5
REPROCESSING IN PROFESSIONAL
HEALTH INSTITUTIONS
Dry the connection tubing after each use [see: Connection
tubing, page 37].

5.1 Reprocessing cycles

Single patient use
Nebuliser excluding
connection tubing and
accessories (e.g.
masks)
Before a change of patient
Nebuliser excluding
connection tubing but
including accessories
(e.g. mask)
Connection tubing

5.2 Processing limits

Nebuliser and accessories,
disinfection
Nebuliser and accessories,
sterilisation
Connection tubing
– 28 –
– Clean immediately after every use
– Disinfect once per week
– Cleaning
– Disinfection
– Sterilisation
Mechanical cleaning with disinfection
300 reprocessing cycles,
max. 1 year
100 reprocessing cycles,
max. 1 year
50 reprocessing cycles,
max. 1 year