INSTALLATION AND OPERATION
2.4.2.

Remote Interlock

Remote interlock is used with a door switch to prevent unintentional entry.
Once the door is opened, the interlock circuit triggers and the laser emission
is immediately shut off. The interlock connector used by iM3 Veterinary
Dental Laser is a standard 3.5mm audio jack. To use this function,
connect the remote interlock cable to the door switch and plug it into the
interlock socket located at the back of the base unit. The door switch
should be a normally closed switch. Therefore, the interlock circuit can be
cut when the door is opened, and then the laser emission is interrupted with
an error message appears on the screen. To clear this error, please close
the door and press "Clear" button on the screen.
NOTE: If you don't need this feature, don't connect the interlock
cable to the base unit.
2.4.3.

Install FastFiber Tip

The iM3 Veterinary Dental Laser provides disposable fiber tips with three
types of core diameters - 200μm, 300μm, and 400μm, in the color of pink,
blue and grey respectively. Please choose proper fiber tip before use.
fiber tips must be cleaned and sterilized before use
how to sterilize the tips, refer to Section 6.4.1.
To install, insert the handpiece sleeve from the front of the handpiece body
and screw it clockwise until snug, then insert the FastFiber tip to the
handpiece sleeve and click into place (Figure 3.6).
Operators can use a bender to bend the tip to the angle needed for best