Standby Error Codes; Code 02 : +15 V Supply Error - Siemens MULTIMOBIL 5C Troubleshooting Manual

Hide thumbs Also See for MULTIMOBIL 5C:
Table of Contents

Advertisement

Service Instructions
Multimobil 5C

8.3 Standby Error codes

8.3.1 CodE 02 : +15 V Supply Error

START
REMOVE CONNECOR X15 ON D915
& SWITCH ON THE UNIT
MEASURE +15 V ON X15
CONNECTOR
IS +15V THERE ?
YES
NO
REPLACE D915 PCB
REPLACE SMPS1
STOP
Siemens Ltd. Med India