Lumera T surgical microscope

Intended use

Description of components

4
G-30-1682-en
Description
The Lumera T surgical microscope has been designed for the magnified
visualization of the field of view during surgical procedures in ophthal-
mology. The illumination system of the surgical microscope features red
reflex illumination and surrounding field illumination, providing very effec-
tive illumination of the field of view and optimum visualization of the red
reflex.
Note:
The illumination system of the Lumera T surgical microscope contains a
UV blocking filter as a standard feature. This helps the surgeon to reduce
the risk of phototoxic retinal injury in the patient.
The Lumera T surgical microscope comprises the following components:
1 X-Y coupling
The X-Y coupling allows motorized fine positioning of the surgical mi-
croscope in a horizontal plane. The range of travel is 40 mm x 40 mm.
The speed of travel can be set on the control panel of the suspension
system.
The X-Y coupling is provided with a recentering mechanism. When
you press reset button (2) or an appropriately programmed button of
the foot control panel (4),
– the X-Y coupling adopts its center position,
– the focusing system of the surgical microscope is reset to its initial
position (3).