The heat circuit, agitation circuit and ultrasound can all be operated simultaneously.
·
The heat circuit and pump filtration circuit can be operated simultaneously.
·
The ultrasound and pump filtration circuits are electrically interlocked and the
filtration system will not operate during the ultrasonic cleaning process.
·
If a low liquid level condition arises, a Red LOW TANK LEVEL flag will appear on the
main START Screen to the left of the Tank Temperature indicator. The low level
sensor protects the ultrasound, heat and pump agitation circuits.
LIFT-TABLE PNEUMATIC ADJUSTMENTS
·
To adjust air pressure to the machine you will pull the grey knob on the left hand side
up and turning it clockwise-CW will increase the air pressure, it should be set to no
less than 50PSI. Once set you will push the knob down to lock in place.
·
To adjust the lift speed in the raising operation, turning the white valve adjustment on
the UP valve counter clockwise –CCW will increase the flow of air through the valve
and make the lifting action faster. Clockwise-CW will inversely cause the lift to raise
more slowly.
·
To adjust the lift speed in the lowering operation, turning the white valve adjustment
on the DOWN valve counter clockwise –CCW will increase the flow of air through the
valve and make the lowering action faster. Clockwise-CW will inversely cause the lift
to lower more slowly.
·
NOTE: The valves are factory adjusted and should not be altered.