6. Type the measured temperature into the keypad.
7. Press the "Set" button in the Temperature panel. This will set the current temperature
to the value you are measuring.
8. Press the "Mode" button again to re-enter calibration mode.
9. Press the "Cal Stats" button.
10. Press the "Save" button.
11. Press "Return" to go back to the main screen.
12. The device temperature is now calibrated!
13. If there are any errors with temperature calibration, you can reset to the Factory
Calibration from the "Cal Stats" page in calibration mode.

Replacing Electrodes

The platinum wire electrodes get thinner over time and eventually will form a break in the wire.
This can take months - years but is expected behavior. Please contact LifeCanvas at
info@lifecanvastech.com
simple:
1. Power off the device and unplug the power cord from the wall. This will ensure that the
electrode swap is safe.
2. Drain out any liquid from the device.
3. Pull the red or black wire out from the Electrode Assembly.
4. Unscrew the 4 bolts that secure the Electrode Assembly to the Chamber.
5. Wipe up any liquid with a paper towel.
6. Place the new electrode assembly in the Chamber slot.
7. Secure the Electrode Assembly with the bolts using a hex key. All 4 bolts should be
finger tight.
8. Plug the wire into the new Electrode Assembly.
9. Power on the device and fill it with 400 mL of distilled water.
10. Turn on the Auxiliary Power and let the system run for ~5 minutes to confirm the
electrode assembly is well sealed to the Chamber.

Replacing Fuse

Caution: Always unplug the power supply before replacing the fuse.
1. Power off the device and unplug the power cord from the device.
2. Remove the Fuse Drawer by pushing a small screwdriver into the slots on the left and
pulling the drawer out.
3. Remove the defective fuse and replace it with the replacement fuse (T 10A 250V).
4. Reinsert the Fuse Drawer by pushing it into the slot until you hear it click into place.
www.lifecanvastech.com
to purchase replacement electrodes. The replacement procedure is
info@lifecanvastech.com