Maintenance and Troubleshooting
5
Replace gas filter for bag version
4 Use tweezers to take out the washer and filter.
5 Replace the filter. (Agilent advises replacement of the filter at regular intervals.
Filter P/N: 3150-1017).
6 To reassemble the pagoda connector, perform
step 4
back to
step
1.
For reassembly: Make sure to manually screw the pagoda connector before using
WA RNING
the 1/4-inch wrench to tighten it to the Selector. Using a wrench directly might
wear out the screw thread and ultimately cause gas leakage, which is NOT
reversible.
Agilent Gas Sample Selector for 8860 and 8890 User Manual