189II/AN
Service Manual
Hazardous Voltage
To alert you to the presence of a potentially hazardous voltage, when the Meter detects a
voltage ≥30 V or a voltage overload (OL), the
Required Equipment
Required equipment is listed in Table 2. If the recommended models are not available,
equipment with equivalent specifications may be used.
Repairs or servicing should be performed only by qualified personnel.
Equipment
Calibrator
Thermocouple Adapter
Accessory
K-type Thermocouple wire
Serial Cable, Infrared
10
Table 2. Required Equipment
Required Characteristics
AC Voltage Range: 0-1000 Vac
Accuracy: ±0.0875 %
Frequency Range: 20 Hz – 100 kHz
Accuracy:±3 %
DC Voltage Range: 0-1000 Vdc
Accuracy: ±0.0075 %
Current Range: 0-10A
AC mA - ±0.15 %
Accuracy:
AC A - ±0.21 %
Frequency range: 1 kHz – 30 kHz
DC mA - +0.0175 %
DC A - ±0.08 %
Frequency Source: 45 Hz-950 kHz
Accuracy: ±0.0026 %
Amplitude: 600 mV
Accuracy: ±5 %
Ohms Range: 0 Ω - 300 MΩ
Accuracy: ±0.0135 %
Capacitance Range: 5 nF
Accuracy: ±0.55 %
Temperature Range: 0 to 1000 °C
Accuracy: 0.2 %
K-type
K-type, mini-plug on both ends
Cable, Adapter, Fluke-Custom IRDA Optical to USB,2M
symbol is displayed.
Y
Recommended
Model
Fluke 5520A
Fluke 80AK
N/A
Fluke P/N 2166275