Controlling the Operating Table
Avoiding collisions
76
The individual functions on the column keypad can be selected only
once it is unlocked, or optionally via the two-key control.
The end position of a function depends on the patient's weight, the
composition of the operating table and which functions have
already been performed on the operating table (current operating
table position). The adjustment ranges of the functions may be
restricted. The speed at which the function is executed is dependent
on the patient's weight. The speed may be lower for a heavier
patient.
The function also provides increased safety in the operating theater.
The operating table monitors the movement range of the operating
table top and prevents collisions with the floor, the running gear and
the operating table column.
Movement ranges are only monitored with the following operating
table equipment:
– single-jointed head section or double-jointed head section on the
two-part table top head end
– upper back section and head section single joint or head section
double joint head end on the one-part table top
– one-part leg section or Carbon 600 table top segment on the
table top foot end
– The mechanical joints on the section segments must be in the level
position.
Different operating table equipment and accessories on the side
rails are not monitored with regard to collisions. In this case,
incorrect warning messages may be produced, or no warning
message when one is required.
A warning tone during the function indicates that the maximum
adjustment range will soon be reached. The function stops
automatically prior to a possible collision with the floor, the running
gear or the operating table column.
Objects (for example, devices or furnishings) within the range of
movement of the operating table top are not detected by the
operating table. In this case the function must be stopped by the
operator before a collision occurs.
The monitoring of collisions takes place regardless of the load on
the operating table.
Instruction manual Operating table TruSystem 7000 – 1 730 263 – 11/2016

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the TruSystem 7000 and is the answer not in the manual?

Subscribe to Our Youtube Channel

This manual is also suitable for:

Trusystem 7000 vTrusystem 7000 uTrusystem 7000 mbwTrusystem 7000 mbw vTrusystem 7000 u mbTrusystem 7000 u14 mbw ... Show all

Table of Contents

Save PDF