Site Preparation - Agilent Technologies 8860 User Manual

Gas sample selector
Hide thumbs Also See for 8860:
Table of Contents

Advertisement

3
Installation

Site preparation

Figure 13 Shipping kit contents for the bag version Gas Sample Selector.
Site preparation
Other materials required (not provided):
34
Agilent Gas Sample Selector for 8860 and 8890 Safety and Installation Manual
1
3
6
8
Leave enough space on the left side of the GC to place the bag version selector
(40 cm).
Installation requires 1 power outlet.
1/8-inch tubing, terminating in a 1/8-inch male Swagelok fitting, that is long
enough to connect the Selector sample exhaust vent to the laboratory exhaust
system or fume hood.
Compressed air supply, terminating in a 1/8-inch male Swagelok connector
(for valve actuation) at 380 kPa (55 psi).
Purge gas supply, nitrogen, air, or helium, terminating in a 1/8-inch male
Swagelok connector.