Siemens MAMMOMAT Novation DR Manual page 26

Hide thumbs Also See for MAMMOMAT Novation DR:
Table of Contents

Advertisement

26
CAUTION
9. Check the temperature logger with a ball pen, as shown in
Fig. 10: Check temperature logger
NOTE
NOTE
NOTE
MAMMOMAT Novation DR
Handle the detector with extreme care!
Any dust or fingerprints in this area may affect final image quality.
Temperature extremes may cause permanent damage.
Avoid touching the thin film top cover that protects the
imaging area. The imaging electronics under the front
cover are extremely sensitive. The actual image area
starts at the front edge and is approximately 25 mm
inside the other three edges.
A green LED indicates that the transportation was OK and the de-
tector can be installed.
A red LED indicates that the detector is defective and has to be re-
placed.
You have to send the temperature logger to the factory using the
regular Customer Services return process.
You'll find all documents required for sending it enclosed in the
detector package (black box).
Enter the detector serial number, your district address and instal-
lation date before you send it.
The detector package has to be disposed of locally after the instal-
lation.
SPB7-250.815.06.02.02
12.05
Detector installation
(Fig. 10 / p.
Page 26 of 116
CS SD 24
26).
Siemens AG