The furnace door can easily be converted to a right hand swing door as fallows:
1. Open and support the furnace door. Remove the four bolts holding the door
assembly to the furnace frame.
2. Remove the contact switch mounted below the door.
3. Install door on right hand side of the chamber frame, using the bolts in the
holes provided. Install the contact switch in the mounting location provided at
the right of old location.
4. Check alignment of the door insulation with the chamber.
Sideways adjustment can be made by loosening the door insulation supports and
moving the insulation plug. Vertical adjustment can be made by placing or
removing spacers on the door hinge.