Gas sample selector
Hide thumbs Also See for 8860:
Table of Contents

Advertisement

3
Installation
Procedure
8 Connect the actuator air supply to the Air/N
9 Connect the Vent fitting on the back of the Selector to the laboratory exhaust
40
Agilent Gas Sample Selector for 8860 and 8890 Safety and Installation Manual
Set the supply pressure to 380 kPa (55 psi).
system. When routing the tubing, avoid sharp bends or obstructions. The
exhaust tubing must be able to vent to atmospheric pressure without clogging
or backpressure. Backpressure in the vent line can unbalance the sampling
system and affect repeatability. Handle sample gas exhaust in accordance with
all laboratory and local safety guidelines.
fitting on the back of the Selector.