ET35 Series Digital Bridges(LCR)Operating Instructions
The page is shown below
Set the reference value
Set the mode of deviation
3.2 Correction page
Press the [Meas Setup] key and then press the [correction] function key to enter the
user correction page. It allows the user to correct open / short / load to compensate for
any errors due to interference factors such as stray admittance and residual impedance.
In addition, users can also choose the proper length of cable on the correction page.
ET35 offers two correction modes. One is the use of interpolation method to correct
the open / short for all frequency points. Another is to correct open / short / load for the
currently set frequency points.
User Correction page
Hangzhou Zhongchuang Electron Co., Ltd.
27