Changing Filters - Omegasonics PowerLift 4560 Operation & Instruction Manual

Ultrasonic cleaning system
Table of Contents

Advertisement

CHANGING FILTERS

·
Prior to changing the filters, the water level in the tank must be below the level of the
filter. Follow the prescribed equipment maintenance procedures above.
·
Filters should be changed periodically depending on the usage of the equipment and
the contaminants being removed.
·
Individually unscrew counterclockwise the two (2) large blue filter cartridge holders.
The cartridges will be full of water.
·
Pour the water out of the cartridges and remove the filters. Be careful not to lose the
rubber O-ring gasket located on the inside of the filter cartridge. Absence of this
gasket will cause the filter(s) to leak.
·
Replace with new filter(s) and re-install the cartridges.
·
The recommended 50 Micron Polycarbonate Filter is the primary filter and is located
on the right side of the filtration unit.
·
The Carbon Filter is the secondary oil filter and is located on the left side of the
filtration system. The carbon filter removes oil and grease and will require changing
more often than the recommended Micron Polycarbonate Filter.