Manual Cleaning:   
Prepare the detergent solution in a tank with cold water (please follow the instructions of 
the detergent manufacturer regarding application, dilution and contact time). 
A) EUP‐V73W probe 
1) The temperature of the detergent solution should be between 15‐30 °C, concentration 
is  1.6%.  Please  note  the  minimum  contact  time  of  the  detergent  in  the 
manufacturer's  instruction.  If  a  differing  detergent  is  used,  please  also  note  the 
approved material compatibility for the medical device. 
2) Immerse  the  immersible  part  of  the  probe  without  connector  into  the  diluted 
detergent  solution  (see  Fig.  17).  Wipe  the  immersible  part  of  the  probe  under  the 
surface of the detergent solution with a soft cloth to remove all visible soil. Be sure 
that all grooves of the probe are implemented during the cleaning process. 
3) The immersible part of the probe should be left in the detergent solution according to 
the specified contact time of the detergent manufacturer. 
4) Wipe  the  un‐immergible  parts  of  the  probe  with  a  soft  cloth  dipped  with  the 
detergent solution. 
5) Rinse  the  probe  with  running  tap  water  for  1  minute.  (alternatively:  immerse  the 
immersible part of the probe in a tray filled with deionized water/tap water (see Fig. 
17) for 5 min.)   
6) Visually  check  the  outer  surface  of  the  probe  for  cleanness.  If  necessary,  use 
magnifying glass for visually check. If there is still soil visible, repeat all above steps. 
B) Attachment 
1) The temperature of the detergent solution should be between 15‐30 °C, concentration 
is  1.6%.  Please  note  the  minimum  contact  time  of  the  detergent  in  the 
manufacturer's  instruction.  If  a  differing  detergent  is  used,  please  also  note  the 
approved material compatibility for the medical device.     
2) Immerse  the  Attachment  into  the  diluted  detergent  solution.  Wipe  it  under  the 
surface of the detergent solution with a soft cloth to remove all visible soil. Be sure 
that all grooves of the attachment are implemented during the cleaning process. 
3) The  Attachment  should  be  left  in  the  detergent  solution  according  to  the  specified 
contact time of the detergent manufacturer. 
4) Rinse the Attachment with running tap water for 1 minute. (alternatively: immerse it 
in a tray filled with deionized water/tap water (see Fig. 17) for 5 min.)   
5) Visually  check  the  outer  surface  of  the  Attachment  for  cleanness.  If  necessary,  use 
magnifying glass for visually check. If there is still soil visible, repeat all above steps. 
 
 
 
 
 
 
‐19‐ 