1) Special  attention  should  be  paid  for  a  probe  cover  made  of  latex.  Latex  may  cause 
allergic  reactions  such  as  itching,  rubor,  urticaria,  swelling,  fever,  anhelation, 
wheezing, depression of blood pressure, and shock. 
For  the  patients  suspected  of  latex  allergy,  do  not  use  the  latex‐containing  medical 
devices.  If  your  patient  shows  any  of  above  mentioned  symptoms  during  the 
operation, stop the use of the latex‐containing medical devices immediately and take 
an appropriate treatment to the patient. 
2) The puncture adapter EZU‐PA7V and the recommended probe covers are disposable 
and must not be reused. 
3) Clean, disinfect and sterilize the probe and the puncture adapter when the probe cover 
is torn. 
4) Be  careful  about  the  use  of  a  needle  cannula  with  the  puncture  adapter.  If  the 
insulation coating of the needle cannula is damaged, it may cause a burn to tissue.   
5) Use a needle whose size is appropriate for the puncture adapter, otherwise looseness 
or  tightness  of  the  needle  could  result  in  biopsy  at  an  unintended  part  and  injury  to 
the patient.   
6) Display the  needle  guideline during  biopsy.  Regarding  the  needle guideline,  refer  to 
the operation manual of the ultrasound diagnostic scanner. 
 
 
 
WARNING
 
‐9‐ 
Q1E‐EP1430