Machine Operation - Omegasonics PowerLift 4560 Operation & Instruction Manual

Ultrasonic cleaning system
Table of Contents

Advertisement

MACHINE OPERATION

·
Press START SCREEN if not currently in that mode.
·
Display will be as shown on page 5.
·
The color shown
indicates
selected button.
Pressing Green or Yellow function buttons indicates that this function is currently
disengaged and will be engaged when pressed.
·
Pressing Red function button indicates that this function is currently active and
will be deactivated when pressed.
·
To activate the heat, press the Yellow HEAT ON button. This button will change
states to Red and will display HEAT OFF.
·
The temperature controller is factory preset for 150° F. To change the preset, follow
the instructions found on page 6.
·
Actual bath temperature will display in the blue box - TANK TEMP. ACTUAL.
·
To activate the ultrasound, press the Lime Green SOUND CYCLE TIME button. The
button color will change to a darker Forest Green color and the cycle will begin
based on the programmed cycle time.
·
Operate the ultrasound for fifteen (15) minutes after the tank has been initially filled
and heated. This process is called degassing and helps eliminate any air from the
water in the tank.
·
To deactivate the ultrasound before the timer has completed its' cycle, press the Red
SOUND CYCLE TIME - STOP button. The Red button will illuminate brighter and the
cycle will cease.
·
The Sound Cycle Time is factory preset for 15:00 minutes.
·
To activate the agitation, press the Lime Green AGITATION CYCLE TIME button.
The button color will change to a darker Forest Green color and the cycle will begin
based on the programmed cycle time.
·
To deactivate the agitation before the timer has completed its' cycle, press the Red
AGITATION CYCLE TIME - STOP button. The Red button will illuminate brighter
and the cycle will cease.
·
If the Agitation Cycle is stopped before the cycle time has expired and in any position
but the complete lower position, the lift will lower to the complete lower position.
·
The Agitation Cycle Time is factory preset for 05:00 minutes.
·
To activate the filtration system, press the Green PUMP ON button. This button will
change states to Red and will display PUMP OFF.
·
You are now ready to begin the cleaning process.
·
Manually open the lid to the full open upright position.
·
Press and hold the Green RAISE LIFT button until the lift reaches the height of the
tank lip. The lift will not rise above this lip elevation.
·
Load parts onto the lift platform.
·
Press and hold the Red LOWER LIFT button until the lift reaches the bottom of the
tank. The lift platform will lower to the tank bottom just above the immersible
transducer packs.
·
Manually close the lid. Lid should not be closed until the lift has been lowered to the
bottom of the tank to ensure the parts being cleaned do not hit the inside of the lid.