5.2.

Power Calibration

Laser calibration is recommended to be performed every 12 months to
maintain the accuracy of the laser output. If you need power calibration
service, contact us or your authorized service representative.
NOTE: It is recommended to use your own laser power meter to
check whether the actual laser output deviates from the set value
by more than ±20%. Do not use a iM3 Laser exceeding this
limit and contact us or your authorized service representative
immediately for power calibration.
5.3.

Cleaning, Disinfection, and Sterilization

5.3.1.
Cleaning and Sterilizing Instructions for the disposable tip, the
handpiece sleeve, and the tip bender
The FastFiber tip, the handpiece sleeve, and the tip bender must be clean
and steam sterilized prior to use. The parameters of autoclaving should be:
Cycle Type: Gravity displacement;
Temperature: 121℃ (250℉);
Sterilization Time: 30 minutes;
Dry Time: 30 minutes;
Packaging Method: Pouch per AAMI recommendations
5.3.2.

Cleaning and Disinfecting Instructions for Therapy Handpiece

In order to prevent cross-contamination, the therapy handpiece should be
cleaned and disinfected prior to each use by a soft cloth dipped with
disinfectant, e.g. CaviWipes
after each use and leave it for at least 5 minutes, and then wipe it with a new
MAINTENANCE AND SERVICE
or other similar products, to remove the dust
TM
PG 26