Setting The Probe Molar Mass; Purging Moisture From Probe (Purification) - Extech Instruments RH600 User Manual

Dew point meter
Hide thumbs Also See for RH600:
Table of Contents

Advertisement

4
Probe Use and Configuration

4.4 Setting the Probe Molar Mass

Set the probe's Molar Mass setting to match the gas under test. See the steps
below. The molar mass of air is 28.96 g/mol and the molar mass of SF6 is
146.05 g/mol.
• Press F3
• Use the left or right arrow to scroll to Environment.
• Press F3
• At the PPM option (2nd row), press F3.
• Use the arrow buttons to set the molar mass value. Press F1
and to return to the previous menu.
Figure 4.2 Use the arrow buttons to set the probe molar mass to match the molar mass of
the gas under test.

4.5 Purging Moisture from Probe (Purification)

When you activate the Purification option, the probe enters a five-minute
purge cycle. See the steps below.
• Press F3
• Use the left or right arrow to scroll to Environment.
• Press F3
• At the Purification option, press F3.
• The display will show an alert message indicating that a five–minute purge
cycle is in process. The meter menus are not accessible during the purge
cycle. When the purge cycle is complete, the alert message will switch off
and the meter menus will be fully operational.
#NAS100113; r. AA/87024/87024; en-US
to open the Programming menu.
to open the Environment menu.
to open the programming menu.
to open the Environment menu.
to confirm