Cerabar PMC11, PMC21, PMP11, PMP21, PMP23
Endress+Hauser
7
Operation options
7.1
Plug-on display PHX20 (optional)
No display or other operation facility is required to operate the device. However, devices
with a valve plug can be fitted with the optional local display PHX20.
A 1-line liquid crystal display (LCD) is used. The local display shows measured values, fault
messages and information messages. The device display can be turned in 90° steps.
Depending on the orientation of the device, it is therefore easy to read the measured
values.
7.1.1
Storage conditions
• Use original packaging.
• Storage temperature range: –30 to +80 °C (–22 to +176 °F)
7.1.2
Installation
1. Place seals between the sensor/plug-on display and the plug-on display/connector.
2. Fit the plug-on display (2) between the connector (3) and connector socket (1) of the
sensor.
3. Replace the securing screw (4) with the longer screw included in the delivery.
4. An adhesive label (included in the delivery) with information on the technical unit
can be fixed beneath the LED display.
Mounting dimensions
7.1.3
Technical data
Display:
Digit height:
Display range:
Accuracy:
1
2
3
42 (1.65)
48 (1.89)
4-digit, red LED display
7.62 mm; programmable decimal point setting
-1999 to 9999
0.2% of the span ±1 digit
Operation options
4
A0022208
A0022210
23