5.2. SETTING THE OPERATING MODE
The Revolution 4DC x-ray unit features the following operating modes:
-
En60: all the x-rays are taken at 60KV and 7mA
-
En63: all the x-rays are taken at 63KV and 6mA
-
En65: all the x-rays are taken at 65KV and 6mA
-
AUTO: the system automatically selects the best setting from amongst En60, En63 and En65 for each tooth-
patient size combination
NOTE: The current setting is displayed on the handheld for approximately 1 second for each tooth-patient
size selected before the relative exposure time is shown.
To set the operative mode, press keys
Scroll the parameters up to parameter P06 and press key
and then press key
Press key
to quit the set up menu.
5.3. RESTORING FACTORY SETTINGS
To restore the factory settings (see paragraph 3.5) press keys
Press keys
and
EN
and
.
simultaneously. "rESS" will briefly appear and the handheld will be rebooted.
OPERATOR'S MANUAL
to go to the set up menu.
. Scroll the options to find the desired operating mode
and
to go to the set up menu.
19