Airway Module Calibration - Datex-Ohmeda CARESCAPE R860 User's Reference Manual

Table of Contents

Advertisement

Airway module calibration

CAUTION
2065490-001
1
Figure 4-11 • Airway module connection to the patient breathing
circuit
1.
Airway module
2.
Gas sample, gas sampling line connector on the water trap
3.
Gas sampling and spirometry tubes
4.
D-lite(+)/Pedi-lite(+) sensor
5.
Spacer (5 ml - minimum)
6.
Heat and moisture exchanger with filter (HMEF) (optional)
7.
Elbow
8.
Gas sampling line
Note
To disconnect, follow the steps in reverse order.
Airway module calibration should be performed once every two
months or when there is an indication of an error in the gas
measurements and waveforms. Use only the calibration gas and
regulator specified for the airway module. See "Calibrating the airway
module" in the "System configuration (Super User) and service"
section.
Use only GE Healthcare authorized calibration gas. Do
not use other calibration gases or the calibration will not
be successful. Dispose of calibration gas containers in
accordance with local environmental procedures.
Note
During gas calibration, CO2 is always in % regardless of the
measurement unit selected.
Install this regulator on the gas calibration cylinder:
755534-HEL (Quick cal regulator)
M1006864 (US only)
4 Setup and connections
2
3
7
6
4
8
5
4-31