report. 4.6.1.3 Eluate blank value The eluate blank value is a special blank value for samples from the purity validation or elu- ate preparation.
  • Page 44: First Commissioning

    First commissioning First commissioning Site requirements 5.1.1 Installation conditions The following requirements are placed on the climatic conditions in the operating room of the analyzer:  Temperature range: +10°C to +35°C Max. humidity: 90% at 30°C  Air pressure: 0.7 bar to 1.06 bar ...
  • Page 45: Power Supply

    First commissioning 5.1.3 Power supply WARNING! ELECTRIC SHOCK! The analyzer multi N/C 2100 must only be connected to a properly grounded mains outlet in accordance with the voltage specifications on the type plate! The multi N/C 2100 is operated from a single phase alternating current mains. The installation of the electrical equipment of the laboratory must comply with the standard DIN VDE 0100.
  • Page 46: Connecting Add-On Devices

    Connecting add-on devi ices Conn necting g add- -on de evices CAUTIO Before con nnecting any y add-on dev vices switch o off the analyz zer. Only eve er connect a dd-on devices to o the multi N/ /C 2100 whe en it is switch hed off! Samp...
  • Page 47 onnecting add d-on devices Connect the s sampler to th e power sup pply in accord dance with th he user man ual. Connect the s serial data ca able supplied d to the "sam pler" interfac ce at the equ uipment ackplate of th he analyzer...
  • Page 48: External Solids Module

    Connecting add-on devices Before the first start, the sampler must be adjusted (see section "Adjust the sampler" p. 59). If the piston does not move fully down after initializing the sampler, you need to adjust the piston (see section "Adjusting the piston position" p. 60). External solids module CAUTION Before connecting any add-on devices switch off the analyzer.
  • Page 49: Operation

    Operation Operation General information on analysis work Observe the following during analysis: When analyzing samples with high acidic or saline content, aerosol can be formed in  the TIC condensation vessel. The capacity of the halogen trap is depleted within a relatively short time.
  • Page 50: Switching On The Analyzer (Standard Commissioning)

    Operation Switching on the analyzer (standard commis- sioning) WARNING Concentrated phosphoric acid is required for the measurement process! Contact with phosphoric acid can cause severe skin and eye injuries! Always wear goggles and protective gloves when handling concentrated phosphoric acid! CAUTION Damage to optical and electronic components (detectors, flow sensors) from aggressive combustion products if the copper wool in the halogen trap is used up!
  • Page 51: Carrying Out The Calibration

    Operation IMPORTANT In the System state window the displays of the components which are not yet ready are shown in red during initialization. During the start-up phase of the analyzer the external communication with the program is blocked. The individual components have different start-up periods: −...
  • Page 52: Fig.21

    Operation  After loading the method to be calibrated or after opening an existing calibra- tion table the Calibration –Data of new calibration window is opened. Fig.21 Window Calibration - Data of new calibration In the group Calibration parameters select the calibration type. Note: Preferably multipoint calibrations with constant sample volume and variable concentra- tions should be carried out.
  • Page 53: Fig.22

    Operation of the sampler. For manual sample supply, the provisioning of the preparation wa- ter is first requested. − Selection field enter: The content of the preparation water can be entered as a value. Note: The preparation blank must be specified standardized to 1 ml. If the preparation blank is not taken into account, enter a 0 in the input field.
  • Page 54: Displaying Calibration Results

    Operation 13. After opening the Measurement window click on the [Start F2] button.  The calibration process starts. 7.3.2 Displaying calibration results After the calibration measurements have been completed, the calibration report is automati- cally opened in the Calibration – Calibration settings window and can be edited. The cali- bration report can also be opened later via the menu command Data Evaluation / Calibra- tionReport / Select CalibrationReport .
  • Page 55: Editing An Existing Calibration

    Operation Linear regression/ Dependent on the methodology selected, the regression calcu- lation and determination of the method characteristics is based Quadratic regression on individual values or mean values of the net integers. For the selected regression type the respective calibration coefficients are displayed.
  • Page 56: Fig.24

    Operation All measuring points enabled by () in the No. column of the result table are includ- ed in the regression calculation. A measuring point can be disabled by removing the () (click in the No. column).  Disabling individual measured values By clicking the button at the end of each line of the results table you can view the in- dividual measured values (see Fig.24).
  • Page 57: Transferring Calibration Parameters To A Method

    Operation 7.3.4 Transferring calibration parameters to a method Transfer calibration parameters to a method as follows: Select an appropriate calibration range for the respective parameters (e.g., NPOC/TN). Note: Up to three linear calibration ranges for each parameter can be stored in a method. It should be ensured that the ranges merge and do not have any gaps! Using calibration function with quadratic regression only one calibration range can be stored in a method.
  • Page 58 Operation The definition of the calibration range depends on those stored in the method and the new calibration range: No calibration − Transfer the currently determined calibration data with the [Ac- range exists cept values] button. The same calibration coefficients appear in the left-hand and right-hand columns.
  • Page 59: Managing Calibration Data

    Operation Fig.26 Window Link with method with three areas 7.3.5 Managing calibration data Printing calibration data Print the calibration report as follows: In the Calibration – Calibration settings window enable the use calibration option. Define the scope of printing under the menu Print options : −...
  • Page 60: Performing Measurements

    Operation Reopening a calibration report In the main window open the menu command Data Evaluation / CalibrationReport . In the Selection CalibrationReport window select the calibration report. Note: In the Selection CalibrationReport window filters can be set, if necessary, and the records sorted by clicking on the respective header.
  • Page 61: Measurement With Sampler

    Operation − If applicable ChD - OK − Gas flow - OK − Temperature - OK Note: If one of the entries is incorrect (shown in red), carry out a fault finding exercise in ac- cordance with the notes in section "Troubleshooting" p. 81. In the System State window select the manual sample supply by clicking the [manual] button.
  • Page 62 Operation − If applicable, ChD - OK − Gas flow - OK − Temperature - OK Note: If one of the entries is incorrect (shown in red), carry out a fault finding exercise in ac- cordance with the notes in section "Troubleshooting" p. 81. In the System State window select the sample preparation with sampler by clicking the [Sampler] button.
  • Page 63: Maintenance And Care

    Maintenance and care Maintenance and care Maintenance intervals Analyzer Maintenance task Maintenance interval Clean and maintain the device Weekly Inspect all hose connections for tight fit Monthly Inspect the fastening screws for tight fit Monthly Water traps Maintenance task Maintenance interval Check the gas flow Daily Exchange the water traps...
  • Page 64 Maintenance and care Locks Maintenance task Maintenance interval Check for leaks 3 months Replace the septum of the TC or TIC lock As required, but no later than after 12 months ChD (optional) Maintenance task Maintenance interval Replace the support battery 12 months IMPORTANT To enable regular inspections and maintenance tasks to be carried out, always ensure that...
  • Page 65: Adjustment And Setup Tasks

    Maintenance and care Adjustment and setup tasks 8.2.1 Adjust the sampler An adjustment of the sampler is necessary:  before the first start  after every syringe replacement  after every manipulation at the locks (e.g, catalyst replacement and maintenance tasks) ...
  • Page 66: 8.2.1.2 Adjusting The Piston Position

    Maintenance and care Carry out the adjustment as follows: Enter the syringe size: − With the menu command Configuration / Edit options to call the Options win- dow/ Analyzer components tab. − In the group Syringe in the list size select the specified syringe volume. −...
  • Page 67: Adjust The Npoc Purge Flow

    Maintenance and care In the group Please select position needing adjustment select the Piston position from the list. Increase or reduce the z coordinate in 0.1 mm increments to align the piston position: Note: Move the syringe piston down in steps until the slit just disappears Start the adjustment for the syringe piston position by clicking on the [Position piston adjust] button.
  • Page 68 Maintenanc ce and care − Select the po osition at wh hich the purg ing flow is to o be monitore ed in the Rac ck posi- tion field be tween 1 and d 60. − Place a sam mple cup with h ultrapure wa ater onto this s position.
  • Page 69: Replacing The Water Traps

    Maintenan nce and care lacing the wa ater tra IMPO ORTANT Replac ce the water traps depen ndent on the sample matr rix, but no lat ter than after r 6 months. Always s replace all water traps! The water tr raps only ser rve their func ction if they a...
  • Page 70 Maintenance and care CAUTION There is a risk of burning! The combustion system remains hot immediately after switch- ing off the analyzer! Wait for the combustion system to cool down before performing maintenance on the internal parts of the analyzer. Carrier gas connection to the combustion system Hose from the gasbox...
  • Page 71: Replacing The Halogen Trap

    Maintenan nce and care lacing the ha alogen trap CAUT TION Damag ge to optical and electron nic componen nts (detector s, flow senso ors) from agg gressive combu stion produc cts if the copp per wool in th he halogen t rap is used u Replac ce the comple...
  • Page 72: Exchanging The Catalyst

    Maintenanc ce and care Excha anging the ca atalyst 8.5.1 Lifetime e of the c catalyst If the effect tiveness of th he catalyst d decreases, th he combustio on tube has t to be filled ag gain. A check has to be perform med after exp piry of the se...
  • Page 73: Filling The Combustion Tube

    Maintenance and care Disconnect the spherical joint (3) holding the bottom end of the combustion tube and the entrance of the condensation coil to- gether. To do so, release the knurled head screw (1) and remove the fork clamp (2). Carefully pull the combustion tube upwards out of the combustion furnace.
  • Page 74 Maintenanc ce and care Filling the combustion n tube for co onventiona l samples You can fix xate the comb bustion tube on a tripod f for fill- ing. Fill the combustion tube in acco ordance with following in struction from m the bottom m up: Insert...
  • Page 75: Inserting The Combustion Tube

    Maintenance and care 8.5.4 Inserting the combustion tube CAUTION Alkaline salts (hand perspiration) cause crystallizations in the quartz glass when heating the combustion furnace which reduce the service life of the combustion tube. Where possible do not touch the cleaned combustions tube with your hands. Wear protec- tive gloves to install the lock on the combustion tube.
  • Page 76 Maintenanc ce and care Position the ceramic hol lder in the to p open- ing of the co ombustion fu urnace. Insert the co ombustion tu ube with the f furnace head into th e combustio on furnace. Screw the lo ock with the A Allen screw tight to...
  • Page 77: Regenerating The Tic Reactor

    Maintenance and care Regenerating the TIC reactor WARNING The TIC reactor is regenerated with 40% phosphoric acid! Phosphoric acid is irritating for eyes, skin and the mucous membrane! Always wear goggles and protective gloves when handling concentrated phosphoric acid! Rinse the affected skin with water immediately! CAUTION A cannula that is too large will damage the septum at the septum port.
  • Page 78: Maintaining The Condensation Coil

    Maintenanc ce and care Open n the doors o of the analyz ll the connec ction hose to o the water tr aps (1) the TIC con densate con ntainer. ll the hose n o. 1 off the F FAST connec ctor.
  • Page 79 Maintenan nce and care Pull the h hose 1 off th e FAST con nector of the cond densation coi Disconne ect the spher rical joint (3) between the lowe r end of the c combustion tube and the entra ance of the co condensation n coil by re-...
  • Page 80: Installing The Condensation Coil

    Maintenanc ce and care 8.8.2 Installin ng the co ondensa tion coil Insert the c condensation n coil as follo ows: Push the ho ose 1 into the e FAST conn nector. Use the thre eading aid, if f necessary ( (see section "Rep placing the h...
  • Page 81: Removing And Installing The Combustion Furnace

    Maintenan nce and care moving and in nstallin ng the c combu ustion furn nace 8.9.1 Remo oving the e combu ustion fur rnace CAUT TION There i is a risk of bu urns on the c combustion f furnace! Only y remove the e combustion n tube...
  • Page 82: Installing The Combustion Furnace

    Maintenanc ce and care 8.9.2 Installin ng the co ombustio on furna Install the c combustion f furnace as fo ollows: ove the top c cover. Open the lef ft side wall o of the analyze er. Un- screw the fo our fastening screws;...
  • Page 83: Removing And Replacing The Pump Hose

    Maintenan nce and care 8.10 moving and re eplacin ng the p pump h hose CAUT TION The pu ump hose co ontains phosp phoric acid! Phosphoric a acid is irritati ng for eyes, skin and the mu ucous memb brane! Always s wear gogg...
  • Page 84: Replacing The Hose Connections

    Maintenanc ce and care Positio on the conve eyor belt arou und the pump body. 10. Press the conveyo or belt upwar rds with one h and and rota ate the brack ket with the ot her hand to t the right unti il it en- gages...
  • Page 85: Fig.30

    Maintenance and care For angled FAST connectors make sure not to slide the hose ends beyond the branch length of the connector in order to guarantee an unimpeded gas flow. Angled FAST connector Hose Connection piece Fig.30 Angled FAST connector with connected hose When replacing damaged finger-tight screw connections, only use straight cut, round and unpinched hose ends for the connection.
  • Page 86: Replacing The Septums

    Maintenance and care 8.12 Replacing the septums The septums have to be exchanged at regular intervals. A change is required when the tightness of the system is not ensured anymore. The septum of the TIC lock The septum of the TIC lock should be replaced when necessary. Unscrew the lock from the plastic knurled head screws in a counter-clockwise direction.
  • Page 87: Troubleshooting

    Troubleshooting Troubleshooting General information The following chapter describes a number of possible problems that the user can partially remedy independently. If such problems occur frequently, the Service Department of Analytik Jena AG must always be informed. As soon as the multi N/C 2100 is switched on, system monitoring takes place. Any errors occurring are displayed in a window after start-up.
  • Page 88: Error Messages In Multiwin

    Troubleshooting Error messages in multiWin Error Error message code VERS Communication error - incorrect command set between PC and device! Cause Remedy − − The internal and external program Update the internal and external versions do not match program VERS1 Communication error - analyzer Cause Remedy...
  • Page 89 Troubleshooting COM 2 not found COM 3 not found COM 4 not found Cause Remedy − − Problems with the internal hard- Switch the analyzer off/on ware Gas pressure error Cause Remedy − Counterpressure in the analysis − Search for and replace the com- system too great: the carrier gas ponent causing the gas pressure supply is automatically interrupted...
  • Page 90 Troubleshooting No connection to optics (NDIR) CRC error optics optics status error optics error; incorrect command Cause Remedy − − Communication error Initialize the analyzer − − The NDIR detector is faulty Inform Service Optics error, analog values outside of range Cause Remedy −...
  • Page 91 Troubleshooting Computer restart in the analysis device Cause Remedy − − Internal computer reset If the front LED display "internal com- puter" lights, initialize the analyzer − Overvoltage − If this is repeated, monitor precisely at − Short-term power failure which time the error occurs (note sta- tus line) Internal program restart...
  • Page 92: Status Errors - Indications In The System State Window

    Troubleshooting Status errors – indications in the System state window Note: Status errors are shown in the System state window in red or yellow. Values of the NDIR detector below opt. bank indicated in yellow Cause Remedy − − The analog values of the detector are Measurements are still possible, but it outside the working range.
  • Page 93 Troubleshooting − − The cannula is damaged Replace the cannula − Use a cannula suitable for particles if the solutions contain particles − − Inhomogeneous sample matrix Warm up cold samples before analysis − Filter samples prior to analysis − −...
  • Page 94 Troubleshooting samples) replace the catalyst − When using a special catalyst for multi N/C, replace the catalyst Note: After a catalyst replacement a calibra- tion must be carried out. − Use a cannula suitable for particles if the solutions contain particles −...
  • Page 95 Troubleshooting Manual metering in locks with septum faulty Cause Remedy − − Uneven dosing Ensure even injection − Do not inject too fast; the injection speed has to be reduced with an increased sample volume − − Carry-over, since the sample is Meter the sample vertically into the reactor metered on the reactor wall Sample is not drawn up without air bubbles...
  • Page 96 Troubleshooting Control lamps at the analyzer do not light: 5 V, 24 V Cause Remedy − − Error in the power supply or in the Check the electrical connections electronics − Check the lab's power supply − − The equipment fuse is faulty Inform Service Front LED display on the analyzer does not light: Lockin Cause...
  • Page 97: Transport And Storage

    Transport and storage Transport and storage 10.1 Transport 10.1.1 Preparing the analyzer for transport WARNING! RISK OF BURNS AT THE HOT FURNACE! There is a risk of burns at the combustion furnace! Only remove the combustion furnace when the device is cold or allow the device to cool down sufficiently! CAUTION! RISK OF INJURY FROM GLASS BREAKAGE! When removing the glass components there is a risk of injury from glass breakage! Re- move all glass components carefully from the analyzer!
  • Page 98: Transport Notes

    Transport and storage 14. Attach the top cover and secure it with adhesive tape. 15. Carefully package the accessories, in particular protect glass components against breakage. 10.1.2 Transport notes Observe the safety instructions in section "Safety instructions, transport and installation" p. 9. Transport the analyzer very carefully to prevent damage from impact or vibration.
  • Page 99: Storage

    Transport and storage 10.2 Storage CAUTION Environmental influences and condensate formation can destroy individual components of the analyzer! The analyzer must only be stored in acclimatized rooms. The atmosphere must be low in dust and free from aggressive vapors. If the analyzer and add-on devices are not positioned immediately after delivery or are not required for a prolonged period of time, they should best be stored in their original packag- ing.
  • Page 100: Connecting The Analyzer

    Transport and storage 10. Install the TIC condensate vessel (see section Cleaning the TC condensate vessel" p. 71). 11. Install the halogen trap and the water traps (see sections "Replacing the water traps" p. 63 and "Replacing the halogen trap" p. 65). 12.
  • Page 101 Transport and storage Connect the mains cable CAUTION Always electrically connect the system components to the multi N/C 2100 when it is switched off! Before connecting the mains cable, ensure that the main switch on the back of the equip- ment is set to "0"! Only use the IEC connection cable included in the scope of delivery for the connection to the mains supply (VDE label, 1.5 m long).
  • Page 102: Disposal

    Disposal Disposal 11.1 Waste water ENVIRONMENTAL PROTECTION The neutralized waste must be brought to the appropriate waste disposal center for correct disposal according to the appropriate legal guidelines. Waste water arises during the ongoing analysis operation of the multi N/C 2100. This water contains hydrochloric acid, diluted phosphoric acid and sample, depending on the measure- ment mode.

Table of Contents