Note

11.1 Moving the System

Warning:
1.
2.
3.
4.
5.

Figure 7 Transport Position Knob

6.
MAN-04858 Revision 006
InSight FD operators should read the Radiation Safety section
before operating the system.
Before moving the system, the flex-arm and C-arm must be in the
locked storage position to prevent tipping.
Ensure that the system is powered off.
See
Section 11.5 Powering Down the System
Unplug the power cord and wrap it around the cord guides on the
back of the cabinet.
Store the foot switch in the foot switch shelf.
Flip the monitor forward and down.
Place the flex-arm and C-arm in the locked position
the Transport Position Knob on the side of the C-arm housing.
Lock the rear casters to prevent the system from moving. The
swiveling angle of the rear casters can be locked to prevent lateral
movement of the system or unlocked to allow lateral movement.
InSight FD Mini C-arm User Guide
on Page 22.)
(Figure
Using the System
7) using
19