Starting The Upgrade Utility; Setting The Communication Link; Programming The Instrument; Uninstalling The Upgrade Software - Hach DR/2400 Manual

Portable spectrophotometer
Hide thumbs Also See for DR/2400:
Table of Contents

Advertisement

Upgrading the Instrument Software

9.4 Starting the Upgrade Utility

9.5 Setting the Communication Link

9.6 Programming the Instrument

9.7 Uninstalling the Upgrade Software

Upgrading the Instrument Software
Page 54
On the computer, click
Start
.
Upgrade Utility
The upgrade information window will appear. It displays information about the
upgrade utility. Any last minute changes made to the utility that might not be
included in this manual will be displayed here. Use the scroll bar to the right of
the window to view additional information.
In order for the upgrade utility to program a DR/2400 Spectrophotometer, it
must be able to establish a communication link with the instrument through one
of the computer's serial ports. Using the serial port selection box in the DR/2400
DR/2400 Upgrade software, choose the serial port used to connect the computer
to the instrument. Click on the down arrow to view a list of the available serial
ports, then click on the appropriate port selection.
The DR/2400 Spectrophotometer must be set up as described in Communications
with a Printer or Computer on page 24.
Once the instrument is connected to the computer and the communication
settings are made, programming can begin. Click on the
the programming sequence.
1. When the upgrade utility completes its system checks, a confirmation
prompt will appear. The estimated time to program the instrument is
also displayed.
2. Click on
to upgrade the instrument, click on
Yes
programming sequence.
After selecting
, the instrument switches to the programming mode and
Yes
the upgrade begins. Do not disconnect or turn off the instrument or
computer while programming. The status bar indicates the progress of the
programming sequence.
3. A dialog box is displayed when programming is complete. Click on the
button to return to the utility form.
4. At this time another instrument can be programmed, or click on the
button to exit the upgrade utility.
Future releases of the DR/2400 Spectrophotometer Upgrade Utility will
automatically replace an older version on the hard drive, so uninstalling this
version is not required. To remove the upgrade utility, use the following
procedure.
From Windows Explorer, click on the DR/2400 Software Upgrade Utility icon to
highlight it — do not double-click on it. Select the
option. Choose
Delete
Yes
,
,
Programs
Hach Utilities
in the dialog box to confirm the deletion.
,
DR/2400 Software
button to start
Program
or
to cancel the
No
Cancel
Close
menu and choose the
File
5940018Upgrade.fm
OK

Advertisement

Table of Contents
loading

Table of Contents