Image Quality
Adjusting the grid switch
The switch S1, see
the grid is in the turning point. The switch must be correctly adjusted. otherwise the high
voltage will be active when the grid is standing still.
To minimize this risk of the high voltage being active when the grid is standing still, the
high voltage should be disabled for as long as possible.
1. Move the grid to the left end position.
2. Loosen the three screws. Now it's possible to adjust the grid switch (S1) alongside the
grid (see
Fig. 44: Adjusting grid switch S1
Siemens AG
Medical Solutions
(Fig. 44 / p.
67), on the detector table disables the high voltage when
(Fig. 44 / p.
67)).