Gas Inlet Tube; Figure 2 Thermocouple - Thermo Scientific BF51841C Operation Manual

Laboratory box furnaces
Table of Contents

Advertisement

Installation
Gas Inlet Tube (Refer Figure 3. Gas Inlet Tube Assembly)
near the front of the furnace. If mechanical adjustment is necessary, slight
bending of the switch arm can be done.
7. Replace and secure the small cover on the Rear panel and right side panel of
the furnace using the necessary screws.
Gas Inlet Tube (Refer Figure 3. Gas Inlet Tube Assembly)
The gas inlet tube assembly has been packaged separately to avoid breakage
during shipping and handling.
Even if you do not intend to use the gas inlet, you must install the assembly before
operating the furnace. The only tool you need is a Phillips head screwdriver.
To install the gas inlet assembly:
1. Carefully remove the assembly from the package and inspect for any damage.
2. Remove the two mounting screws from the rear housing panel of the furnace.
3. Insert the ceramic tube end through the access hole in the rear of the furnace
and guide the tube into the back of the chamber.
4. Align the mounting holes in the rear housing panel with the holes in the gas
inlet tube assembly and secure the assembly with the mounting screws.
4-2
Figure 2. Thermocouple